Factors affecting the accuracy of non-invasive blood glucose measurement by short-wavelength near infrared spectroscopy in the determination of the glycaemic index of foods

被引:17
作者
Uwadaira, Yasuhiro [1 ,2 ]
Adachi, Norihiko [2 ]
Ikehata, Akifumi [3 ]
Kawano, Sumio [1 ,3 ]
机构
[1] Univ Tsukuba, Grad Sch Life & Environm Sci, Tsukuba, Ibaraki 3058572, Japan
[2] WACCORD Co, Tsukuba, Ibaraki 3050033, Japan
[3] Natl Food Res Inst, Tsukuba, Ibaraki 3058642, Japan
关键词
blood glucose; non-invasive analysis; glycaemic index; near infrared spectroscopy; interactance; hand-palm; fibre-optic probe; HUMAN SERUM-ALBUMIN; CALIBRATION MODELS; GAMMA-GLOBULIN; SUGAR CONTENT; WHOLE-BLOOD; COMPONENTS; MATRIX;
D O I
10.1255/jnirs.895
中图分类号
O69 [应用化学];
学科分类号
081704 ;
摘要
Non-invasive techniques for the analysis of blood glucose are desirable not only for clinical use but also for the determination of the glycaemic index (GI) of foods. Factors that affect the accuracy of non-invasive blood glucose measurements by near infrared (NIR) spectroscopy were identified and examined to establish the optimal conditions for the analysis. Calibration models were obtained by means of partial least squares regression. Several experiments revealed that a good calibration model could be obtained using spectra (7001100 nm) acquired from the palm of the hand with a large interactance probe fixed to a table provided that spectral acquisition and blood glucose analysis are performed simultaneously. The accuracy of the calibration model developed is clinically acceptable; the coefficient of multiple determination was 0.85 and the standard error of cross-validation was 9.7 mg dL(-1). An analysis of a model solution confirmed that NIR spectroscopy in the short-wavelength region 700-1100 nm is capable of measuring glucose concentrations of the order of 0.01% (10 mg dL(-1)) in blood. The determination of the GI of foods by using the present method will be examined in the near future.
引用
收藏
页码:291 / 300
页数:10
相关论文
共 30 条
[1]   Scattering and absorption effects in the determination of glucose in whole blood by near-infrared spectroscopy [J].
Amerov, AK ;
Chen, J ;
Small, GW ;
Arnold, MA .
ANALYTICAL CHEMISTRY, 2005, 77 (14) :4587-4594
[2]   Phantom glucose calibration models from simulated noninvasive human near-infrared spectra [J].
Arnold, MA ;
Burmeister, JJ ;
Small, GW .
ANALYTICAL CHEMISTRY, 1998, 70 (09) :1773-1781
[3]   DETERMINATION OF PHYSIOLOGICAL LEVELS OF GLUCOSE IN AN AQUEOUS MATRIX WITH DIGITALLY FILTERED FOURIER-TRANSFORM NEAR-INFRARED SPECTRA [J].
ARNOLD, MA ;
SMALL, GW .
ANALYTICAL CHEMISTRY, 1990, 62 (14) :1457-1464
[4]  
BIRTH GS, 1985, J AM SOC HORTIC SCI, V110, P297
[5]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[6]   INFLUENTIAL OBSERVATIONS IN LINEAR-REGRESSION [J].
COOK, RD .
JOURNAL OF THE AMERICAN STATISTICAL ASSOCIATION, 1979, 74 (365) :169-174
[7]   Current development in non-invasive glucose monitoring [J].
do Amaral, Carlos Eduardo Ferrante ;
Wolf, Benhard .
MEDICAL ENGINEERING & PHYSICS, 2008, 30 (05) :541-549
[8]   Enhancing calibration models for non-invasive near-infrared spectroscopical blood glucose determination [J].
Fischbacher, C ;
Jagemann, KU ;
Danzer, K ;
Muller, UA ;
Papenkordt, L ;
Schuler, J .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1997, 359 (01) :78-82
[9]  
Food and Agriculture Organization, 1998, JOINT FAO WHO EXP CO
[10]   Short-wavelength near-infrared spectra of sucrose, glucose, and fructose with respect to sugar concentration and temperature [J].
Golic, M ;
Walsh, K ;
Lawson, P .
APPLIED SPECTROSCOPY, 2003, 57 (02) :139-145