EBT2 radiochromic film for quality assurance of complex IMRT treatments of the prostate: micro-collimated IMRT, RapidArc, and TomoTherapy

被引:39
作者
Kairn, T. [1 ]
Hardcastle, N. [2 ,3 ]
Kenny, J. [1 ]
Meldrum, R. [1 ]
Tome, W. A. [2 ,3 ]
Aland, T. [1 ]
机构
[1] Wesley Med Ctr, Auchenflower, Qld 4066, Australia
[2] Univ Wisconsin, Dept Human Oncol, Madison, WI 53703 USA
[3] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
关键词
Dosimetry; Radiation therapy; Arc therapy; Quality control; HELICAL TOMOTHERAPY; MULTILEAF COLLIMATOR; ENERGY-DEPENDENCE; RADIOTHERAPY; SYSTEM; VERIFICATION; DOSIMETRY; DELIVERY; MATRIXX; PHOTON;
D O I
10.1007/s13246-011-0087-z
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In response to the clinical need for a dosimetry system with both high resolution and minimal angular dependence, this study demonstrates the utility of Gafchromic EBT2 radiochromic dosimetry film for the quality assurance of micro-collimated IMRT, RapidArc and TomoTherapy treatments. Firstly, preliminary measurements indicated that the dose response of EBT2 film does not appreciably vary with either the angle of incidence of the radiation beam or the depth in water at which the film is placed. Secondly, prostate treatment plans designed for delivery using static-beam IMRT (collimated using the BrainLab m3 microMLC), RapidArc and TomoTherapy were investigated by comparing dose planes obtained from treatment planning calculations with EBT2 film measurements. For all treatment plans, the proportion of dose points agreeing with the film measurements to within c (3%, 3 mm) was found to be above 95%, with all points agreeing within 5%. The film images provided sufficient information to verify that the treatments could be delivered with an acceptable level of accuracy, while also providing additional information on low-level dose variations that were not predicted by the treatment planning systems. This information included: the location and extent of dose from inter-leaf leakage (in the RapidArc plan) and helical field junctioning (in the TomoTherapy plan), as well as the existence of small regions where the treatment planning system under-predicted the dose from very small treatment segments (in the micro-collimated IMRT plan).
引用
收藏
页码:333 / 343
页数:11
相关论文
共 36 条
  • [1] Evaluation of a Gafchromic EBT2 film dosimetry system for radiotherapy quality assurance
    Aland, T.
    Kairn, T.
    Kenny, J.
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2011, 34 (02) : 251 - 260
  • [2] Prostate planning treatment volume margin calculation based on the exactrac X-ray 6D image-guided system:: Margins for various clinical implementations
    Alonso-Arrizabalaga, Sara
    Gonzalez, Luis Brualla
    Ferrando, Juan V. Rosello
    Peidro, Jorge Pastor
    Torrecilla, Jose Lopez
    Meseguer, Domingo Planes
    Hernandez, Trinidad Garcia
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : 936 - 943
  • [3] Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies
    Arjomandy, Bijan
    Tailor, Ramesh
    Anand, Aman
    Sahoo, Narayan
    Gillin, Michael
    Prado, Karl
    Vicic, Milos
    [J]. MEDICAL PHYSICS, 2010, 37 (05) : 1942 - 1947
  • [4] Energy response of the new EBT2 radiochromic film to x-ray radiation
    Butson, Martin J.
    Yu, Peter K. N.
    Cheung, Tsang
    Alnawaf, Hani
    [J]. RADIATION MEASUREMENTS, 2010, 45 (07) : 836 - 839
  • [5] Measurement of high energy x-ray beam penumbra with Gafchromic™ EBT radiochromic film
    Cheung, Tsang
    Butson, Martin J.
    Yu, Peter K. N.
    [J]. MEDICAL PHYSICS, 2006, 33 (08) : 2912 - 2914
  • [6] Fractionated stereotactic radiotherapy: A method to evaluate geometric and dosimetric uncertainties using radiochromic films
    Coscia, Gianluca
    Vaccara, Elena
    Corvisiero, Roberta
    Cavazzani, Paolo
    Ruggieri, Filippo Grillo
    Taccini, Gianni
    [J]. MEDICAL PHYSICS, 2009, 36 (07) : 2870 - 2880
  • [7] THE DEVELOPMENT OF A MONTE CARLO SYSTEM TO VERIFY RADIOTHERAPY TREATMENT DOSE CALCULATIONS
    Crowe, S.
    Kairn, T.
    Fielding, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S71 - S71
  • [8] Small fields:: Nonequilibrium radiation dosimetry
    Das, Indra J.
    Ding, George X.
    Ahnesjo, Anders
    [J]. MEDICAL PHYSICS, 2008, 35 (01) : 206 - 215
  • [9] A study on beams passing through hip prosthesis for pelvic radiation treatment
    Ding, GX
    Yu, CW
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (04): : 1167 - 1175
  • [10] Searching for Optimal Dose-Volume Constraints to Reduce Rectal Toxicity after Hypofractionated Radiotherapy for Prostate Cancer
    Faria, S.
    Joshua, B.
    Patrocinio, H.
    Dal Pra, A.
    Cury, F.
    Velly, A. M.
    Souhami, L.
    [J]. CLINICAL ONCOLOGY, 2010, 22 (10) : 810 - 817