The staircase phenomenon - Implications for monitoring of neuromuscular transmission

被引:62
作者
Kopman, AF [1 ]
Kumar, S [1 ]
Klewicka, MM [1 ]
Neuman, GG [1 ]
机构
[1] St Vincents Hosp & Med Ctr, Dept Anesthesiol, New York, NY 10011 USA
关键词
D O I
10.1097/00000542-200108000-00023
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background, Repeated indirect stimulation enhances the evoked mechanical response of muscle (the staircase phenomenon). There are few data that document the magnitude of this effect in man. Inexpensive acceleromyographic monitors of neuromuscular function are now available. If these units are to be used as scientific tools or clinical monitors, additional information regarding how to achieve proper baseline stabilization and calibration is needed. Methods: Anesthesia was induced and maintained with nitrous oxide, propofol, and an opioid. Tracheal intubation or laryngeal mask insertion was accomplished without muscle relaxants. Thirty adult patients classified as American Society of Anesthesiologists physical status I or II were divided into groups of 10. The mechanical response of the thumb to supramaximal ulnar nerve stimulation was recorded continuously with an acceleromyograph. Group I had train-of-four stimuli at 15-s intervals for 25 min. Group 2 had single stimuli at 1.0 HZ for 10 min. Group 3 had the same stimuli as group I except that a 50-Hz tetanus of 5 seconds' duration immediately preceded instrument calibration. Results: In group 1, average twitch height (T-1) increased rapidly to 148 +/- 19% (mean SD) of control at 15 min and then more slowly to reach 158 +/- 26% of control at 25 min. The train-of-four fade ratio did not vary with the duration of stimulation. In group 2, T-1 increased to 172 +/- 19% of control after 400 stimuli (6.7 min) and 180 +/- 22% of control at 10 min. In group 3, average T-1 did not decrease below 97.5% or increase above 105 +/- 15% of control at any time. Conclusions: A 5-s, 50-Hz tetanus administered before initial twitch calibration considerably shortens the time required to achieve baseline stability.
引用
收藏
页码:403 / 407
页数:5
相关论文
共 16 条
[1]   REAL-TIME VERSUS SLOW-MOTION TRAIN-OF-4 MONITORING - A THEORY TO EXPLAIN THE INACCURACY OF VISUAL ASSESSMENT [J].
BRULL, SJ ;
SILVERMAN, DG .
ANESTHESIA AND ANALGESIA, 1995, 80 (03) :548-551
[2]  
Engbaek J, 1996, DAN MED BULL, V43, P301
[3]   The effect of stabilization on the onset of neuromuscular block when assessed using accelerometry [J].
Girling, KJ ;
Mahajan, RP .
ANESTHESIA AND ANALGESIA, 1996, 82 (06) :1257-1260
[4]   ENHANCEMENT OF MUSCLE CONTRACTION AFTER TETANUS [J].
Guttman, S. A. ;
Horton, R. G. ;
Wilber, D. T. .
AMERICAN JOURNAL OF PHYSIOLOGY, 1937, 119 (03) :463-473
[5]   THE INFLUENCE OF CHANGES IN HAND TEMPERATURE ON THE INDIRECTLY EVOKED ELECTROMYOGRAM OF THE FIRST DORSAL INTEROSSEOUS MUSCLE [J].
KOPMAN, AF ;
JUSTO, MD ;
MALLHI, MU ;
ABARA, CE ;
NEUMAN, GG .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1995, 42 (12) :1090-1095
[6]   ENHANCEMENT AND DIMINUTION OF MECHANICAL TENSION EVOKED BY STAIRCASE AND BY TETANUS IN RAT MUSCLE [J].
KRARUP, C .
JOURNAL OF PHYSIOLOGY-LONDON, 1981, 311 (FEB) :355-372
[7]   Improving the design of muscle relaxant studies - Stabilization period and tetanic recruitment [J].
Lee, GC ;
Iyengar, S ;
Szenohradszky, J ;
Caldwell, JE ;
Wright, PMC ;
Brown, R ;
Lau, M ;
Luks, A ;
Fisher, DM .
ANESTHESIOLOGY, 1997, 86 (01) :48-54
[8]   THE TOF-GUARD NEUROMUSCULAR-TRANSMISSION MONITOR - A COMPARISON WITH THE MYOGRAPH-2000 [J].
LOAN, PB ;
PAXTON, LD ;
MIRAKHUR, RK ;
CONNOLLY, FM ;
MCCOY, EP .
ANAESTHESIA, 1995, 50 (08) :699-702
[9]   THE INFLUENCE OF THE DURATION OF CONTROL STIMULATION ON THE ONSET AND RECOVERY OF NEUROMUSCULAR BLOCK [J].
MCCOY, EP ;
MIRAKHUR, RK ;
CONNOLLY, FM ;
LOAN, PB .
ANESTHESIA AND ANALGESIA, 1995, 80 (02) :364-367
[10]   Evaluation of neuromuscular and cardiovascular effects of two doses of rapacuronium (ORG 9487) versus mivacurium and succinylcholine [J].
Miguel, R ;
Witkowski, T ;
Nagashima, H ;
Fragen, R ;
Bartkowski, R ;
Foldes, FF ;
Shanks, C .
ANESTHESIOLOGY, 1999, 91 (06) :1648-1654