Building an internal model of a myoelectric prosthesis via closed-loop control for consistent and routine grasping

被引:34
作者
Dosen, Strahinja [1 ]
Markovic, Marko [2 ]
Wille, Nicola [2 ,3 ]
Henkel, Markus [2 ,3 ]
Koppe, Mario [2 ,3 ]
Ninu, Andrei [1 ]
Froemmel, Cornelius [4 ]
Farina, Dario [1 ]
机构
[1] Univ Gottingen, Univ Med Ctr Goettingen, Dept Neurorehabil Engn, D-37075 Gottingen, Germany
[2] Otto Bock Healthcare Gmbh, D-37115 Duderstadt, Germany
[3] PFH Private Univ Appl Sci Goettingen, D-37073 Gottingen, Germany
[4] Univ Gottingen, Univ Med Ctr Goettingen, Dept Orthobion, D-37075 Gottingen, Germany
关键词
Upper limb prostheses; Grasping force; Closed-loop control; Internal models; Feedforward; Feedback; MOTOR CONTROL; SENSORY FEEDBACK; GRIP; HAND; FEEDFORWARD; OBJECTS; DESIGN; SYSTEM; LIMB;
D O I
10.1007/s00221-015-4257-1
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Prosthesis users usually agree that myoelectric prostheses should be equipped with somatosensory feedback. However, the exact role of feedback and potential benefits are still elusive. The current study investigates the nature of human control processes within a specific context of routine grasping. Although the latter includes a fast feedforward control of the grasping force, the assumption was that the feedback would still be useful; it would communicate the outcome of the grasping trial, which the subjects could use to learn an internal model of feedforward control. Nine able-bodied subjects produced repeatedly a desired level of grasping force using different control configurations: feedback versus no-feedback, virtual versus real prosthetic hand, and joystick versus myocontrol. The outcome measures were the median and dispersion of the relative force errors. The results demonstrated that the feedback was successful in limiting the variability of the routine grasping due to uncertainties in the system and/or the command interface. The internal models of feedforward control could be employed by the subjects to control the prosthesis without the loss of performance even after the force feedback was removed. The models were, however, unstable over time, especially with myocontrol. Overall, the study demonstrates that the prosthesis system can be learned by the subjects using feedback. The feedback is also essential to maintain the model, and it could be delivered intermittently. This approach has practical advantages, but the level to which this mechanism can be truly exploited in practice depends directly on the consistency of the prosthesis control interface.
引用
收藏
页码:1855 / 1865
页数:11
相关论文
共 32 条
[1]  
Antfolk C, 2013, EXPERT REV MED DEVIC, V10, P45, DOI [10.1586/ERD.12.68, 10.1586/erd.12.68]
[2]   Importance of cutaneous feedback in maintaining a secure grip during manipulation of hand-held objects [J].
Augurelle, AS ;
Smith, AM ;
Lejeune, T ;
Thonnard, JL .
JOURNAL OF NEUROPHYSIOLOGY, 2003, 89 (02) :665-671
[3]   Mechanical design and performance specifications of anthropomorphic prosthetic hands: A review [J].
Belter, Joseph T. ;
Segil, Jacob L. ;
Dollar, Aaron M. ;
Weir, Richard F. .
JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT, 2013, 50 (05) :599-617
[4]  
Chatterjee A., 2008, JPO Journal of Prosthetics and Orthotics, V20, P27, DOI [10.1097/01.JPO.0000311041.61628.be, DOI 10.1097/01.JP0.0000311041.61628.BE]
[5]   On the shared control of an EMG-controlled prosthetic hand: Analysis of user-prosthesis interaction [J].
Cipriani, Christian ;
Zaccone, Franco ;
Micera, Silvestro ;
Carrozza, M. Chiara .
IEEE TRANSACTIONS ON ROBOTICS, 2008, 24 (01) :170-184
[6]   HyVE: Hybrid Vibro-Electrotactile Stimulation for Sensory Feedback and Substitution in Rehabilitation [J].
D'Alonzo, Marco ;
Dosen, Strahinja ;
Cipriani, Christian ;
Farina, Dario .
IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2014, 22 (02) :290-301
[7]   Widespread access to predictive models in the motor system: a short review [J].
Davidson, Paul R. ;
Wolpert, Daniel M. .
JOURNAL OF NEURAL ENGINEERING, 2005, 2 (03) :S313-S319
[8]  
Dosen S., 2014, IEEE T NEURAL SYST R
[9]   Hybrid force-velocity sliding mode control of a prosthetic hand [J].
Engeberg, Erik D. ;
Meek, Sanford G. ;
Minor, Mark A. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (05) :1572-1581
[10]   Improved grasp force sensitivity for prosthetic hands through force-derivative feedback [J].
Engeberg, Erik D. ;
Meek, Sanford .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (02) :817-821