X-ray micro-tomography of a coronary stent deployed in a model artery

被引:14
|
作者
Connolley, Thomas [1 ]
Nash, Derek [1 ]
Buffiere, Jean-Yves [2 ]
Sharif, Faisal [1 ]
McHugh, Peter Edward [1 ,3 ]
机构
[1] Natl Univ Ireland, Natl Ctr Biomed Eng Sci, Galway, Ireland
[2] Inst Natl Sci Appl, GEMPPM, F-69621 Villeurbanne, France
[3] Natl Univ Ireland, Dept Mech & Biomed Engn, Galway, Ireland
关键词
tomographic reconstruction; phase contrast; synchrotron source; stent deployment;
D O I
10.1016/j.medengphy.2006.10.016
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Coronary stents are small scaffolds routinely implanted during angioplasty procedures to re-open coronary arteries which have become narrowed by an atherosclerotic plaque. Despite the advent of drug-eluting stents to reduce restenosis (re-narrowing) after the procedure, complications still arise and mechanical factors may be partly the cause. Stents are being used in increasingly complex cases, so new forms of pre-clinical testing may be helpful in evaluating stent designs and deployment techniques. With this in mind, an in vitro experiment was conducted to evaluate the use of X-ray micro-tomography to image stents at various stages of deployment. A stent was deployed in an artificial artery while using a synchrotron X-ray source to obtain the tomography scans. A volume element (voxel) size of 5.3 mu m was achieved, with a vertical field of view of 4.5 mm. The imaging of the stent and artificial artery materials was better than expected, despite some attenuation artifacts and the high monochromatic beam energy used (25 keV). Experimental problems are discussed, together with recommendations for improving the technique. It is intended that this technique will be of interest to engineers and clinicians as a pre-clinical test. (C) 2006 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1132 / 1141
页数:10
相关论文
共 50 条
  • [21] Micro-Tomography of Coherent X-ray Scatter Using an X-ray Collimator and Spectral Imaging Array
    Eaker, Diane R.
    Jorgensen, Steven M.
    Cui, Congwu
    Ritman, Erik L.
    TOMOGRAPHY AND INVERSE TRANSPORT THEORY, 2011, 559 : 171 - 180
  • [22] Optimisation of X-ray micro-tomography for the in situ study of the development of plant roots
    Jenneson, PM
    Gilboy, WB
    Morton, EJ
    Luggar, RD
    Gregory, PJ
    Hutchison, D
    1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1999, : 429 - 432
  • [23] Statistical iterative reconstruction for multi-contrast x-ray micro-tomography
    Allner, S.
    Velroyen, A.
    Fehringer, A.
    Pfeiffer, F.
    Noel, P. B.
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [24] Overcoming the fragility - X-ray computed micro-tomography elucidates brachiopod endoskeletons
    Seidel, Ronald
    Lueter, Carsten
    FRONTIERS IN ZOOLOGY, 2014, 11
  • [25] Characterisation of fretting fatigue damage using synchrotron X-ray micro-tomography
    Proudhon, H.
    Buffire, J. -Y.
    Fouvry, S.
    TRIBOLOGY INTERNATIONAL, 2006, 39 (10) : 1106 - 1113
  • [26] Characterization of Oil Nano-Structures with Monochromatic X-ray Micro-Tomography
    Asadchikov, Victor E.
    Buzmakov, Alexey V.
    Osadchaya, Anna S.
    Zolotov, Denis A.
    Rafailov, Michael K.
    BIOSENSING AND NANOMEDICINE V, 2012, 8460
  • [27] Moldavite porosity: a 3-D X-ray micro-tomography study
    Rantzsch, Ulrike
    Franz, Alexandra
    Kloess, Gert
    EUROPEAN JOURNAL OF MINERALOGY, 2013, 25 (04) : 705 - 710
  • [28] Quantifying the morphology of crushed sand particles using X-ray micro-tomography
    Wu, Mengmeng
    Lu, Jiatai
    Li, Xintong
    Pan, Sicheng
    Wang, Jianfeng
    Yin, Zhenyu
    GRANULAR MATTER, 2023, 25 (04)
  • [29] Quantifying the morphology of crushed sand particles using X-ray micro-tomography
    Mengmeng Wu
    Jiatai Lu
    Xintong Li
    Sicheng Pan
    Jianfeng Wang
    Zhenyu Yin
    Granular Matter, 2023, 25
  • [30] Micromechanisms of inelastic deformation in sandstones: an insight using x-ray micro-tomography
    Fonseca, J.
    Besuelle, P.
    Viggiani, G.
    GEOTECHNIQUE LETTERS, 2013, 3 : 78 - 83