Density measurements of single granules using the atomic force microscope

被引:3
作者
Andersson, KM [1 ]
Bergström, L [1 ]
机构
[1] Inst Surface Chem, YKI, SE-11486 Stockholm, Sweden
关键词
D O I
10.1111/j.1551-2916.2005.00419.x
中图分类号
TQ174 [陶瓷工业]; TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
The density of single spray-dried granules has been determined with a new method based on atomic force microscopy (AFM). Spherical granules with a well-defined diameter are attached to the AFM cantilever, which acts as a beam-type spring, and the mass of a granule is estimated from the shift in the resonant frequency. The error of the measurements associated with the method was estimated to vary between 1% and 5%, depending on the size and shape of the granule. Density measurements of spray-dried WC-Co granules are presented, and the effect of a polymeric binder and dispersant on the consolidation during drying is discussed.
引用
收藏
页码:2322 / 2324
页数:3
相关论文
共 10 条