Simplifying Size-Specific Radiation Dose Estimates in Pediatric CT

被引:45
作者
Khawaja, Ranish Deedar Ali [1 ]
Singh, Sarabjeet [1 ]
Vettiyil, Beth [1 ]
Lim, Ruth [2 ]
Gee, Michael [1 ]
Westra, Sjirk [2 ]
Kalra, Mannudeep K. [1 ]
机构
[1] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Dept Radiol, Boston, MA 02114 USA
[2] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Dept Radiol,Div Pediat Imaging, Boston, MA 02114 USA
关键词
body diameter; body weight; pediatric CT; size-specific dose estimate; COMPUTED-TOMOGRAPHY; PROTOCOLS; INDEX;
D O I
10.2214/AJR.13.12191
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. Size-specific dose estimates (SSDEs) require manual measurement of torso diameters for individual patients-anteroposterior (hereafter, D-AP), lateral (hereafter, D-LAT), and effective (hereafter, D-E)-which can be tedious in clinical settings. We aimed to determine whether body weight can be used as a surrogate for measuring diameter in children. MATERIALS AND METHODS. D-AP and D-LAT were measured in 522 consecutive CT examinations (chest, 187 and abdomen-pelvis, 335) performed in 483 (+/- SD) children (13 +/- 7 years). Effective diameter (D-E1) was calculated as the square root of the product of D-AP and D-LAT. A second measurement of effective diameter (D-E2) was obtained using automated software. Correlation coefficients between patient body weight, age, and diameter were measured in addition to 95% prediction interval analysis for diameters corresponding to body weight. RESULTS. Median body weight was 51 kg, and mean D-AP, D-LAT, D-E1, and D-E2 were 207.1 +/- 50.8 mm, 289.8 +/- 72.6 mm, 243.3 +/- 62.0 mm, and 233.6 +/- 55.4 mm, respectively. Overall body weight had a strong correlation with diameter (0.88, 0.85, 0.86, and 0.93 respectively; all p < 0.0001). SSDE measured using body weight was statistically not different than SSDE measured using effective diameters (p = 0.9). Children weighing less than 27 kg and between 46 and 100 kg had statistically significant correlations with torso diameters, whereas only anteroposterior and effective diameters were correlated with children weighing between 27 and 45 kg. Children less than 4 years old had strong correlation with all diameters. Adolescents (15-18 years) did not have statistically significant correlation with any of the diameters. CONCLUSION. Body weight, instead of body diameter, can be used as a surrogate to estimate size-specific dose in children, making dose estimation clinically simpler and more rapid.
引用
收藏
页码:167 / 176
页数:10
相关论文
共 10 条
[1]  
American Association of Physicists in Medicine website, 2011, SIZ SPEC DOS EST SSD
[2]   Investigation of American Association of Physicists in Medicine Report 204 Size-specific Dose Estimates for Pediatric CT Implementation [J].
Brady, Samuel L. ;
Kaufman, Robert A. .
RADIOLOGY, 2012, 265 (03) :832-840
[3]   It is time to retire the computed tomography dose index (CTDI) for CT quality assurance and dose optimization [J].
Brenner, DJ ;
McCollough, CH ;
Orton, CG .
MEDICAL PHYSICS, 2006, 33 (05) :1189-1191
[4]   Size-specific Dose Estimates for Adult Patients at CT of the Torso [J].
Christner, Jodie A. ;
Braun, Natalie N. ;
Jacobsen, Megan C. ;
Carter, Rickey E. ;
Kofler, James M. ;
McCollough, Cynthia H. .
RADIOLOGY, 2012, 265 (03) :841-847
[5]   Size-appropriate radiation doses in pediatric body CT: a study of regional community adoption in the United States [J].
Hopkins, Katharine L. ;
Pettersson, David R. ;
Koudelka, Caroline W. ;
Spinning, Kristopher ;
Vajtai, Petra L. ;
Beckett, Brooke R. ;
Bardo, Dianna M. E. .
PEDIATRIC RADIOLOGY, 2013, 43 (09) :1128-1135
[6]   Effective dose determination using an anthropomorphic phantom and metal oxide semiconductor field effect transistor technology for clinical adult body multidetector array computed tomography protocols [J].
Hurwitz, Lynne M. ;
Yoshizumi, Terry T. ;
Goodman, Philip C. ;
Frush, Donald P. ;
Nguyen, Giao ;
Toncheva, Greta ;
Lowry, Carolyn .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2007, 31 (04) :544-549
[7]   CT Dose Index and Patient Dose: They Are Not the Same Thing [J].
McCollough, Cynthia H. ;
Leng, Shuai ;
Yu, Lifeng ;
Cody, Dianna D. ;
Boone, John M. ;
McNitt-Gray, Michael F. .
RADIOLOGY, 2011, 259 (02) :311-316
[8]   A METHOD FOR DESCRIBING THE DOSES DELIVERED BY TRANSMISSION-X-RAY COMPUTED-TOMOGRAPHY [J].
SHOPE, TB ;
GAGNE, RM ;
JOHNSON, GC .
MEDICAL PHYSICS, 1981, 8 (04) :488-495
[9]   Dose Reduction and Compliance with Pediatric CT Protocols Adapted to Patient Size, Clinical Indication, and Number of Prior Studies [J].
Singh, Sarabjeet ;
Kalra, Mannudeep K. ;
Moore, Michael A. ;
Shailam, Randheer ;
Liu, Bob ;
Toth, Thomas L. ;
Grant, Ellen ;
Westra, Sjirk J. .
RADIOLOGY, 2009, 252 (01) :200-208
[10]   Image Gently Vendor Summit: Working Together for Better Estimates of Pediatric Radiation Dose from CT [J].
Strauss, Keith J. ;
Goske, Marilyn J. ;
Frush, Donald P. ;
Butler, Priscilla F. ;
Morrison, Gregory .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 192 (05) :1169-1175