A Robot Assisted High-flow Portable Cyclone Sampler for Bacterial and SARS-CoV-2 Aerosols

被引:20
作者
Li, Xinyue [1 ]
Chen, Haoxuan [1 ]
Qi, Xiao [2 ]
Peng, Yijiao [1 ]
Zhou, Lian [3 ]
Ma, Jianxin [2 ]
Yao, Maosheng [1 ]
机构
[1] Peking Univ, Coll Environm Sci & Engn, State Key Joint Lab Environm Simulat & Pollut Con, Beijing 100871, Peoples R China
[2] Ctr Dis Control & Prevent Chaoyang Dist Beijing, Beijing, Peoples R China
[3] Jiangsu Prov Ctr Dis Control & Prevent, Dept Environm & Hlth, Nanjing, Peoples R China
基金
中国国家自然科学基金;
关键词
Bioaerosol; Cyclone sampler; High-flow sampling; Portable sampler; BIOAEROSOL; PERFORMANCE; AIR; EFFICIENCIES; THREATS; SYSTEM;
D O I
10.4209/aaqr.210130
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Airborne transmission of infectious diseases attracts great attention since the COVID-19 pandemic. Yet, there has been an intense dispute about aerosol transmission of the disease, which is largely due to lack of qualified instruments for studying the subject. Air sampling plays a critical role in all air pollution related study, and particularly critical for airborne pathogen detection. Here, we designed and evaluated a portable and high volume (400 L min-1) cyclone sampler named as Yao-CSpler using aerosolized Polystyrene (PS) uniform microspheres, Bacillus subtilis var. niger, Pseudomonas fluorescens, and both indoor and outdoor air particles. The experimental cutoff size of the Yao-CSpler was demonstrated to be 0.58 m (while the calculated theoretical value is 1.84 m), and the sampler has shown stable microbial collection performances for bacteria, fungi, and even viruses. The sampler had a physical collection efficiency of close 100% for particles of larger than 1 m. Jet-to-liquid distance and sampling duration were shown to substantially influence the sampler performance. Given the same sampling duration, the performances of the Yao-CSpler were significantly higher than those of the traditional BioSampler (SKC Inc.) in terms of samples' bacterial diversity. The developed sampler coupled with a robot has been successfully applied to sampling airborne SARS-CoV-2 in both Wuhan and Beijing during the COVID-19 outbreaks. With a high sampling flow, the Yao-CSpler was shown to be able to collect the SARS-CoV-2 with a detectable concentration level down to 9-219 viruses m-3 in clinical settings housing COVID-19 patients. Further more efficient bioaerosol sampler, which is able to rapidly capture low level pathogenic agents, is urgently required to better understand and confront airborne transmission of infectious diseases.
引用
收藏
页数:13
相关论文
共 58 条
[1]   Evaluation of bioaerosol samplers for the detection and quantification of influenza virus from artificial aerosols and influenza virus-infected ferrets [J].
Bekking, Christian ;
Yip, Lily ;
Groulx, Nicolas ;
Doggett, Nathan ;
Finn, Mairead ;
Mubareka, Samira .
INFLUENZA AND OTHER RESPIRATORY VIRUSES, 2019, 13 (06) :564-573
[2]   Performance of the Coriolis air sampler, a high-volume aerosol-collection system for quantification of airborne spores and pollen grains [J].
Carvalho, E. ;
Sindt, C. ;
Verdier, A. ;
Galan, C. ;
O'Donoghue, L. ;
Parks, S. ;
Thibaudon, M. .
AEROBIOLOGIA, 2008, 24 (04) :191-201
[3]   Collection efficiency of liquid-based samplers for fungi in indoor air [J].
Chang, Ching-Wen ;
Ting, Yun-Tzu ;
Horng, Yu-Ju .
INDOOR AIR, 2019, 29 (03) :380-389
[4]   Development of a personal sampler for collecting fungal spores [J].
Chen, B ;
Feather, GA ;
Maynard, A ;
Rao, CY .
AEROSOL SCIENCE AND TECHNOLOGY, 2004, 38 (09) :926-937
[5]   A high-flow portable biological aerosol trap (HighBioTrap) for rapid microbial detection [J].
Chen, Haoxuan ;
Yao, Maosheng .
JOURNAL OF AEROSOL SCIENCE, 2018, 117 :212-223
[6]   Detection of air and surface contamination by SARS-CoV-2 in hospital rooms of infected patients [J].
Chia, Po Ying ;
Coleman, Kristen Kelli ;
Tan, Yian Kim ;
Ong, Sean Wei Xiang ;
Gum, Marcus ;
Lau, Sok Kiang ;
Lim, Xiao Fang ;
Lim, Ai Sim ;
Sutjipto, Stephanie ;
Lee, Pei Hua ;
Son, Than The ;
Young, Barnaby Edward ;
Milton, Donald K. ;
Gray, Gregory C. ;
Schuster, Stephan ;
Barkharn, Timothy ;
De, Partha Pratim ;
Vasoo, Shawn ;
Chan, Monica ;
Ang, Brenda Sze Peng ;
Tan, Boon Huan ;
Leo, Yee-Sin ;
Ng, Oon-Tek ;
Wong, Michelle Su Yen ;
Marimuthu, Kalisvar .
NATURE COMMUNICATIONS, 2020, 11 (01)
[7]   Development of an automated wet-cyclone system for rapid, continuous and enriched bioaerosol sampling and its application to real-time detection [J].
Cho, Yu Sung ;
Hong, Seung Chan ;
Choi, Jeongan ;
Jung, Jae Hee .
SENSORS AND ACTUATORS B-CHEMICAL, 2019, 284 :525-533
[8]   Field sampling of indoor bioaerosols [J].
Cox, Jennie ;
Mbareche, Hamza ;
Lindsley, William G. ;
Duchaine, Caroline .
AEROSOL SCIENCE AND TECHNOLOGY, 2020, 54 (05) :572-584
[9]   Operative and Technical Modifications to the Coriolis® μ Air Sampler That Improve Sample Recovery and Biosafety During Microbiological Air Sampling [J].
de Sousa, Nuno Rufino ;
Shen, Lei ;
Silcott, David ;
Call, Charles J. ;
Rothfuchs, Antonio Gigliotti .
ANNALS OF WORK EXPOSURES AND HEALTH, 2020, 64 (08) :852-865
[10]   Comparative Testing and Evaluation of Nine Different Air Samplers: End-to-End Sampling Efficiencies as Specific Performance Measurements for Bioaerosol Applications [J].
Dybwad, Marius ;
Skogan, Gunnar ;
Blatny, Janet Martha .
AEROSOL SCIENCE AND TECHNOLOGY, 2014, 48 (03) :282-295