Implantation of a biventricular pacing and defibrillator device via a persistent left superior vena cava

被引:9
作者
Kapetanopoulos, Athanasios [1 ]
Peckham, Gene [1 ]
Kiernan, Francis [1 ]
Clyne, Christopher [1 ]
Kluger, Jeffrey [1 ]
Migeed, Magdy A. [1 ]
机构
[1] Univ Connecticut, Sch Med, Hartford Hosp, Farmington, CT USA
关键词
congenital heart disease; congestive heart failure; defibrillation; pacemaker;
D O I
10.2459/01.JCM.0000228695.14902.4c
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A persistent left superior vena cava was discovered in a 66-year-old man with heart failure undergoing implantation of a biventricular pacing and defibrillator device. An active fixation right ventricular defibrillator lead was placed through a curved guiding catheter. A sub-selection catheter and a guidewire allowed the engagement of a posterior-lateral branch of the coronary sinus, performance of an angiogram without an occlusive balloon, and optimal lead placement. The right atrial lead was positioned using a standard stylet. Despite the technical challenges, implantation of a biventricular pacing and defibrillator device via a persistent left superior vena cava is safe and feasible. J Cardiovasc Med 7:430-433 (C) 2006 Italian Federation of Cardiology.
引用
收藏
页码:430 / 433
页数:4
相关论文
共 19 条
  • [1] Failure of left-sided implantable cardioverter defibrillator implantation due to absence of left brachiocephalic vein
    Antonelli, D
    Freedberg, NA
    Rosenfeld, T
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (12): : 1842 - 1843
  • [2] Implantation of dual chamber pacemaker in a patient with persistent left superior vena cava
    Antonelli, D
    Rosenfeld, T
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (06): : 1737 - 1738
  • [3] Left superior vena cava persistence in patients undergoing pacemaker or cardioverter-defibrillator implantation - A 10-year experience
    Biffi, M
    Boriani, G
    Frabetti, L
    Bronzetti, G
    Branzi, A
    [J]. CHEST, 2001, 120 (01) : 139 - 144
  • [4] CAMPBELL M, 1954, BRIT HEART J, V16, P423
  • [5] IMPLANTATION OF A DUAL CHAMBER PACEMAKER IN A PATIENT WITH PERSISTENT LEFT SUPERIOR VENA-CAVA
    DIRIX, LY
    KERSSCHOT, IE
    FIERENS, H
    GOETHALS, MA
    VANDAELE, G
    CLAESSEN, G
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (03): : 343 - 345
  • [6] ELEVATED DEFIBRILLATION THRESHOLD WHEN RIGHT-SIDED VENOUS ACCESS IS USED FOR NONTHORACOTOMY IMPLANTABLE DEFIBRILLATOR LEAD IMPLANTATION
    EPSTEIN, AE
    KAY, GN
    PLUMB, VJ
    VOSHAGESTAHL, L
    HULL, ML
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1995, 6 (11) : 979 - 986
  • [7] TRANSVENOUS DEFIBRILLATOR IMPLANTATION IN PATIENTS WITH PERSISTENT LEFT SUPERIOR VENA-CAVA AND RIGHT SUPERIOR VENA-CAVA ATRESIA
    FAVALE, S
    BARDY, GH
    PITZALIS, MV
    DICANDIA, CD
    TRAVERSA, M
    RIZZON, P
    [J]. EUROPEAN HEART JOURNAL, 1995, 16 (05) : 704 - 707
  • [8] Comparison of right- and left-sided pectoral implantation parameters with the jewel active can cardiodefibrillator
    Flaker, GC
    Tummala, R
    Wilson, J
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (02): : 447 - 451
  • [9] Biventricular pacing via a persistent left superior vena cava: Report of four cases
    Gasparini, M
    Mantica, M
    Galimberti, P
    Coltorti, F
    Simonini, S
    Ceriotti, C
    Gronda, E
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (01): : 192 - 196
  • [10] Hanna-Moussa Shafik, 2002, J Invasive Cardiol, V14, P192