A Real-Time Vital-Sign Monitoring in the Physical Domain on a Mixed-Signal Reconfigurable Platform

被引:8
作者
Shah, Sahil [1 ]
Toreyin, Hakan [2 ]
Gungor, Cihan Berk [2 ,3 ]
Hasler, Jennifer [4 ]
机构
[1] CALTECH, Pasadena, CA 92115 USA
[2] San Diego State Univ, Dept Elect & Comp Engn, San Diego, CA 92182 USA
[3] San Diego State Univ, San Diego, CA 92182 USA
[4] Georgia Inst Technol, Sch Elect & Comp Engn, Atlanta, GA 30308 USA
关键词
Field programmable analog arrays; Electrocardiography; Biomedical monitoring; Real-time systems; Monitoring; Energy efficiency; Feature extraction; Energy-efficient cardiac signal processing; real-time physical signal processing; wearable sensor nodes; QRS DETECTION; LOW-POWER; ECG; SYSTEM; IMPLEMENTATION; ACQUISITION; DESIGN; SOC;
D O I
10.1109/TBCAS.2019.2949778
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work presents a mixed-signal physical-compu-tation-electronics for monitoring three vital signs; namely heart rate, blood pressure, and blood oxygen saturation; from electrocardiography, arterial blood pressure, and photoplethysmography signals in real-time. The computational circuits are implemented on a reconfigurable and programmable signal-processing platform, namely field-programmable analog array (FPAA). The design leverages the core enabling technology of FPAA, namely floating-gate CMOS devices, and an on-chip low-power microcontroller to achieve energy-efficiency while not compromising accuracy. The custom physical-computation-electronics operating in CMOS subthreshold region, performs low-level (i.e., physiologically-relevant feature extraction) and high-level (i.e., detecting arrhythmia) signal processing in an energy-efficient manner. The on-chip microcontroller is used (1) in the programming mode for controlling the charge storage at the analog-memory elements to introduce patient-dependency into the system and (2) in the run mode to quantify the vital signs. The system has been validated against digital computation results from MATLAB using datasets collected from three healthy subjects and datasets from the MIT/BIH open source database. Based on all recordings in the MIT/BIH database, ECG R-peak detection sensitivity is 94.2. The processor detects arrhythmia in three MIT/BIH recordings with an average sensitivity of 96.2. The cardiac processor achieves an average percentage mean error bounded by 3.75, 6.27, and 7.3 for R-R duration, systolic blood pressure, and oxygen saturation level calculations; respectively. The power consumption of the ECG, blood-pressure and photo-plethysmography processing circuitry are 126 nW, 251nW and 1.44W respectively in a 350nm process. Overall, the cardiac processor consumes 1.82W.
引用
收藏
页码:1690 / 1699
页数:10
相关论文
共 59 条
  • [21] QUANTITATIVE INVESTIGATION OF QRS DETECTION RULES USING THE MIT/BIH ARRHYTHMIA DATABASE
    HAMILTON, PS
    TOMPKINS, WJ
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (12) : 1157 - 1165
  • [22] Finding a roadmap to achieve large neuromorphic hardware systems
    Hasler, Jennifer
    Marr, Bo
    [J]. FRONTIERS IN NEUROSCIENCE, 2013, 7
  • [23] Higuera V., 2016, IS HEART DIS DIAGNOS
  • [24] A 48.6-to-105.2 μW Machine Learning Assisted Cardiac Sensor SoC for Mobile Healthcare Applications
    Hsu, Shu-Yu
    Ho, Yingchieh
    Chang, Po-Yao
    Su, Chauchin
    Lee, Chen-Yi
    [J]. IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2014, 49 (04) : 801 - 811
  • [25] MIMIC-III, a freely accessible critical care database
    Johnson, Alistair E. W.
    Pollard, Tom J.
    Shen, Lu
    Lehman, Li-wei H.
    Feng, Mengling
    Ghassemi, Mohammad
    Moody, Benjamin
    Szolovits, Peter
    Celi, Leo Anthony
    Mark, Roger G.
    [J]. SCIENTIFIC DATA, 2016, 3
  • [26] Strategies for Implementing Screening for Critical Congenital Heart Disease
    Kemper, Alex R.
    Mahle, William T.
    Martin, Gerard R.
    Cooley, W. Carl
    Kumar, Praveen
    Morrow, W. Robert
    Kelm, Kellie
    Pearson, Gail D.
    Glidewell, Jill
    Grosse, Scott D.
    Howell, R. Rodney
    [J]. PEDIATRICS, 2011, 128 (05) : E1259 - E1267
  • [27] A Configurable and Low-Power Mixed Signal SoC for Portable ECG Monitoring Applications
    Kim, Hyejung
    Kim, Sunyoung
    Van Helleputte, Nick
    Artes, Antonio
    Konijnenburg, Mario
    Huisken, Jos
    Van Hoof, Chris
    Yazicioglu, Refet Firat
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2014, 8 (02) : 257 - 267
  • [28] Calibration of Floating-Gate SoC FPAA System
    Kim, Sihwan
    Shah, Sahil
    Hasler, Jennifer
    [J]. IEEE TRANSACTIONS ON VERY LARGE SCALE INTEGRATION (VLSI) SYSTEMS, 2017, 25 (09) : 2649 - 2657
  • [29] Integrated Floating-Gate Programming Environment for System-Level ICs
    Kim, Sihwan
    Hasler, Jennifer
    George, Suma
    [J]. IEEE TRANSACTIONS ON VERY LARGE SCALE INTEGRATION (VLSI) SYSTEMS, 2016, 24 (06) : 2244 - 2252
  • [30] A Multi(bio)sensor Acquisition System With Integrated Processor, Power Management, 8 x 8 LED Drivers, and Simultaneously Synchronized ECG, BIO-Z, GSR, and Two PPG Readouts
    Konijnenburg, Mario
    Stanzione, Stefano
    Yan, Long
    Jee, Dong-Woo
    Pettine, Julia
    van Wegberg, Roland
    Kim, Hyejung
    van Liempd, Chris
    Fish, Ram
    Schuessler, James
    de Groot, Harmke
    Van Hoof, Chris
    Yazicioglu, Refet Firat
    Van Helleputte, Nick
    [J]. IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2016, 51 (11) : 2584 - 2595