Simplifying Size-Specific Radiation Dose Estimates in Pediatric CT

被引:43
|
作者
Khawaja, Ranish Deedar Ali [1 ]
Singh, Sarabjeet [1 ]
Vettiyil, Beth [1 ]
Lim, Ruth [2 ]
Gee, Michael [1 ]
Westra, Sjirk [2 ]
Kalra, Mannudeep K. [1 ]
机构
[1] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Dept Radiol, Boston, MA 02114 USA
[2] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Dept Radiol,Div Pediat Imaging, Boston, MA 02114 USA
关键词
body diameter; body weight; pediatric CT; size-specific dose estimate; COMPUTED-TOMOGRAPHY; PROTOCOLS; INDEX;
D O I
10.2214/AJR.13.12191
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. Size-specific dose estimates (SSDEs) require manual measurement of torso diameters for individual patients-anteroposterior (hereafter, D-AP), lateral (hereafter, D-LAT), and effective (hereafter, D-E)-which can be tedious in clinical settings. We aimed to determine whether body weight can be used as a surrogate for measuring diameter in children. MATERIALS AND METHODS. D-AP and D-LAT were measured in 522 consecutive CT examinations (chest, 187 and abdomen-pelvis, 335) performed in 483 (+/- SD) children (13 +/- 7 years). Effective diameter (D-E1) was calculated as the square root of the product of D-AP and D-LAT. A second measurement of effective diameter (D-E2) was obtained using automated software. Correlation coefficients between patient body weight, age, and diameter were measured in addition to 95% prediction interval analysis for diameters corresponding to body weight. RESULTS. Median body weight was 51 kg, and mean D-AP, D-LAT, D-E1, and D-E2 were 207.1 +/- 50.8 mm, 289.8 +/- 72.6 mm, 243.3 +/- 62.0 mm, and 233.6 +/- 55.4 mm, respectively. Overall body weight had a strong correlation with diameter (0.88, 0.85, 0.86, and 0.93 respectively; all p < 0.0001). SSDE measured using body weight was statistically not different than SSDE measured using effective diameters (p = 0.9). Children weighing less than 27 kg and between 46 and 100 kg had statistically significant correlations with torso diameters, whereas only anteroposterior and effective diameters were correlated with children weighing between 27 and 45 kg. Children less than 4 years old had strong correlation with all diameters. Adolescents (15-18 years) did not have statistically significant correlation with any of the diameters. CONCLUSION. Body weight, instead of body diameter, can be used as a surrogate to estimate size-specific dose in children, making dose estimation clinically simpler and more rapid.
引用
收藏
页码:167 / 176
页数:10
相关论文
共 50 条
  • [1] Advancing size-specific dose estimates in CT examinations: Dose estimates at longitudinal positions of scans
    Li, Xinhua
    Marschall, Theodore A.
    Yang, Kai
    Liu, Bob
    MEDICAL PHYSICS, 2022, 49 (02) : 1303 - 1311
  • [2] Size-specific dose estimates in pediatric brain CT in relation to age and weight
    Fujii, Kaoru
    Inoue, Yusuke
    Itoh, Hiroyasu
    Sasa, Ryosuke
    Hata, Hirofumi
    Mitsui, Kohei
    RADIATION PROTECTION DOSIMETRY, 2023, 199 (10) : 1075 - 1081
  • [3] RADIATION DOSE DETERMINATION IN ABDOMINAL CT EXAMINATIONS OF CHILDREN AT SUDANESE HOSPITALS USING SIZE-SPECIFIC DOSE ESTIMATES
    Bashier, Einas H.
    Suliman, I. I.
    RADIATION PROTECTION DOSIMETRY, 2019, 183 (04) : 444 - 449
  • [4] Optimizing CT radiation dose based on patient size and image quality: the size-specific dose estimate method
    Larson, David B.
    PEDIATRIC RADIOLOGY, 2014, 44 : 501 - 505
  • [5] Size-specific Dose Estimates for Adult Patients at CT of the Torso
    Christner, Jodie A.
    Braun, Natalie N.
    Jacobsen, Megan C.
    Carter, Rickey E.
    Kofler, James M.
    McCollough, Cynthia H.
    RADIOLOGY, 2012, 265 (03) : 841 - 847
  • [6] Consideration of the usefulness of a size-specific dose estimate in pediatric CT examination
    Tsujiguchi, Takakiyo
    Obara, Hideki
    Ono, Shuichi
    Saito, Yoko
    Kashiwakura, Ikuo
    JOURNAL OF RADIATION RESEARCH, 2018, 59 (04) : 430 - 435
  • [7] Size-Specific Dose Estimates for Evaluation of Individual Patient Dose in CT Protocol for Renal Colic
    Waszczuk, Lukasz A.
    Guzinski, Maciej
    Czarnecka, Anna
    Sasiadek, Marek J.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 205 (01) : 100 - 105
  • [8] How accurate is size-specific dose estimate in pediatric body CT examinations?
    Karmazyn, Boaz
    Ai, Huisi
    Klahr, Paul
    Ouyang, Fangqian
    Jennings, S. Gregory
    PEDIATRIC RADIOLOGY, 2016, 46 (09) : 1234 - 1240
  • [9] Radiation dose in chest CT: Assessment of size-specific dose estimates based on water-equivalent correction
    Gabusi, M.
    Riccardi, L.
    Aliberti, C.
    Vio, S.
    Paiusco, M.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (02): : 393 - 397
  • [10] Optimizing CT radiation dose based on patient size and image quality: the size-specific dose estimate method
    David B. Larson