Portable I2-stabilized Nd:YAG laser for international comparisons

被引:41
作者
Hong, FL [1 ]
Ishikawa, J
Bi, ZY
Zhang, J
Seta, K
Onae, A
Yoda, J
Matsumoto, H
机构
[1] AIST, Natl Res Lab Metrol, Tsukuba, Ibaraki 3058563, Japan
[2] E China Normal Univ, Lab Quantum Opt, Shanghai 200062, Peoples R China
[3] Shanxi Univ, Inst Optoelect, Taiyuan, Peoples R China
关键词
international comparison; laser frequency stabilization; molecular iodine; Nd : YAG laser; optical frequency standard;
D O I
10.1109/19.918172
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have established a compact and transportable I-2-stabilized Nd:YAG laser for international comparisons of laser frequency. The root Allan variance of the portable laser has reached 3.9 x 10(-14) when the integration time is longer than 200 s, The results of an international comparison between the National Research Laboratory of Metrology (NRLM), Tsukuba, Japan and the JILA (formerly the Joint Institute for Laboratory Astrophysics), Boulder, CO, USA show that the frequency difference of the portable laser NRLM-Y1 and the JILA laser JILA-W (f(NRLM-Y1) - f(JILA-W)) was -2.5 kHz, when the cold-finger temperatures of NRLM-Y1 and JILA-W were kept at -10 degreesC and -15 degreesC, respectively. The averaged frequency offset between two NRLM lasers (f(NRLM-Y1) - f(NRLM-Y2)) was -1.1 kHz, A frequency variation of about 1.2 kHz was found for the frequency offset between two NRLM lasers, after NRLM-Y1 was taken for a round trip to Sydney for a comparison organized by the National Measurement Laboratory, (NML), Australia.
引用
收藏
页码:486 / 489
页数:4
相关论文
共 14 条
[1]   STATISTICS OF ATOMIC FREQUENCY STANDARDS [J].
ALLAN, DW .
PROCEEDINGS OF THE INSTITUTE OF ELECTRICAL AND ELECTRONICS ENGINEERS, 1966, 54 (02) :221-&
[2]   HETERODYNE SATURATION SPECTROSCOPY THROUGH FREQUENCY-MODULATION OF THE SATURATING BEAM [J].
CAMY, G ;
BORDE, CJ ;
DUCLOY, M .
OPTICS COMMUNICATIONS, 1982, 41 (05) :325-330
[3]   Direct link between microwave and optical frequencies with a 300 THz femtosecond laser comb [J].
Diddams, SA ;
Jones, DJ ;
Ye, J ;
Cundiff, ST ;
Hall, JL ;
Ranka, JK ;
Windeler, RS ;
Holzwarth, R ;
Udem, T ;
Hänsch, TW .
PHYSICAL REVIEW LETTERS, 2000, 84 (22) :5102-5105
[4]   Optical frequency measurement across a 104-THz gap with a femtosecond laser frequency comb [J].
Diddams, SA ;
Jones, DJ ;
Ma, LS ;
Cundiff, ST ;
Hall, JL .
OPTICS LETTERS, 2000, 25 (03) :186-188
[5]   Stabilization and frequency measurement of the I2-stabilized Nd:YAG laser [J].
Hall, JL ;
Ma, LS ;
Taubman, M ;
Tiemann, B ;
Hong, FL ;
Pfister, O ;
Ye, J .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1999, 48 (02) :583-586
[6]   SHIFT AND BROADENING OF SATURATED ABSORPTION RESONANCES DUE TO CURVATURE OF LASER WAVE FRONTS [J].
HALL, JL ;
BORDE, CJ .
APPLIED PHYSICS LETTERS, 1976, 29 (12) :788-790
[7]   A portable I2-stabilized Nd:YAG laser for wavelength standards at 532 nm and 1064 nm [J].
Hong, FL ;
Ishikawa, J ;
Yoon, TH .
RECENT DEVELOPMENTS IN OPTICAL GAUGE BLOCK METROLOGY, 1998, 3477 :2-10
[8]   Frequency comparison of 127I2-stabilized Nd:YAG lasers [J].
Hong, FL ;
Ishikawa, J ;
Yoda, J ;
Ye, J ;
Ma, LS ;
Hall, JL .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1999, 48 (02) :532-536
[9]   ABSOLUTE FREQUENCY OF THE MOLECULAR-IODINE TRANSITION R(56)32-0 NEAR 532 NM [J].
JUNGNER, PA ;
SWARTZ, S ;
EICKHOFF, M ;
YE, J ;
HALL, JL ;
WALTMAN, S .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1995, 44 (02) :151-154
[10]  
*NML NRLM CMS ITRI, 2001, INT COMP ORG NML PAR