Phased-array intracardiac Echocardiography monitoring during pulmonary vein isolation in patients with atrial fibrillation - Impact on outcome and complications

被引:375
作者
Marrouche, NF [1 ]
Martin, DO [1 ]
Wazni, O [1 ]
Gillinov, AM [1 ]
Klein, A [1 ]
Bhargava, M [1 ]
Saad, E [1 ]
Bash, D [1 ]
Yamada, H [1 ]
Jaber, W [1 ]
Schweikert, R [1 ]
Tchou, P [1 ]
Abdul-Karim, A [1 ]
Saliba, W [1 ]
Natale, A [1 ]
机构
[1] Cleveland Clin Fdn, Dept Cardiovasc Med, Ctr Atrial Fibrillat, Cleveland, OH 44195 USA
关键词
fibrillation; atrial; ablation; echocardiography;
D O I
10.1161/01.CIR.0000070541.83326.15
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background - The objective of this study was to assess the impact of intracardiac echocardiography ( ICE) on the long-term success and complications in patients undergoing pulmonary vein isolation ( PVI) for treatment of atrial fibrillation (AF). Methods and Results - Three hundred fifteen patients underwent PVI for treatment of AF. Each patient underwent ostial isolation of all PVs using a cooled-tip ablation catheter. PVI was performed using circular mapping ( CM) alone ( group 1, 56 patients), CM and ICE ( group 2, 107 patients), and CM and ICE with titration of radiofrequency energy based on visualization of microbubbles by ICE ( group 3, 152 patients). After a mean follow-up time of 417 +/- 145 days, 19.6% (11 of 56), 16.8% (18 of 107), and 9.8% (15 of 152) of patients in groups 1, 2, and 3 experienced recurrence of AF, respectively. Moreover, whereas no group 3 patient experienced severe (> 70%) PV stenosis, severe PV stenosis was documented in 3 (3.5%) of 56 patients in group 1 and in 2 (1.8%) of 107 patients in group 2 ( P < 0.05). No embolic events were detected in group 3 patients. Conclusions - Intracardiac echocardiography improves the outcome of cooled-tip PVI. Power adjustment guided by direct visualization of microbubble formation reduces the risk of PV stenosis and improves long-term cure.
引用
收藏
页码:2710 / 2716
页数:7
相关论文
共 17 条
[1]  
[Anonymous], 2001, CIRCULATION, V104, P2118
[2]   Initiation of atrial fibrillation by ectopic beats originating from the pulmonary veins - Electrophysiological characteristics, pharmacological responses, and effects of radiofrequency ablation [J].
Chen, SA ;
Hsieh, MH ;
Tai, CT ;
Tsai, CF ;
Prakash, VS ;
Yu, WC ;
Hsu, TL ;
Ding, YA ;
Chang, MS .
CIRCULATION, 1999, 100 (18) :1879-1886
[3]   INTRACARDIAC ECHOCARDIOGRAPHY DURING RADIOFREQUENCY CATHETER ABLATION OF CARDIAC-ARRHYTHMIAS IN HUMANS [J].
CHU, E ;
KALMAN, JM ;
KWASMAN, MA ;
JUE, JCY ;
FITZGERALD, PJ ;
EPSTEIN, LM ;
SCHILLER, NB ;
YOCK, PG ;
LESH, MD .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (05) :1351-1357
[4]   Catheter tip orientation affects radiofrequency ablation lesion size in the canine left ventricle [J].
Chugh, SS ;
Chan, RC ;
Johnson, SB ;
Packer, DL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (03) :413-420
[5]   Comparative study of fluoroscopy and intracardiac echocardiographic guidance for the creation of linear atrial lesions [J].
Epstein, LM ;
Mitchell, MA ;
Smith, TW ;
Haines, DE .
CIRCULATION, 1998, 98 (17) :1796-1801
[6]   TISSUE HEATING DURING RADIOFREQUENCY CATHETER ABLATION - A THERMODYNAMIC MODEL AND OBSERVATIONS IN ISOLATED PERFUSED AND SUPERFUSED CANINE RIGHT VENTRICULAR FREE WALL [J].
HAINES, DE ;
WATSON, DD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (06) :962-976
[7]   Mapping-guided ablation of pulmonary veins to cure atrial fibrillation [J].
Haïssaguerre, M ;
Shah, DC ;
Jaïs, P ;
Hocini, M ;
Yamane, T ;
Deisenhofer, I ;
Garrigue, S ;
Clémenty, J .
AMERICAN JOURNAL OF CARDIOLOGY, 2000, 86 (9A) :9K-19K
[8]   Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins [J].
Haïssaguerre, M ;
Jaïs, P ;
Shah, DC ;
Takahashi, A ;
Hocini, M ;
Quiniou, G ;
Garrigue, S ;
Le Mouroux, A ;
Le Métayer, P ;
Clémenty, J .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 339 (10) :659-666
[9]   Electrophysiological breakthroughs from the left atrium to the pulmonary veins [J].
Haïssaguerre, M ;
Shah, DC ;
Jaïs, P ;
Hocini, M ;
Yamane, T ;
Deisenhofer, I ;
Chauvin, M ;
Garrigue, S ;
Clémenty, J .
CIRCULATION, 2000, 102 (20) :2463-2465
[10]   Biophysical characteristics of radiofrequency lesion formation in vivo: Dynamics of catheter tip-tissue contact evaluated by intracardiac echocardiography [J].
Kalman, JM ;
Fitzpatick, AP ;
Olgin, JE ;
Chin, MC ;
Lee, RJ ;
Scheinman, MM ;
Lesh, MD .
AMERICAN HEART JOURNAL, 1997, 133 (01) :8-18