Evaluation of the Contact Area in Total Knee Arthroplasty Designed for Deep Knee Flexion

被引:6
作者
Triwardono, Joko [1 ,2 ]
Supriadi, Sugeng [1 ]
Whulanza, Yudan [1 ]
Saragih, Agung Shamsuddin [1 ]
Novalianita, Deva Ariana [1 ]
Utomo, Muhammad Satrio [2 ]
Kartika, Ika [2 ]
机构
[1] Univ Indonesia, Fac Engn, Dept Mech Engn, Kampus Ui Depok, Depok 16424, Indonesia
[2] Natl Res & Innovat Agcy, Banten 15314, Indonesia
关键词
Dislocation; Finite element analysis; Hyperflexion; Polyethylene; Subluxation; Total knee arthroplasty; FINITE-ELEMENT-ANALYSIS; POLYETHYLENE; JOINT; KINEMATICS; THICKNESS; FAILURE; REPLACEMENT; REVISION; FRACTURE; RISK;
D O I
10.14716/ijtech.v12i6.5193
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Total knee arthroplasty (TKA) implants are becoming an interesting subject in implant design research and development activities due to their complexity. They should be able to facilitate knee movement while supporting body weight during daily usage. Meanwhile, incidents such as hyperflexion in TKA implants outside their designated configuration can lead to subluxation and dislocation in this study, a polyethylene component of a posterior-stabilized right knee joint implant was developed to facilitate a high range of motion (ROM). Finite element analysis (FEA) was used to analyze the contact area on the polyethylene component. FEA was used to simulate weightbearing conditions at 0 degrees, 30 degrees, 60 degrees, 90 degrees, 120 degrees, and 150 degrees of knee flexion. The modified polyethylene component resulted in better performance in terms of contact area, especially at 120 degrees of knee flexion. The two dominant contact areas on the polyethylene component were 733 mm(2) at 0 degrees of knee flexion and 576 mm(2) at 120 degrees of knee flexion. Furthermore, the current design of the polyethylene component can maintain a contact area of 65 mm(2) at 150 degrees of knee flexion. The current design is expected to accommodate deep knee flexion movement in daily activities and reduce the possibility of subluxation and dislocation at the polyethylene component during deep knee flexion. In addition, a large contact area can reduce the potential wear on or fracture of the polyethylene component. Finally, the result of FEA was validated using a simulator of knee kinematic motion; there was no indication of subluxation and dislocation at any degree of knee flexion.
引用
收藏
页码:1312 / 1322
页数:11
相关论文
共 33 条
[1]  
AHLBERG A, 1988, CLIN ORTHOP RELAT R, P229
[2]   Finite Element Analysis of Proximal Cement Fixation in Total Hip Arthroplasty [J].
Ahmad, Maizatul Afirah ;
Zulkifli, Nurul Nadhirah Mohamed Elias ;
Shuib, Solehuddin ;
Sulaiman, Shahrul Hisham ;
Abdullah, Abdul Halim .
INTERNATIONAL JOURNAL OF TECHNOLOGY, 2020, 11 (05) :1046-1055
[3]  
Azam N.M. A., 2018, MATEC WEB C, V225, P2
[4]   PFC knee replacement - Osteolytic failures from extreme polyethylene degradation [J].
Casey, David ;
Cottrell, Jocelyn ;
DiCarlo, Edward ;
Windsor, Russell ;
Wright, Timothy .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (464) :157-163
[5]   Polyethylene post failure in posterior stabilized total knee arthroplasty [J].
Clarke, HD ;
Math, KR ;
Scuderi, GR .
JOURNAL OF ARTHROPLASTY, 2004, 19 (05) :652-657
[6]   Impact of Polyethylene Thickness on Clinical Outcomes and Survivorship in Medial Mobile-Bearing Unicondylar Knee Arthroplasty [J].
Crawford, David A. ;
Lapsley, Lauren ;
Hurst, Jason M. ;
Morris, Michael J. ;
Lombardi, Adolph V., Jr. ;
Berend, Keith R. .
JOURNAL OF ARTHROPLASTY, 2021, 36 (07) :2440-2444
[7]   The Effect of Polyethylene Liner Thickness on Patient Outcomes and Failure After Primary Total Knee Arthroplasty [J].
Garceau, Simon P. ;
Warschawski, Yaniv S. ;
Tang, Alex ;
Sanders, Ethan B. ;
Schwarzkopf, Ran M. ;
Backstein, David J. .
JOURNAL OF ARTHROPLASTY, 2020, 35 (08) :2072-2075
[8]   Kinematics of the knee joint in deep flexion: a radiographic assessment [J].
Hefzy, MS ;
Kelly, BP ;
Cooke, TDV .
MEDICAL ENGINEERING & PHYSICS, 1998, 20 (04) :302-307
[9]   Kinematic alignment produces near-normal knee motion but increases contact stress after total knee arthroplasty: A case study on a single implant design [J].
Ishikawa, Masahiro ;
Kuriyama, Shinichi ;
Ito, Hiromu ;
Furu, Moritoshi ;
Nakamura, Shinichiro ;
Matsuda, Shuichi .
KNEE, 2015, 22 (03) :206-212
[10]   Risk of Bone Fracture in Resurfacing Hip Arthroplasty at Varus and Valgus Implant Placements [J].
Izmin, Nor Aiman Nor ;
Hazwani, Fatin ;
Todo, Mitsugu ;
Abdullah, Abdul Halim .
INTERNATIONAL JOURNAL OF TECHNOLOGY, 2020, 11 (05) :1025-1035