A programmable pre-emphasis system

被引:29
作者
Gach, HM
Lowe, IJ
Madio, DP
Caprihan, A
Altobelli, SA
Kuethe, DO
Fukushima, E
机构
[1] Univ Pittsburgh, Dept Phys & Astron, Pittsburgh, PA 15260 USA
[2] Schlumberger Doll Res Ctr, Ridgefield, CT 06877 USA
[3] Lovelace REsp Res Inst, Albuquerque, NM USA
[4] Carnegie Mellon Univ, Pittsburgh NMR Ctr Biomed Res, Mellon Inst, Pittsburgh, PA 15213 USA
关键词
pre-emphasis; eddy current; gradient compensation; time constant;
D O I
10.1002/mrm.1910400313
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
MRI systems often use magnetic field gradient and shim pulse-shaping networks (pre-emphasis) to correct for magnetic field distortions caused by eddy currents. A pre-emphasis system that uses up to 16 fixed resistor-capacitor (RC) time constants per channel with programmable amplitude coefficients is described. The magnetic fields induced by the pre-emphasis RC time constants serve as a set of basis functions for compensating eddy-current fields induced by the gradient set. The resultant time-varying magnetic field gradient accurately reflects the gradient specified by the pulse programmer. Reductions in eddy-current fields are demonstrated for actively shielded and unshielded gradient sets.
引用
收藏
页码:427 / 431
页数:5
相关论文
共 6 条
[1]   ANALYTICAL METHOD FOR THE COMPENSATION OF EDDY-CURRENT EFFECTS INDUCED BY PULSED MAGNETIC-FIELD GRADIENTS IN NMR SYSTEMS [J].
JEHENSON, P ;
WESTPHAL, M ;
SCHUFF, N .
JOURNAL OF MAGNETIC RESONANCE, 1990, 90 (02) :264-278
[2]   QUANTITATIVE CHARACTERIZATION OF THE EDDY-CURRENT FIELDS IN A 40-CM BORE SUPERCONDUCTING MAGNET [J].
LIU, Q ;
HUGHES, DG ;
ALLEN, PS .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (01) :73-76
[3]   OPTIMIZATION OF EDDY-CURRENT COMPENSATION [J].
VANVAALS, JJ ;
BERGMAN, AH .
JOURNAL OF MAGNETIC RESONANCE, 1990, 90 (01) :52-70
[4]   A SIMPLE METHOD OF MEASURING GRADIENT INDUCED EDDY CURRENTS TO SET COMPENSATION NETWORKS [J].
WYSONG, RE ;
LOWE, IJ .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (01) :119-121
[5]   A NOVEL EDDY-CURRENT COMPENSATION SCHEME FOR PULSED GRADIENT SYSTEMS [J].
WYSONG, RE ;
MADIO, DP ;
LOWE, IJ .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (05) :572-575
[6]   An algorithm for eddy currents symmetrization and compensation [J].
Zur, Y ;
Stokar, S .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (02) :252-260