A method to measure the accuracy of loads in knee-ankle-foot orthoses using conventional gait analysis, applied to persons with poliomyelitis

被引:8
作者
Andrysek, Jan [1 ]
Redekop, Susan [1 ]
Matsui, Naomi C. [2 ]
Kooy, John [2 ]
Hubbard, Sheila [2 ]
机构
[1] Bloorview Kids Rehab, Bloorview Res Inst, Toronto, ON M4G 1R8, Canada
[2] Bloorview Kids Rehab, Dept Clin Technol, Toronto, ON M4G 1R8, Canada
来源
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION | 2008年 / 89卷 / 07期
基金
加拿大健康研究院;
关键词
gait; orthotic devices; rehabilitation;
D O I
10.1016/j.apmr.2007.11.045
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Objectives: To determine (1) the forces and moments passing through knee-ankle-foot orthoses (KAFOs) during walking and (2) the accuracy with which these loads can be measured using conventional gait analysis techniques. Design: Comparative case series. Setting: Rehabilitation facility with human movement laboratory (gait lab). Participants: Four patients with poliomyelitis wearing KAFOs. Interventions: KAFOs were instrumented with a load cell, and walking data were concurrently collected using conventional gait analysis. Main Outcome Measures: Load measurements and gait parameters. Results: Predominant orthotic loads (knee joint forces and moments) were composed of knee flexion moments and axial compression forces. With conventional gait analysis, peak knee joint moments were substantially underestimated compared with those directly measured using the load cell. Defining the knee axis anatomically versus at the orthotic axis, tracking it dynamically, and compensating for each patient's corrected knee flexion contracture resulted in considerable improvements in the gait lab estimates of knee joint moments. Conclusions: A practical method that directly measures moments and forces in conventional KAFOs has been applied to show that conventional gait analysis techniques substantially underestimate knee joint moments in the KAFOs of persons with poliomyelitis. Underestimation of orthotic loads could result in underdesigned orthotic components and ultimately higher incidence of component failure in clinical applications.
引用
收藏
页码:1372 / 1379
页数:8
相关论文
共 18 条
[1]  
[Anonymous], 2000, Phys Med Rehabil
[2]   Pelvis and lower limb anatomical landmark calibration precision and its propagation to bone geometry and joint angles [J].
Della Croce, U ;
Cappozzo, A ;
Kerrigan, DC .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1999, 37 (02) :155-161
[3]  
Goldberg B, 1997, ATLAS ORTHOSIS ASSIS
[4]   Gait evaluation of an automatic stance-control knee orthosis in a patient with postpoliomyelitis [J].
Hebert, JS ;
Liggins, AB .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2005, 86 (08) :1676-1680
[5]   Performance specification for lower limb orthotic devices [J].
Johnson, GR ;
Ferrarin, M ;
Harrington, M ;
Hermens, H ;
Jonkers, I ;
Mak, P ;
Stallard, J .
CLINICAL BIOMECHANICS, 2004, 19 (07) :711-718
[6]   MEASUREMENT OF LOWER-EXTREMITY KINEMATICS DURING LEVEL WALKING [J].
KADABA, MP ;
RAMAKRISHNAN, HK ;
WOOTTEN, ME .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1990, 8 (03) :383-392
[7]  
KAUFMAN KR, 1996, J PROSTHET ORTHOT, V8, P79
[8]  
Kendall F.P., 1993, MUSCLES TESTING FUNC
[9]  
Lehmann J F, 1970, Arch Phys Med Rehabil, V51, P331
[10]  
Lehmann J F, 1971, Arch Phys Med Rehabil, V52, P366