A multi institutional comparison of imaging dose and technique protocols for neonatal chest radiography

被引:8
|
作者
Gunn, C. [1 ]
O'Brien, K. [2 ,3 ]
Fossa, K. [4 ]
Tonkopi, E. [2 ]
Lanca, L. [5 ,6 ,7 ]
Martins, C. T. [5 ,8 ]
Muller, H. [9 ,10 ]
Friedrich-Nel, H. [10 ]
Abdolell, M. [2 ]
Johansen, S. [11 ,12 ]
机构
[1] Dalhousie Univ, Sch Hlth Sci, Halifax, NS, Canada
[2] Dalhousie Univ, Fac Med, Halifax, NS, Canada
[3] IWK Hlth Ctr, Diagnost Imaging, Halifax, NS, Canada
[4] Oslo Univ Hosp, Rikshosp, Div Diagnost & Intervent, Oslo, Norway
[5] Inst Poliecn Lisboa, ESTeSL Escola Super Tecnol Saude Lisboa, Lisbon, Portugal
[6] Karolinska Inst, Stockholm, Sweden
[7] Singapore Inst Technol, Hlth & Social Sci Cluster, Singapore, Singapore
[8] Hosp Santa Maria, Radiol Dept, Ctr Hosp Lisboa Norte, EPE, Lisbon, Portugal
[9] Cent Univ Technol, Fac Hlth & Environm Sci, Free State CUT, Bloemfontein, South Africa
[10] Univ Acad Hosp, Dept Clin Imaging Sci, Bloemfontein, South Africa
[11] Oslo Metropolitan Univ OsloMet, Fac Hlth Sci, Oslo, Norway
[12] Oslo Univ Hosp, Dept Oncol, Div Canc Med, Oslo, Norway
关键词
Chest x-ray; Paediatric imaging; Exposure parameters; BREAST-CANCER; OPTIMIZATION; QUALITY; WOMEN; RISK;
D O I
10.1016/j.radi.2019.10.013
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: The focus on paediatric radiation dose reduction supports reevaluation of paediatric imaging protocols. This is particularly important in the neonates where chest radiographs are frequently requested to assess respiratory illness and line placement. This study aims to assess the impact of neonatal chest radiographic protocols on patient dose in four hospitals in different countries. Methods: Exposure parameters, collimation, focus to skin distance (FSD) and radiation dose from 200 neonatal chest radiographs were registered prospectively. Inclusion criteria consisted of both premature and full-term neonates weighing between 1000 and 5000 g. Only data from the examinations meeting diagnostic criteria and approved for the clinical use were included. Radiation dose was assessed using dose area product (DAP). Results: The lowest DAP value (4.58 mGy cm(2)) was recorded in the Norwegian hospital, employing a high kVp, low mAs protocol using a DR system. The Canadian hospital recorded the highest DAP (9.48), using lower kVp and higher mAs with a CR system, including the addition of a lateral projection. The difference in the mean DAP, weight, field of view (FOV) and kVp between the hospitals is statistically significant (p < 0.001). Conclusion: Use of non-standardised imaging protocols in neonatal chest radiography results in differences in patient dose across hospitals included in the study. Using higher kVp, lower mAs and reducing the number of lateral projections to clinically relevant indications result in a lower DAP measured in the infant sample studied. Further studies to examine image quality based on exposure factors and added filtration are recommended. Implications for practice: Reevaluation of paediatric imaging protocols presents an opportunity to reduce patient dose in a population with increased sensitivity to ionising radiation. (c) 2019 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:E66 / E72
页数:7
相关论文
共 50 条
  • [1] Neonatal chest radiography: Influence of standard clinical protocols and radiographic equipment on pathology visibility and radiation dose using a neonatal chest phantom
    Al-Murshedi, S.
    Hogg, Peter
    England, A.
    RADIOGRAPHY, 2020, 26 (04) : 282 - 287
  • [2] A multi-institutional assessment of low-dose protocols in chest computed tomography: Dose and image quality
    Tonkopi, Elena
    Tetteh, Mercy Afadzi
    Gunn, Catherine
    Ashraf, Haseem
    Rusten, Sigrid Lia
    Safi, Perkhah
    Tinsoe, Nora Suu
    Colford, Kylie
    Ouellet, Olivia
    Naimi, Salma
    Johansen, Safora
    ACTA RADIOLOGICA OPEN, 2024, 13 (01)
  • [3] Optimisation of radiation dose and image quality in mobile neonatal chest radiography
    Hinojos-Armendariz, V. I.
    Mejia-Rosales, S. J.
    Franco-Cabrera, M. C.
    RADIOGRAPHY, 2018, 24 (02) : 104 - 109
  • [4] Optimization of chest radiographic imaging parameters: a comparison of image quality and entrance skin dose for digital chest radiography systems
    Sun, Zhonghua
    Lin, Chenghsun
    Tyan, YeuSheng
    Ng, Kwan-Hoong
    CLINICAL IMAGING, 2012, 36 (04) : 279 - 286
  • [5] DUAL-SCREEN COMPUTED RADIOGRAPHY TECHNIQUE FOR IMAGING OF THE CHEST
    SHAW, CC
    WANG, T
    BREITENSTEIN, DS
    RADIOLOGY, 1995, 197 : 325 - 325
  • [6] Low-Dose Chest CT Protocols for Imaging COVID-19 Pneumonia: Technique Parameters and Radiation Dose
    Suliman, Ibrahim I.
    Khouqeer, Ghada A.
    Ahmed, Nada A.
    Abuzaid, Mohamed M.
    Sulieman, Abdelmoneim
    LIFE-BASEL, 2023, 13 (04):
  • [7] Cu filtration for dose reduction in neonatal chest imaging
    Smans, K.
    Struelens, L.
    Smet, M.
    Bosmans, H.
    Vanhavere, F.
    RADIATION PROTECTION DOSIMETRY, 2010, 139 (1-3) : 281 - 286
  • [8] Chest Computed Radiography Imaging Parameters for Nodule Detection: A Comparison of Image Quality and Entrance Skin Dose
    Sun, Zhonghua
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2014, 4 (04) : 587 - 592
  • [9] DIGITAL RADIOGRAPHY AND CONVENTIONAL IMAGING OF THE CHEST - A COMPARISON OF OBSERVER PERFORMANCE
    THAETE, FL
    FUHRMAN, CR
    OLIVER, JH
    BRITTON, CA
    CAMPBELL, WI
    FEIST, JH
    STRAUB, WH
    DAVIS, PL
    PLUNKETT, MB
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1994, 162 (03) : 575 - 581
  • [10] Anthropomorphic chest phantom imaging - The potential for dose creep in computed radiography
    Ma, W. K.
    Hogg, P.
    Tootell, A.
    Manning, D.
    Thomas, N.
    Kane, T.
    Kelly, J.
    McKenzie, M.
    Kitching, J.
    RADIOGRAPHY, 2013, 19 (03) : 207 - 211