A systematic study on the strip-line readout method for SiPM-based TOF PET

被引:0
作者
Kim, Heejong [1 ]
Hua, Yuexuan [2 ]
Lyu, Xudong [3 ]
Xu, Feng [3 ]
Wang, Yunbo [3 ]
Lyu, Jine [2 ]
Eclov, Neville [1 ]
Chen, Chin-Tu [1 ]
Xie, Qingguo [2 ]
Kao, Chien-Min [1 ]
机构
[1] Univ Chicago, Dept Radiol, Chicago, IL 60637 USA
[2] Huazhong Univ Sci & Technol, Biomed Engn Dept, Wuhan, Hubei, Peoples R China
[3] Huazhong Univ Sci & Technol, Sch Opt & Elect Informat, Wuhan, Hubei, Peoples R China
来源
2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC) | 2017年
基金
中国国家自然科学基金;
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Signal multiplexing for SiPM-based TOF PET to achieve high multiplexing ratio and keep the fast timing performance is a challenging issue. We have developed a multiplexing method using a strip-line; multiple SiPMs are connected to a single micro-strip transmission line, and the position of hit SiPM is determined by the signal arrival time difference on the strip-line. The strip-line method is scalable; high multiplexing ratio can be easily achieved by increasing the number of SiPMs connected on a strip-line. To test the scalable feature of the strip-line readout systematically, we have implemented several strip-lines with different trace lengths and varied the number of SiPMs (Hamamatsu S13361-3050NE) on the strip-lines up to 32. Experimental tests have been conducted to assess the influences of the strip-lines on detector performance: position, energy, and coincidence time resolution. Preliminary results from the study are presented, and our ongoing efforts on strip-line readout for TOF PET are discussed.
引用
收藏
页数:3
相关论文
共 3 条
[1]   A silicon photo-multiplier signal readout using strip-line and waveform sampling for Positron Emission Tomography [J].
Kim, H. ;
Chen, C. -T. ;
Eclov, N. ;
Ronzhin, A. ;
Murat, P. ;
Ramberg, E. ;
Los, S. ;
Kao, C. -M. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 830 :119-129
[2]   A feasibility study of a PET/MRI insert detector using strip-line and waveform sampling data acquisition [J].
Kim, H. ;
Chen, C. -T. ;
Eclov, N. ;
Ronzhin, A. ;
Murat, P. ;
Ramberg, E. ;
Los, S. ;
Wyrwicz, A. M. ;
Li, L. ;
Kao, C. -M. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 784 :557-564
[3]   Application of the DRS chip for fast waveform digitizing [J].
Ritt, Stefan ;
Dinapoli, Roberto ;
Hartmann, Ueli .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 623 (01) :486-488