Improved screening capabilities in forensic toxicology by application of liquid chromatography-tandem mass spectrometry

被引:49
作者
Thieme, D
Sachs, H
机构
[1] Inst Doping Anal & Sports Biochem, D-01731 Kreischa, Germany
[2] Inst Forens Med, Munich, Germany
关键词
liquid chromatography-tandem mass spectrometry; drug monitoring; forensic toxicology; hair analysis;
D O I
10.1016/S0003-2670(03)00563-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The introduction of the upcoming instrumentation liquid chromatography-mass spectrometry (LC-MS) into the field of forensic toxicology was primarily focused on crucial substances, undetectable by current analytical techniques. Typically, polar substances (which cannot be derivatised properly) thermolabile and high molecular weight compounds were first targets of the analytical technique. Successful application of LC-tandem mass spectrometry (API 2000, Applied Biosystems) in multiple reaction monitoring (MRM) mode to the identification of alprazolam (>1 pg/mg), dothiepine (>0.4 pg/mg) and piritramide (>2 pg/mg) in hair is presented in this paper. Moreover, the quantification of cocaine and benzoylecgonine was performed in the grey zone of suspicious hair samples (66 cases with positive response below cut-off in immunoassay screening but negative GC-MS confirmation). A high correlation between both techniques and a reliable substance confirmation by LC-MS-MS could be verified down to cocaine concentrations of 20 pg/mg in hair. An extension of the concept of MRM experiments to larger groups of analytes was attempted successfully. An adaptable combination of experiment modules (e.g. stimulants, narcotics, benzodiazepines, antidepressants/neuroleptics) to screening methods provides an efficient tool for a sensitive and specific systematic toxicological analysis (STA) in 'general unknown' cases. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:171 / 186
页数:16
相关论文
共 74 条
  • [11] Decaestecker TN, 2000, RAPID COMMUN MASS SP, V14, P1787, DOI 10.1002/1097-0231(20001015)14:19<1787::AID-RCM94>3.0.CO
  • [12] 2-S
  • [13] Confirmatory analysis of residues of stanozolol and its major metabolite in bovine urine by liquid chromatography-tandem mass spectrometry
    Draisci, R
    Palleschi, L
    Marchiafava, C
    Ferretti, E
    Quadri, FD
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2001, 926 (01) : 69 - 77
  • [14] Chromatographic screening techniques in systematic toxicological analysis
    Drummer, OH
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 1999, 733 (1-2): : 27 - 45
  • [15] Fitzgerald RL, 1999, CLIN CHEM, V45, P1224
  • [16] LC-EI-MS determination of veratridine and cevadine in two fatal cases of Veratrum album poisoning
    Gaillard, Y
    Pepin, G
    [J]. JOURNAL OF ANALYTICAL TOXICOLOGY, 2001, 25 (06) : 481 - 485
  • [17] Simultaneous screening and quantitation of 18 antihistamine drugs in blood by liquid chromatography ionspray tandem mass spectrometry
    Gergov, M
    Robson, JN
    Ojanperä, I
    Heinonen, OP
    Vuori, E
    [J]. FORENSIC SCIENCE INTERNATIONAL, 2001, 121 (1-2) : 108 - 115
  • [18] Gergov M, 2000, J MASS SPECTROM, V35, P912, DOI 10.1002/1096-9888(200007)35:7<912::AID-JMS19>3.0.CO
  • [19] 2-4
  • [20] Postmortem tissue concentrations of venlafaxine
    Goeringer, KE
    McIntyre, IM
    Drummer, OH
    [J]. FORENSIC SCIENCE INTERNATIONAL, 2001, 121 (1-2) : 70 - 75