Clinical evaluation of an expert system for arteriovenous fistula assessment

被引:8
作者
Chanliau, Jacques [1 ]
Charasse, Christophe [2 ]
Rose, Cedric [3 ]
Bene, Bernard [4 ]
机构
[1] ALTIR, Vandoeuvre Les Nancy, France
[2] St Brieuc Hosp Complex, St Brieuc, France
[3] Diatelic, F-54600 Villers Les Nancy, France
[4] Gambro Ind, Meyzieu, France
关键词
Hemodialysis; Expert systems; Arteriovenous surgical shunt; Renal dialysis machine; Markov chains; Bayes theorem; ACCESS BLOOD-FLOW; IONIC DIALYSANCE; VASCULAR ACCESS; MODELS;
D O I
10.5301/ijao.5000364
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The monitoring of ionic dialysance in hemodialysis allows early detection of arterio-venous fistula stenosis. One limitation to the practical use of ionic dialysance is that the analysis is very time consuming on a majority of normal cases. The purpose of the study is to evaluate the utility of an expert system reproducing a human analysis and allowing continuous monitoring of the ionic dialysance by helping the physician to focus his or her expertise on the abnormal cases. The method is based on a Bayesian model that analyzes the blood flow rate, the ionic dialysance, and the venous and arterial pressures measured on the extra corporeal circuit. The clinical evaluation was performed on 90 dialysis patients at the hospital dialysis center of Saint Brieux in France with a history of at least four consecutive months of validated recording. The retrospective automated analysis was evaluated in comparison to vascular access problems identified from invasive investigation or treatment. The sensitivity of the automated analysis is 92% with a specificity of 75%. As a conclusion we suggest that this expert system could be used in a continuous vascular access monitoring procedure consisting in a weekly review of the patient population at the dialysis center. The patients with the highest risk score need a further investigation of their historical data and their medical history in order to decide whether or not to perform an invasive intervention.
引用
收藏
页码:809 / 815
页数:7
相关论文
共 11 条
  • [1] Is ionic dialysance useful for early detection of vascular access dysfunction? Six illustrative cases
    Fontsere, Nestor
    Blasco, Miquel
    Arias, Marta
    Maduell, Francesc
    Vera, Manel
    Burrel, Marta
    Barrufet, Marta
    Isabel Real, Maria
    Mestres, Gaspar
    Campistol, Josep M.
    [J]. HEMODIALYSIS INTERNATIONAL, 2011, 15 (01) : 108 - 111
  • [2] Telehomecare for vulnerable populations: The evaluation of new models of care
    Gagnon, Marie-Pierre
    Lamothe, Lise
    Hebert, Marilynne
    Chanliau, Jacques
    Fortin, Jean-Paul
    [J]. TELEMEDICINE JOURNAL AND E-HEALTH, 2006, 12 (03): : 324 - 331
  • [3] Hanoy M, 2007, NEPHROL THER, V3
  • [4] Detection of vascular access stenosis by measurement of access blood flow from ionic dialysance
    Mercadal, L
    Challier, E
    Cluzel, P
    Hamani, A
    Boulechfar, H
    Boukhalfa, Z
    Izzedine, H
    Bassilios, N
    Barrou, B
    Deray, G
    Petitclerc, T
    [J]. BLOOD PURIFICATION, 2002, 20 (02) : 177 - 181
  • [5] Determination of access blood flow from ionic dialysance:: Theory and validation
    Mercadal, L
    Hamani, A
    Béné, B
    Petitclerc, T
    [J]. KIDNEY INTERNATIONAL, 1999, 56 (04) : 1560 - 1565
  • [6] Merino JL, 2006, NEFROLOGIA, V26, P481
  • [7] Pearl J., 1988, PROBABILISTIC REASON, DOI DOI 10.1016/C2009-0-27609-4
  • [8] A TUTORIAL ON HIDDEN MARKOV-MODELS AND SELECTED APPLICATIONS IN SPEECH RECOGNITION
    RABINER, LR
    [J]. PROCEEDINGS OF THE IEEE, 1989, 77 (02) : 257 - 286
  • [9] Rose C, 2010, INT J ARTIF ORGANS, V33, P485
  • [10] Rose C., 2005, P 17 IEEE INT C TOOL