Validity, Practical Utility, and Reliability of the activPAL™ in Preschool Children

被引:76
作者
Davies, Gwyneth [1 ]
Reilly, John I. [2 ]
McGowan, Amy J. [1 ]
Dall, Philippa M. [3 ]
Granat, Malcolm H. [3 ]
Paton, James Y. [1 ]
机构
[1] Univ Glasgow, Sch Med, Glasgow, Lanark, Scotland
[2] Univ Strathclyde, Phys Act Hlth Res Grp, Glasgow, Lanark, Scotland
[3] Glasgow Caledonian Univ, Sch Hlth, Glasgow G4 0BA, Lanark, Scotland
关键词
ACTIVITY MONITORING; POSTURE; CHILD; VALIDATION; BODY-MASS INDEX; TOTAL-ENERGY EXPENDITURE; DAILY PHYSICAL-ACTIVITY; ACTIVITY MONITOR; AMBULATORY ACCELEROMETRY; SEDENTARY BEHAVIOR; POSTURE ALLOCATION; REFERENCE CURVES; US ADULTS; OBESITY;
D O I
10.1249/MSS.0b013e31823b1dc7
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
DAVIES, G., J. J. REILLY, A. J. MCGOWAN, P. M. DALL, M. H. GRANAT, and J. Y. PATON. Validity, Practical Utility, and Reliability of the activPAL (TM) in Preschool Children. Med. Sci. Sports Exerc., Vol. 44, No. 4, pp. 761-768, 2012. Purpose: With the increasing global prevalence of childhood obesity, it is important to have appropriate measurement tools for investigating factors (e.g., sedentary time) contributing to positive energy balance in early childhood. For preschool children, single-unit monitors such as the activPAL (TM) are promising. However, validation is required because activity patterns differ from adults. Methods: Thirty preschool children participated in a validation study. Children undertaking usual nursery activity while wearing an activPAL (TM) 1 h were recorded using a video camera. Video (criterion method) was analyzed on a second-by-second basis to categorize posture and activity. This was compared with the corresponding activPAL (TM) output. In a subsequent substudy investigating practical utility and reliability, 20 children wore an activPAL (TM) for seven consecutive 24-h periods. Results: A total of 97,750 s of direct observation from 30 children were categorized as sit/lie (46%), stand (35%), and walk (16%), with 3% of time in non-sit/lie/upright postures (e.g., crawl/crouch/kneel-up). Sensitivity for the overall total time-matched seconds detected as activPAL (TM) "sit/lie" was 86.7%, specificity was 97.1%, and positive predictive value was 96.3%. For individual children, the median (interquartile range) sensitivity for activPAL (TM) sit/lie was 92.8% (76.1%-97.4%), specificity was 97.3% (94.9%-99.2%), and positive predictive value was 97.0% (91.5%-99.1%). The activPAL (TM) underestimated total time spent sitting (mean difference = -4.4%, P < 0.01) and overestimated time standing (mean difference = 7.1%, P < 0.01). There was no difference in overall percent time categorized as "walk" (P = 0.2). The monitors were well tolerated by children during a 7-d period of free-living activity. In the reliability study, at least 5 d of monitoring was required to obtain an intraclass correlation coefficient of >= 0.8 for time spent "sit/lie" according to activPAL (TM) output. Conclusions: The activPAL (TM) had acceptable validity, practical utility, and reliability for the measurement of posture and activity during free-living activities in preschool children.
引用
收藏
页码:761 / 768
页数:8
相关论文
共 40 条
[1]   Physical activity monitoring based on accelerometry: validation and comparison with video observation [J].
Aminian, K ;
Robert, P ;
Buchser, EE ;
Rutschmann, B ;
Hayoz, D ;
Depairon, M .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1999, 37 (03) :304-308
[2]   Validity of the ActiReg system in assessing energy requirement in chronic obstructive pulmonary disease patients [J].
Arvidsson, D ;
Slinde, F ;
Nordenson, A ;
Larsson, S ;
Hulthén, L .
CLINICAL NUTRITION, 2006, 25 (01) :68-74
[3]   THE LEVEL AND TEMPO OF CHILDRENS PHYSICAL ACTIVITIES - AN OBSERVATIONAL STUDY [J].
BAILEY, RC ;
OLSON, J ;
PEPPER, SL ;
PORSZASZ, J ;
BARSTOW, TJ ;
COOPER, DM .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 1995, 27 (07) :1033-1041
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Design of Posture and Activity Detector (PAD) [J].
Bliley, Kara E. ;
Schwab, Daniel J. ;
Zahn, Sharon K. ;
Rowley, Katharine L. ;
Kane, Paul H. ;
Levine, James A. ;
Daniel, Erik S. ;
Gilbert, Barry K. .
2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, :2659-+
[6]   Ambulatory monitoring of children's activity [J].
Busser, HJ ;
Ott, J ;
vanLummel, RC ;
Uiterwaal, M ;
Blank, R .
MEDICAL ENGINEERING & PHYSICS, 1997, 19 (05) :440-445
[7]   Quantification of physical activities by means of ambulatory accelerometry: A validation study [J].
Bussmann, JBJ ;
Tulen, JHM ;
van Herel, ECG ;
Stam, HJ .
PSYCHOPHYSIOLOGY, 1998, 35 (05) :488-496
[8]   Measuring daily behavior using ambulatory accelerometry: The activity monitor [J].
Bussmann, JBJ ;
Martens, WLJ ;
Tulen, JHM ;
Schasfoort, FC ;
van den Berg-Emons, HJG ;
Stam, HJ .
BEHAVIOR RESEARCH METHODS INSTRUMENTS & COMPUTERS, 2001, 33 (03) :349-356
[9]   BODY-MASS INDEX REFERENCE CURVES FOR THE UK, 1990 [J].
COLE, TJ ;
FREEMAN, JV ;
PREECE, MA .
ARCHIVES OF DISEASE IN CHILDHOOD, 1995, 73 (01) :25-29
[10]   Global prevalence and trends of overweight and obesity among preschool children [J].
de Onis, Mercedes ;
Blossner, Monika ;
Borghi, Elaine .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 2010, 92 (05) :1257-1264