PURPOSE: The accuracy of intraocular lens (IOL) calculation is limited by the variation of the measurement of corneal power, axial length, anterior chamber depth, and effective postoperative lens position. Therefore, IOL predictability is limited. The aim of this study was to evaluate the performance of the novel light adjustable lens (LAL, Calhoun Vision). METHODS: Ten eyes with cataracts (seven patients) were included. After standard phacoemulsification, the LAL was implanted. Two to 3 weeks after surgery, all IOL adjustments had been completed. Preoperatively, pre-adjustment, pre-lock-in, and 1 month post-lock-in, a full ophthalmic examination including objective (autorefraction) and subjective refraction (ie, manifest uncorrected visual acuity [UCVA] and manifest best spectacle-corrected visual acuity [BSCVA]) and topography were performed. RESULTS: All cataract surgeries were performed with no complications. Two weeks after surgery, the mean refraction was +0.78 +/- 0.79 diopters (D). After adjustment and lock-in, the refraction was -0.07 +/- 0.21 D. In 4 of 10 eyes, second adjustments were necessary. One month after surgery, all patients were within +/- 0.50 D of the intended refraction. Preoperatively, the cylinder was 0.88 +/- 0.77 D, but was reduced after adjustments to 0.15 +/- 0.20 D. Mean UCVA was 1.02, and all eyes gained at least two lines on the Snellen visual acuity test (ie, BSCVA) after surgery. CONCLUSIONS: The LAL is a new IOL with the ability to correct up to 2.00 D of sphere and cylinder after implantation. Our first clinical results are promising and indicate effective safety and stability. U Refract Surg. 2010;26:314-3201 doi:10.3928/1081597X-20090617-12
引用
收藏
页码:314 / 320
页数:7
相关论文
共 50 条
[41]   Small-aperture intraocular lens combined with secondary piggyback intraocular lens during cataract surgery after previous radial keratotomy [J].
Barnett, Victoria ;
Barsam, Allon ;
Than, Jonathan ;
Srinivasan, Sathish .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2018, 44 (08) :1042-1045
[42]   Clinical outcomes of cataract surgery after bag-in-the-lens intraocular lens implantation following ISO standard 11979-7:2006 [J].
Tassignon, Marie-Jose ;
Gobin, Laure ;
Mathysen, Danny ;
Van Looveren, Jan ;
De Groot, Veva .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (12) :2120-2129
[43]   Symbolic algebra approach to the calculation of intraocular lens power following cataract surgery [J].
Hjelmstad, David ;
Sayegh, Samir I. .
OPHTHALMIC TECHNOLOGIES XXIII, 2013, 8567
[44]   Secondary glaucoma and visual outcome after paediatric cataract surgery with primary bag-in-the-lens intraocular lens [J].
Nystrom, Alf ;
Magnusson, Gunilla ;
Zetterberg, Madeleine .
ACTA OPHTHALMOLOGICA, 2020, 98 (03) :296-304
[45]   Effects of Residual Anterior Lens Epithelial Cell Removal on Axial Position of Intraocular Lens after Cataract Surgery [J].
Bang, Seung Pil ;
Yoo, Young-Sik ;
Jun, Jong Hwa ;
Joo, Choun-Ki .
JOURNAL OF OPHTHALMOLOGY, 2018, 2018
[46]   Visual results after implantation of a trifocal intraocular lens in high myopes [J].
Steinwender, Gernot ;
Schwarz, Luisa ;
Boehm, Myriam ;
Slavik-Lencova, Anna ;
Hemkeppler, Eva ;
Shajari, Mehdi ;
Kohnen, Thomas .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2018, 44 (06) :680-685
[47]   Optical quality of toric intraocular lens implantation in cataract surgery [J].
Xiao, Xian-Wen ;
Hao, Jing ;
Zhang, Hong ;
Tian, Fang .
INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2015, 8 (01) :66-71
[48]   Corneal Topography for Intraocular Lens Selection in Refractive Cataract Surgery [J].
Goto, So ;
Maeda, Naoyuki .
OPHTHALMOLOGY, 2021, 128 (11) :E142-E152
[49]   TORIC INTRAOCULAR LENS IMPLANTATION FOR ASTIGMATISM CORRECTION IN CATARACT SURGERY [J].
Vickovic, Ivanka Petric ;
Loncar, Valentina Lacmanovic ;
Mandic, Zdravko ;
Ivekovic, Renata ;
Herman, Jelena Skunca ;
Sesar, Antonio .
ACTA CLINICA CROATICA, 2012, 51 (02) :293-297
[50]   Accuracy of intraocular lens power calculation for cataract surgery after deep anterior lamellar keratoplasty [J].
Pellegrini, Marco ;
Furiosi, Luca ;
Salgari, Niccolo ;
D'Angelo, Sergio ;
Zauli, Giorgio ;
Yu, Angeli Christy ;
Busin, Massimo .
CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2022, 50 (01) :17-22