A subject-specific pelvic bone model and its application to cemented acetabular replacements

被引:44
作者
Zhang, Qing-Hang [1 ]
Wang, Ji-Yuan [1 ]
Lupton, Colin [1 ]
Heaton-Adegbile, Phillip [1 ]
Guo, Zi-Xue [2 ]
Liu, Qin [2 ]
Tong, Jie [1 ]
机构
[1] Univ Portsmouth, Dept Mech & Design Engn, Mech Behav Mat Lab, Portsmouth PO1 3DJ, Hants, England
[2] Beijing Huatuo Biomech Lab, Beijing, Peoples R China
关键词
Subject-specific; Pelvic bone; Finite element; Cup position; Cement mantle thickness; Bone-cement interface; Acetabular replacement; FINITE-ELEMENT MODEL; HIP-JOINT; VALIDATION; COMPONENT; SPACERS; MANTLE; WEAR; CUP;
D O I
10.1016/j.jbiomech.2010.06.023
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A subject-specific three-dimensional finite element (FE) pelvic bone model has been developed and applied to the study of bone-cement interfacial response in cemented acetabular replacements. The pelvic bone model was developed from CT scan images of a cadaveric pelvis and validated against the experiment data obtained from the same specimen at a simulated single-legged stance. The model was then implanted with a cemented acetabular cup at selected positions to simulate some typical implant conditions due to the misplacement of the cup as well as a standard cup condition. For comparison purposes, a simplified FE model with homogeneous trabecular bone material properties was also generated and similar implant conditions were examined. The results from the homogeneous model are found to underestimate significantly both the peak von Mines stress and the area of the highly stressed region in the cement near the bone-cement interface, compared with those from the subject-specific model. Non-uniform cement thickness and non-standard cup orientation seem to elevate the highly stressed region as well as the peak stress near the bone-cement interface. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:2722 / 2727
页数:6
相关论文
共 29 条
[1]   Subject-specific finite element model of the pelvis: Development, validation and sensitivity studies [J].
Anderson, AE ;
Peters, CL ;
Tuttle, BD ;
Weiss, JA .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (03) :364-373
[2]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[3]   Poor accuracy of freehand cup positioning during total hip arthroplasty [J].
Bosker, B. H. ;
Verheyen, C. C. P. M. ;
Horstmann, W. G. ;
Tulp, N. J. A. .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2007, 127 (05) :375-379
[4]   Computational Assessment of the Effect of Polyethylene Wear Rate, Mantle Thickness, and Porosity on the Mechanical Failure of the Acetabular Cement Mantle [J].
Coultrup, Oliver J. ;
Hunt, Chris ;
Wroblewski, B. Michael ;
Taylor, Mark .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2010, 28 (05) :565-570
[5]   MECHANICAL AND TEXTURAL PROPERTIES OF PELVIC TRABECULAR BONE [J].
DALSTRA, M ;
HUISKES, R ;
ODGAARD, A ;
VANERNING, L .
JOURNAL OF BIOMECHANICS, 1993, 26 (4-5) :523-535
[6]   DEVELOPMENT AND VALIDATION OF A 3-DIMENSIONAL FINITE-ELEMENT MODEL OF THE PELVIC BONE [J].
DALSTRA, M ;
HUISKES, R ;
VANERNING, L .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1995, 117 (03) :272-278
[7]   LOAD-TRANSFER ACROSS THE PELVIC BONE [J].
DALSTRA, M ;
HUISKES, R .
JOURNAL OF BIOMECHANICS, 1995, 28 (06) :715-724
[8]  
DALSTRA M, 1993, THESIS U NJEMEGEN NE
[9]  
DELEE JG, 1976, CLIN ORTHOP RELAT R, P20
[10]   Prediction of mechanical properties of cortical bone by quantitative computed tomography [J].
Duchemin, L. ;
Bousson, V. ;
Raossanaly, C. ;
Bergot, C. ;
Laredo, J. D. ;
Skalli, W. ;
Mitton, D. .
MEDICAL ENGINEERING & PHYSICS, 2008, 30 (03) :321-328