Evaluation of two-dimensional dose distributions for pre-treatment patient-specific IMRT dosimetry

被引:5
|
作者
Radojcic, Deni Smilovic [1 ]
Rajlic, David [1 ]
Casar, Bozidar [2 ]
Kolacio, Manda Svabic [1 ]
Obajdin, Nevena [1 ]
Faj, Dario [3 ,4 ]
Jurkovic, Slaven [1 ,5 ]
机构
[1] Univ Hosp Rijeka, Med Phys Dept, Kresimirova 42, Rijeka, Croatia
[2] Inst Oncol LJubljana, Dept Radiat Phys, Ljubljana, Slovenia
[3] Univ Osijek, Fac Med, Osijek, Croatia
[4] Univ Osijek, Fac Dent Med & Hlth, Osijek, Croatia
[5] Univ Rijeka, Dept Med Phys & Biophys, Fac Med, Rijeka, Croatia
关键词
IMRT; 2D dose verification; gamma method; anthropomorphic phantom; WATER; VERIFICATION; PHOTON; TISSUE; BEAMS; FILM;
D O I
10.2478/raon-2018-0019
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background. The accuracy of dose calculation is crucial for success of the radiotherapy treatment. One of the methods that represent the current standard for patient-specific dosimetry is the evaluation of dose distributions measured with an ionization chamber array inside a homogeneous phantom using gamma method. Nevertheless, this method does not replicate the realistic conditions present when a patient is undergoing therapy. Therefore, to more accurately evaluate the treatment planning system (TPS) capabilities, gamma passing rates were examined for beams of different complexity passing through inhomogeneous phantoms. Materials and methods. The research was performed using Siemens Oncor Expression linear accelerator, Siemens Somatom Open CT simulator and Elekta Monaco TPS. A 2D detector array was used to evaluate dose distribution accuracy in homogeneous, semi-anthropomorphic and anthropomorphic phantoms. Validation was based on gamma analysis with 3%/3mm and 2%/2mm criteria, respectively. Results. Passing rates of the complex dose distributions degrade depending on the thickness of non-water equivalent material. They also depend on dose reporting mode used. It is observed that the passing rate decreases with plan complexity. Comparison of the data for all set-ups of semi-anthropomorphic and anthropomorphic phantoms shows that passing rates are higher in the anthropomorphic phantom. Conclusions. Presented results raise a question of possible limits of dose distribution verification in assessment of plan delivery quality. Consequently, good results obtained using standard patient specific dosimetry methodology do not guarantee the accuracy of delivered dose distribution in real clinical cases.
引用
收藏
页码:346 / 352
页数:7
相关论文
共 50 条
  • [31] Stereotactic body radiation therapy patient specific quality assurance using a two-dimensional array at extended source to surface distance
    Stathakis, Sotirios
    Mavroidis, Panayiotis
    Narayanasamy, Ganesh
    Markovic, Miljenko
    Myers, Pamela
    Papanikolaou, Niko
    JOURNAL OF BUON, 2015, 20 (04): : 1154 - 1163
  • [32] Examination of the properties of IMRT and VMAT beams and evaluation against pre-treatment quality assurance results
    Crowe, S. B.
    Kairn, T.
    Middlebrook, N.
    Sutherland, B.
    Hill, B.
    Kenny, J.
    Langton, C. M.
    Trapp, J. V.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (06): : 2587 - 2601
  • [33] PATIENT-SPECIFIC 3D PRETREATMENT AND POTENTIAL 3D ONLINE DOSE VERIFICATION OF MONTE CARLO-CALCULATED IMRT PROSTATE TREATMENT PLANS
    Boggula, Ramesh
    Jahnke, Lennart
    Wertz, Hansjoerg
    Lohr, Frank
    Wenz, Frederik
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (04): : 1168 - 1175
  • [34] Patient-specific voxel phantom dosimetry during the prostate treatment with high-energy linac
    Mohammadi, Najmeh
    Miri-Hakimabad, Hashem
    Rafat-Motavalli, Laleh
    Akbari, Fatemeh
    Abdollahi, Sara
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2015, 304 (02) : 785 - 792
  • [35] A feasibility study on an automated method to generate patient-specific dose distributions for radiotherapy using deep learning
    Chen, Xinyuan
    Men, Kuo
    Li, Yexiong
    Yi, Junlin
    Dai, Jianrong
    MEDICAL PHYSICS, 2019, 46 (01) : 56 - 64
  • [36] Analysis of clinical patient-specific pre-treatment quality assurance with the new helical tomotherapy platform, following the AAPM TG-218 report
    Fusella, Marco
    Cavinato, Samuele
    Germani, Alessandra
    Paiusco, Marta
    Pivato, Nicola
    Rossato, Marco Andrea
    Scott, Anthony
    Scaggion, Alessandro
    RADIATION ONCOLOGY, 2021, 16 (01)
  • [37] Evaluation of automated pre-treatment and transit in-vivo dosimetry in radiotherapy using empirically determined parameters
    Bossuyt, Evy
    Weytjens, Reinhilde
    Nevens, Daan
    De Vos, Sarah
    Verellen, Dirk
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2020, 16 : 113 - 129
  • [38] Two-dimensional in vivo rectal dosimetry during high-dose-rate brachytherapy for cervical cancer: a phantom study
    Jeang, Eun Hee
    Goh, Youngmoon
    Cho, Kwan Ho
    Min, Soonki
    Choi, Sang Hyoun
    Jeong, Hojin
    Jo, Kwanghyun
    Lee, Nuri
    Song, Sanghyeon
    Lee, Se Byeong
    Shin, Dongho
    Kim, Yeon-Joo
    Kim, Joo-Young
    Kim, Dae Yong
    Hwang, Ui-Jung
    Lim, Young Kyung
    ACTA ONCOLOGICA, 2018, 57 (10) : 1359 - 1366
  • [39] Feasibility study of patient-specific four-dimensional in vivo tracking system for high-dose-rate brachytherapy: Experimental evaluation
    Oh, Geon
    Lee, Jeongshim
    Kim, Hunjung
    Cheon, Wonjoong
    Shin, Dong-seok
    Seo, Jaehyeon
    Sung, Jiwon
    Shin, Dongho
    Yoon, Myonggeun
    Chung, Jin-beom
    Lee, Boram
    MEDICAL PHYSICS, 2025,
  • [40] On the evaluation of edgeless diode detectors for patient-specific QA in high-dose stereotactic radiosurgery
    De Martin, Elena
    Alhujaili, Sultan
    Fumagalli, Maria Luisa
    Ghielmetti, Francesco
    Marchetti, Marcello
    Gallo, Pasqualina
    Aquino, Domenico
    Padelli, Francesco
    Davis, Jeremy
    Alnaghy, Saree
    Carrara, Mauro
    Fariselli, Laura
    Rosenfeld, Anatoly B.
    Petasecca, Marco
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 89 : 20 - 28