Comparability of three intraocular pressure measurement: iCare pro rebound, non-contact and Goldmann applanation tonometry in different IOP group

被引:53
作者
Chen, Min [1 ,2 ]
Zhang, Lina [1 ,3 ]
Xu, Jia [1 ,2 ]
Chen, Xinyi [1 ,2 ]
Gu, Yuxiang [1 ,2 ]
Ren, Yuping [4 ]
Wang, Kaijun [1 ,2 ]
机构
[1] Zhejiang Univ, Coll Med, Affiliated Hosp 2, Eye Ctr, 88 Jiefang Rd, Hangzhou 310009, Zhejiang, Peoples R China
[2] Zhejiang Prov Key Lab Ophthalmol, Hangzhou, Zhejiang, Peoples R China
[3] Lishui Peoples Hosp, Lishui, Zhejiang, Peoples R China
[4] Shaoxing Tradit Chinese Med Hosp, Shaoxing, Zhejiang, Peoples R China
基金
中国国家自然科学基金;
关键词
Intraocular pressure; iCare rebound tonometer; Non-contact tonometer; Goldmann applanation tonometer; Central corneal thickness; OCULAR RESPONSE ANALYZER; DYNAMIC CONTOUR; HEALTHY-SUBJECTS;
D O I
10.1186/s12886-019-1236-5
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background Measurement of intraocular pressure (IOP) is essential for glaucoma patients. Many factors such as central corneal thickness (CCT) can affect the accuracy of IOP measurement. The purpose of this study was to evaluate the agreement of IOP measured by non-contact tonometer (NCT), iCare pro rebound tonometer (iCare), and Goldmann applanation tonometer (GAT) in different IOP group. Methods This was a Hospital-based cross-sectional study. Two hundred subjects were enrolled in this study. All subjects underwent IOP measurement using an NCT-iCare-GAT sequence. Bland-Altman, Pearson correlation and intraclass correlation analysis were performed using SPSS 17.0 software. The influence of CCT on each IOP measurement methods was evaluated by linear regression analysis. Results The mean difference (Delta) of NCT-GAT did not differ from (Delta) iCare-GAT in IOP < 10 and 10-21 mmHg group. However, (Delta) NCT-GAT was significantly higher than (Delta) iCare-GAT in IOP 22-30 and > 30 mmHg group (P < 0.05). Bland-Altman analysis showed significant agreement between the three devices (P < 0.01). IOP measurements of the three methods were significantly correlated with CCT (P < 0.01). Conclusions ICare pro shows a higher agreement with GAT over a wide range of IOP compared with NCT. The consistency between the three tonometers was similar in a low and normal IOP range. However, NCT shows a greater overestimate of IOP in moderate and higher IOP group. The variability of IOP measurement affected by CCT is NCT > iCare pro > GAT.
引用
收藏
页数:10
相关论文
共 42 条
  • [21] A comparison of Icare PRO and Tono-Pen XL tonometers in anesthetized children
    McKee, Elliot C.
    Ely, Amanda L.
    Duncan, Jared E.
    Dosunmu, Eniolami O.
    Freedman, Sharon F.
    [J]. JOURNAL OF AAPOS, 2015, 19 (04): : 332 - 337
  • [22] Mohan Shalini, 2014, J Optom, V7, P86, DOI 10.1016/j.optom.2013.06.002
  • [23] Molina E, 2006, MED INTERNA MEXICO, V22, P165
  • [24] Clinical Evaluation of the New Rebound Tonometers Icare PRO and Icare ONE Compared With the Goldmann Tonometer
    Moreno-Montanes, Javier
    Martinez-de-la-Casa, Jose M.
    Sabater, Alfonso L.
    Morales-Fernandez, Laura
    Saenz, Cristina
    Garcia-Feijoo, Julian
    [J]. JOURNAL OF GLAUCOMA, 2015, 24 (07) : 527 - 532
  • [25] Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range
    Munkwitz, S.
    Elkarmouty, A.
    Hoffmann, E. M.
    Pfeiffer, N.
    Thieme, H.
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2008, 246 (06) : 875 - 879
  • [26] Icare (R) rebound tonometers: review of their characteristics and ease of use
    Nakakura, Shunsuke
    [J]. CLINICAL OPHTHALMOLOGY, 2018, 12 : 1245 - 1253
  • [27] The 24-hour intraocular pressure control by tafluprost/timolol fixed combination after switching from the concomitant use of tafluprost and timolol gel-forming solution, in patients with primary open-angle glaucoma
    Nakamoto, Kenji
    Takeshi, Masahiko
    Hiraoka, Toshihiko
    Eguchi, Mayuko
    Nakano, Yuichiro
    Otsuka, Naomi
    Hizaki, Hiroko
    Akai, Hiromi
    Hashimoto, Masayo
    [J]. CLINICAL OPHTHALMOLOGY, 2018, 12 : 359 - 367
  • [28] Nuyen Brenda, 2016, Open Ophthalmol J, V10, P44, DOI 10.2174/1874364101610010044
  • [29] Assessment of the accuracy and reliability of the Topcon CT80 non-contact tonometer
    Ogbuehi, Kelechi C.
    [J]. CLINICAL AND EXPERIMENTAL OPTOMETRY, 2006, 89 (05) : 310 - 314
  • [30] Comparison of IOP measurement by ocular response analyzer, dynamic contour, Goldmann applanation, and noncontact tonometry
    Oncel, Banu
    Dinc, Umut Asli
    Orge, Faruk
    Yalvac, Belkis Ilgaz
    [J]. EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2009, 19 (06) : 936 - 941