[1] Inst Canc Res, Radiotherapy & Imaging Div, London, England
[2] Imperial Coll London, Dept Mech Engn, London, England
来源:
INTERNATIONAL ULTRASONICS SYMPOSIUM (IEEE IUS 2021)
|
2021年
关键词:
Attenuation correction;
Ultrasound Backscatter Coefficient;
Finite Element Method;
Numerical Simulation;
D O I:
10.1109/IUS52206.2021.9593686
中图分类号:
O42 [声学];
学科分类号:
070206 ;
082403 ;
摘要:
The methods and results presented here test the ability of the multi-band finite element method to simulate the frequency dependence of attenuation of ultrasound within soft tissue like materials, and the potential translation of the approach from its previous applications in non-destructive testing into medical ultrasound. Using a variety of materials, the approach was tested by comparing the theoretical attenuation (based on the input attenuation constants and path length through the materials) to the attenuation of the signal energy within the simulation. In two models, hypothetical materials were devised, with arbitrary attenuation properties and acoustic impedance matched to a background with the density and sound speed of water. In the third model, acoustic impedance mismatches were introduced to simulate the signal energy loss due to reflection and attenuation as an ultrasound beam passes from water into parallel layers of human skin, fat and muscle. In all cases, the calculated signal loss (from the reflection coefficients and attenuation constants) agreed with the signal loss measured within the simulation to a high degree, reproducing the input parameters to four significant figures in all cases.