Monte Carlo simulation of the standardization of 22Na using scintillation detector arrays

被引:10
作者
Sato, Y. [1 ]
Murayama, H. [2 ]
Yamada, T. [1 ,3 ,6 ]
Hasegawa, T. [4 ]
Oda, K. [5 ]
Unno, Y. [1 ]
Yunoki, A. [1 ]
机构
[1] Natl Inst Adv Ind Sci & Technol, Natl Metrol Inst Japan, Quantum Radiat Div, Radioact & Neutron Sect, Tsukuba, Ibaraki 3058568, Japan
[2] Natl Inst Radiol Sci, Inage Ku, Chiba 2638555, Japan
[3] Japan Radioisotope Assoc, Bunkyo Ku, Tokyo 1138941, Japan
[4] Kitasato Univ, Kanagawa 2288555, Japan
[5] Tokyo Metropolitan Inst Gerontol, Itabashi Ku, Tokyo 1730022, Japan
[6] Tohoku Univ, Aoba Ku, Sendai, Miyagi 9808579, Japan
关键词
Multi-channel scintillation detectors; Na-22; Monte Carlo; EGS5; Positron emission tomography;
D O I
10.1016/j.apradiso.2009.12.028
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In order to calibrate PET devices by a sealed point source, we contrived an absolute activity measurement method for the sealed point source using scintillation detector arrays This new method was verified by EGS5 Monte Carlo simulation. (C) 2009 Elsevier Ltd All rights reserved