Quantitative image quality evaluation for Cardiac CT reconstructions

被引:0
作者
Tseng, Hsin-Wu [1 ,2 ]
Fan, Jiahua [2 ]
Kupinski, Matthew A. [1 ]
Balhorn, William [2 ]
Okerlund, Darin R. [2 ]
机构
[1] Univ Arizona, Coll Opt Sci, Tucson, AZ 85721 USA
[2] GE Healthcare, CT Engn, Waukesha, WI 53188 USA
来源
MEDICAL IMAGING 2016: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT | 2016年 / 9787卷
关键词
Cardiac CT; CT; Motion Compensation; SnapShot Freeze (SSF); Image Quality;
D O I
10.1117/12.2208341
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Maintaining image quality in the presence of motion is always desirable and challenging in clinical Cardiac CT imaging. Different image-reconstruction algorithms are available on current commercial CT systems that attempt to achieve this goal. It is widely accepted that image-quality assessment should be task-based and involve specific tasks, observers, and associated figures of merits. In this work, we developed an observer model that performed the task of estimating the percentage of plaque in a vessel from CT images. We compared task performance of Cardiac CT image data reconstructed using a conventional FBP reconstruction algorithm and the SnapShot Freeze (SSF) algorithm, each at default and optimal reconstruction cardiac phases. The purpose of this work is to design an approach for quantitative image-quality evaluation of temporal resolution for Cardiac CT systems. To simulate heart motion, a moving coronary type phantom synchronized with an ECG signal was used. Three different percentage plaques embedded in a 3 mm vessel phantom were imaged multiple times under motion free, 60 bpm, and 80 bpm heart rates. Static (motion free) images of this phantom were taken as reference images for image template generation. Independent ROIs from the 60 bpm and 80 bpm images were generated by vessel tracking. The observer performed estimation tasks using these ROIs. Ensemble mean square error (EMSE) was used as the figure of merit. Results suggest that the quality of SSF images is superior to the quality of FBP images in higher heart-rate scans.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Dose performance and image quality: Dual source CT versus single source CT in cardiac CT angiography
    Wang, Min
    Qi, Heng-Tao
    Wang, Xi-Ming
    Wang, Tao
    Chen, Jiu-Hong
    Liu, Cheng
    EUROPEAN JOURNAL OF RADIOLOGY, 2009, 72 (03) : 396 - 400
  • [22] Quantitative CT study of martial arts sports injuries based on image quality
    Ai Zexiu
    JOURNAL OF VISUAL COMMUNICATION AND IMAGE REPRESENTATION, 2019, 60 : 417 - 425
  • [23] Motion compensated reconstructions of calcified coronary plaques in cardiac CT
    King, Martin
    Pan, Xiaochuan
    Giger, Maryellen
    Suzuki, Kenji
    MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [24] Deep learning-based and hybrid-type iterative reconstructions for CT: comparison of capability for quantitative and qualitative image quality improvements and small vessel evaluation at dynamic CE-abdominal CT with ultra-high and standard resolutions
    Matsukiyo, Ryo
    Ohno, Yoshiharu
    Matsuyama, Takahiro
    Nagata, Hiroyuki
    Kimata, Hirona
    Ito, Yuya
    Ogawa, Yukihiro
    Murayama, Kazuhiro
    Kato, Ryoichi
    Toyama, Hiroshi
    JAPANESE JOURNAL OF RADIOLOGY, 2021, 39 (02) : 186 - 197
  • [25] IMAGE QUALITY IN DYNAMIC CT - A CLINICAL DISCUSSION
    FOLEY, WD
    RADIOGRAPHICS, 1993, 13 (01) : 225 - 233
  • [26] Deep learning-based and hybrid-type iterative reconstructions for CT: comparison of capability for quantitative and qualitative image quality improvements and small vessel evaluation at dynamic CE-abdominal CT with ultra-high and standard resolutions
    Ryo Matsukiyo
    Yoshiharu Ohno
    Takahiro Matsuyama
    Hiroyuki Nagata
    Hirona Kimata
    Yuya Ito
    Yukihiro Ogawa
    Kazuhiro Murayama
    Ryoichi Kato
    Hiroshi Toyama
    Japanese Journal of Radiology, 2021, 39 : 186 - 197
  • [27] Impact of Increasing Levels of Advanced Iterative Reconstruction on Image Quality in Low-Dose Cardiac CT Angiography
    Kroepil, P.
    Bigdeli, A. H.
    Nagel, H. D.
    Antoch, G.
    Cohnen, M.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2014, 186 (06): : 567 - 575
  • [28] Influence of reconstruction techniques on PET/CT image quality and quantitative accuracy: a phantom study
    Su, Xuesong
    Geng, Jianhua
    Liu, Jianing
    Liu, Fengshuo
    Wu, Yichen
    Zheng, Rong
    Wang, Xuejuan
    RADIATION DETECTION TECHNOLOGY AND METHODS, 2024, 8 (02) : 1171 - 1186
  • [29] Impact of patient body habitus on image quality and quantitative value in bone SPECT/CT
    Fukai, Shohei
    Daisaki, Hiromitsu
    Umeda, Takuro
    Shimada, Naoki
    Miyaji, Noriaki
    Ito, Ryoma
    Takiguchi, Tomohiro
    Terauchi, Takashi
    ANNALS OF NUCLEAR MEDICINE, 2022, 36 (06) : 586 - 595
  • [30] CT image quality over time: comparison of image quality for six different CT scanners over a six-year period
    Roa, Ana Maria A.
    Andersen, Hilde K.
    Martinsen, Anne Catrine T.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (02): : 350 - 365