3D ultrasound navigation in syrinx surgery - a feasibility study

被引:17
作者
Bonsanto, MM [1 ]
Metzner, R [1 ]
Aschoff, A [1 ]
Tronnier, V [1 ]
Kunze, S [1 ]
Wirtz, CR [1 ]
机构
[1] Heidelberg Univ, Dept Neurosurg, D-69120 Heidelberg, Germany
关键词
3D ultrasound; navigation; syrinx; spine surgery; intraoperative imaging; sonography;
D O I
10.1007/s00701-005-0505-7
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background. The prototype of a 3D ultrasound navigation system, with a trackable 4- to 8-MHz phased-array ultrasound probe was used in syrinx-surgery. The aim of this study was to evaluate the advantages offered by 3D ultrasound and navigation in syringomyelial surgery. Methods and materials. After laminectomy, with a free-handed tilt of the ultrasound probe, the 3D volume of the target area is acquired within 15 seconds. The data are visualized after reconstruction in an axial, coronal, and sagittal view, offering the possibility of ultrasound-based guided surgery. Results. Based on the intraoperative volume information, it was possible to navigate with the 3D ultrasound images in all cases. The orientation and image quality with respect to resolution, spatial information, and the identification of anatomical structures facilitated the surgery in all seven cases. The navigation tool, with a length of 12 cm and a tip diameter of 1 mm, was simple to place into the surgical site. The availability of an up-to-date 3D-image resulted in less interruption of the surgical procedure, with no need to repeatedly fill the cavity with sterile saline for new ultrasound acquisitions. New ultrasound images were only required if shift occurred. The coronal and "trajectory-plane" views, offer additional information about the syrinx cavity. The target borders are easier to determine and orientation in separated cavities was possible. Particularly in syringomyelial surgery it was helpful to determine the surface point of the syrinx to place the myelotomy or insert a catheter. Conclusion. 3D ultrasound offers the advantages of visualizing the third dimension of the target. For orientation and border determination navigation within the 3D ultrasound volume is very helpful and can take place with the ultrasound probe out of the way. Any disruption in the surgical procedure is minimized by not having to repeatedly fill the cavity with a sterile saline solution, there are fewer difficulties with image orientation because of new image adjustments.
引用
收藏
页码:533 / 541
页数:9
相关论文
共 51 条
  • [1] CLIPPING OF PROXIMAL PARACLINOID ANEURYSMS WITH SUPPORT OF THE BALLOON-CATHETER TRAPPING-EVACUATION TECHNIQUE
    ALBERT, FK
    FORSTING, M
    ASCHOFF, A
    KRIEGER, D
    KUNZE, S
    [J]. ACTA NEUROCHIRURGICA, 1993, 125 (1-4) : 138 - 141
  • [2] INTRAOPERATIVE ULTRASOUND (US) IMAGING - COMPARISON OF PATHOMORPHOLOGICAL FINDINGS IN US AND CT
    AUER, LM
    VANVELTHOVEN, V
    [J]. ACTA NEUROCHIRURGICA, 1990, 104 (3-4) : 84 - 95
  • [3] CHADDUCK WM, 1985, NEUROSURGERY, V16, P477, DOI 10.1227/00006123-198504000-00008
  • [4] Chandler W F, 1983, Clin Neurosurg, V31, P550
  • [5] INTRA-OPERATIVE USE OF REAL-TIME ULTRASONOGRAPHY IN NEUROSURGERY
    CHANDLER, WF
    KNAKE, JE
    MCGILLICUDDY, JE
    LILLEHEI, KO
    SILVER, TM
    [J]. JOURNAL OF NEUROSURGERY, 1982, 57 (02) : 157 - 163
  • [6] Validation of object-induced MR distortion correction for frameless stereotactic neurosurgery
    Dean, D
    Kamath, J
    Duerk, JL
    Ganz, E
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) : 810 - 816
  • [7] DEROSIER C, 1991, Journal de Radiologie (Paris), V72, P349
  • [8] Dohrmann G J, 1988, Clin Neurosurg, V34, P378
  • [9] INTRAOPERATIVE ULTRASOUND IMAGING OF THE SPINAL-CORD - SYRINGOMYELIA, CYSTS, AND TUMORS - A PRELIMINARY-REPORT
    DOHRMANN, GJ
    RUBIN, JM
    [J]. SURGICAL NEUROLOGY, 1982, 18 (06): : 395 - 399
  • [10] Virtual fluoroscopy: Computer-assisted fluoroscopic navigation
    Foley, KT
    Simon, DA
    Rampersaud, YR
    [J]. SPINE, 2001, 26 (04) : 347 - 351