Excimer laser for pacemaker and defibrillator lead extraction: Techniques and clinical results

被引:27
作者
Gilligan, DM
Dan, D
机构
[1] Hunter Holmes McGuire VA Med Ctr, Dept Internal Med, Div Cardiol, Cardiac Electrophysiol Lab, Richmond, VA 23249 USA
[2] Virginia Commonwealth Univ, Med Coll Virginia, Richmond, VA 23298 USA
关键词
defibrillator; electrophysiology; excimer; laser; lead extraction; pacemaker; ultraviolet;
D O I
10.1007/PL00011340
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Pacing and defibrillation leads may need to be removed for several reasons including infection, interference with others leads, lack of vascular access or redundancy. However, the removal of chronically implanted leads is a major technical challenge because of the extensive adhesions that develop along the course of the leads over time. The techniques to remove chronic leads have been greatly facilitated by the development of an excimer laser sheath. We undertook an analysis of our experience with laser extraction in the first 50 leads attempted. An excimer laser sheath system, developed by the Spectranetics Corporation, was used to extract 50 chronically implanted leads in 34 patients. The mean patient age was 64 +/- 12 years, all were male and the average duration that the leads had been implanted was 5.0 +/- 3.9 years, Two-thirds of the leads were pacemaker and one-third were defibrillator leads. There was a 100% clinical success rate and 48 of the 50 leads were completely removed. There were no major complications. There was one minor complication of subclavian vein thrombosis and two haemodynamically non-significant episodes of air embolism. The main limitation observed was failure of the excimer laser sheath to advance in 18% of cases, probably due to the presence of calcified adherences on leads. Two strategies were found useful to deal with this problem: under the clavicle stainless-steel sheaths were used to break up calcified adherences and within the venous system the laser sheath was upsized in order to advance over the calcification on the lead. It was concluded that excimer laser has greatly facilitated the removal of chronically implanted pacemaker and defibrillator leads. There is a high success rate and low complication rate in our experience. The main limitation of laser is the presence of calcified adherences.
引用
收藏
页码:113 / 121
页数:9
相关论文
共 22 条
[1]   Partial rupture of the tricuspid valve after extraction of permanent pacemaker leads: Detection by transesophageal echocardiography [J].
Assayag, P ;
Thuaire, C ;
Benamer, H ;
Sebbah, J ;
Leport, C ;
Brochet, E .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (06) :971-974
[2]   Learning curve characteristics of pacing lead extraction with a laser sheath [J].
Bracke, FA ;
Meijer, A ;
Van Gelder, B .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (11) :2309-2313
[3]   Intravascular extraction of problematic or infected permanent pacemaker leads: 1994-1996 [J].
Byrd, CL ;
Wilkoff, BL ;
Love, CJ ;
Sellers, TD ;
Turk, KT ;
Reeves, R ;
Young, R ;
Crevey, B ;
Kutalek, SP ;
Freedman, R ;
Friedman, R ;
Trantham, J ;
Watts, M ;
Schutzman, J ;
Oren, J ;
Wilson, J ;
Gold, F ;
Fearnot, NE ;
Van Zandt, HJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (09) :1348-1357
[4]  
BYRD CL, 1985, J THORAC CARDIOV SUR, V89, P142
[5]   INTRAVASCULAR LEAD EXTRACTION USING LOCKING STYLETS AND SHEATHS [J].
BYRD, CL ;
SCHWARTZ, SJ ;
HEDIN, NB ;
GOODE, LB ;
FEARNOT, NE ;
SMITH, HJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12) :1871-1875
[6]  
BYRD CL, 1996, PACING CLIN ELECTROP, V19, P567
[7]   INTRAVASCULAR EXTRACTION OF CHRONIC PACEMAKER LEADS - EFFICACY AND FOLLOW-UP [J].
COLAVITA, PG ;
ZIMMERN, SH ;
GALLAGHER, JJ ;
FEDOR, JM ;
AUSTIN, WK ;
SMITH, HJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (12) :2333-2336
[8]   Gross and microscopic pathological changes associated with nonthoracotomy implantable defibrillator leads [J].
Epstein, AE ;
Kay, GN ;
Plumb, VJ ;
Dailey, SM ;
Anderson, PG .
CIRCULATION, 1998, 98 (15) :1517-1524
[9]   Initial experience with larger laser sheaths for the removal of transvenous pacemaker and implantable defibrillator leads [J].
Epstein, LM ;
Byrd, CL ;
Wilkoff, BL ;
Love, CJ ;
Sellers, TD ;
Hayes, DL ;
Reiser, C .
CIRCULATION, 1999, 100 (05) :516-525
[10]   THE DOTTER RETRIEVER AND PIGTAIL CATHETER - EFFICACY IN EXTRACTION OF CHRONIC TRANSVENOUS PACEMAKER LEADS [J].
ESPINOSA, RE ;
HAYES, DL ;
VLIETSTRA, RE ;
OSBORN, MJ ;
MCGOON, MD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (12) :2337-2342