共 1 条
Simulations of wavelength-multiplexed holography for single-shot spatiotemporal characterization of NIF's advanced radiographic capability (ARC) laser
被引:1
|作者:
Grace, E.
[1
,2
]
Ma, T.
[2
]
Guang, Z.
[1
]
Alessi, D.
[2
]
Herriot, S.
[2
]
Rhodes, M.
[2
]
Park, J.
[3
]
Trebino, R.
[1
]
机构:
[1] Georgia Inst Technol, Sch Phys, 837 State St NW, Atlanta, GA 30332 USA
[2] Lawrence Livermore Natl Lab, 7000 East Ave, Livermore, CA 94550 USA
[3] Colorado State Univ, Elect & Comp Engn, 400 Isotope Dr, Ft Collins, CO 80523 USA
基金:
美国国家科学基金会;
关键词:
ULTRASHORT PULSES;
ELECTRIC-FIELD;
DISTORTIONS;
PHASE;
TIME;
METROLOGY;
CHIRP;
D O I:
10.1063/5.0043659
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
We simulate the use of a newly developed single-shot wavelength-multiplexed holography-based diagnostic, STRIPED FISH, to fully characterize the as-delivered laser pulses of the National Ignition Facility's Advanced Radiographic Capability (NIF-ARC) laser. To that end, we have performed simulations of the NIF-ARC pulse incorporating (a) a time-integrated spatial-profile measurement and a complete temporal-intensity-and-phase measurement using a frequency resolved optical gating, but without any spatiotemporal pulse characterizations, and (b) simulated first-order spatiotemporal distortions, which could be measured on a single shot if a STRIPED FISH device were deployed. Published under license by AIP Publishing.