Beta versus gamma dosimetry close to Ir-192 brachytherapy sources

被引:59
作者
Baltas, D [1 ]
Karaiskos, P
Papagiannis, P
Sakelliou, L
Loeffler, E
Zamboglou, N
机构
[1] Klinikum Offenbach, Dept Med Phys & Engn, Strahlenkin, D-63069 Offenbach, Germany
[2] Natl Tech Univ Athens, Inst Commun & Comp Syst, GR-15773 Athens, Greece
[3] Univ Athens, Dept Phys, Nucl & Particle Phys Sect, GR-15771 Athens, Greece
[4] Nucletron BV, NL-3905 TH Veenendaal, Netherlands
关键词
beta particles; dosimetry; endobronchial brachytherapy; intravascular brachytherapy; Ir-192; Monte Carlo simulation;
D O I
10.1118/1.1395038
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The relative importance of the dose rate component owing to the beta spectrum emitted by Ir-192 brachytherapy sources at the short radial distances of interest in intravascular and endobronchial applications is investigated. Separate dosimetric calculations, using Monte Carlo simulations, were performed for the gamma and beta dose rate components of an Ir-192 ideal point source as weld as real Ir-192 source designs used in clinical practice including wire and seed sources and both Nucletron and Varian, old and new, high dose rate (HDR) source designs. A significant dose rate enhancement due to the beta spectrum emitted by Ir-192, greater than 50% for radial distances r < 2 mm, was observed for an ideal point source. For real source designs, however, the magnitude of this enhancement was found to depend strongly on the sources' geometric as well as compositional details of the active core and encapsulation. A detectable effect was found for the majority of the investigated sources at radial distances less than I nim, but overall findings suggest that the contribution of beta particles is not significant in Ir-192 clinical intravascular applications that are currently carried out. However, since treatment of vessels with smaller diameters, in the future, may lead to the development of Ir-192 sources and catheters of reduced diameters, the potential effect of the beta spectrum in terms of dose enhancement to tissues in close proximity to Ir-192 sources should not be ignored. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:1875 / 1882
页数:8
相关论文
共 39 条
[1]   Monte Carlo dosimetry of a new 192Ir high dose rate brachytherapy source [J].
Angelopoulos, A ;
Baras, P ;
Sakelliou, L ;
Karaiskos, P ;
Sandilos, P .
MEDICAL PHYSICS, 2000, 27 (11) :2521-2527
[2]   ACCURATE MONTE-CARLO CALCULATIONS OF THE COMBINED ATTENUATION AND BUILDUP FACTORS, FOR ENERGIES (20-1500 KEV) AND DISTANCES (0-10 CM) RELEVANT IN BRACHYTHERAPY [J].
ANGELOPOULOS, A ;
PERRIS, A ;
SAKELLARIOU, K ;
SAKELLIOU, L ;
SARIGIANNIS, K ;
ZARRIS, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (06) :763-778
[3]   Monte Carlo calculation of dose rate distributions around Ir-192 wires [J].
Ballester, F ;
Hernandez, C ;
PerezCalatayud, J ;
Lliso, F .
MEDICAL PHYSICS, 1997, 24 (08) :1221-1228
[4]   Application of the Monte Carlo integration (MCI) method for calculation of the anisotropy of 192Ir brachytherapy sources [J].
Baltas, D ;
Giannouli, S ;
Garbi, A ;
Diakonos, F ;
Geramani, K ;
Ioannidis, GT ;
Tsalpatouros, A ;
Uzunoglu, N ;
Kolotas, C ;
Zamboglou, N .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (06) :1783-1801
[5]   Comparison of calibration procedures for 192IR high-dose-rate brachytherapy sources [J].
Baltas, D ;
Geramani, K ;
Ioannidis, GT ;
Hierholz, K ;
Rogge, B ;
Kolotas, C ;
Müller-Sievers, K ;
Milickovic, N ;
Kober, B ;
Zamboglou, N .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (03) :653-661
[6]  
Borg J, 1999, PIRS629R NRC
[7]  
Briesmeister J.F., 1997, LA12625M
[8]   A SHORT ATLAS OF BETA-RAY SPECTRA [J].
CROSS, WG ;
ING, H ;
FREEDMAN, N .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (11) :1251-1260
[9]   Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source [J].
Daskalov, GM ;
Loffler, E ;
Williamson, JF .
MEDICAL PHYSICS, 1998, 25 (11) :2200-2208
[10]  
GLASGOW G P, 1979, Medical Physics (Woodbury), V6, P49, DOI 10.1118/1.594551