Comparison of two swept-source optical coherence tomography biometers and a partial coherence interferometer

被引:52
作者
Yang, Chan Min [1 ]
Lim, Dong Hui [1 ,2 ]
Kim, Hyo Jeong [1 ]
Chung, Tae-Young [1 ]
机构
[1] Sungkyunkwan Univ, Samsung Med Ctr, Dept Ophthalmol, Sch Med, Seoul, South Korea
[2] Catholic Univ Korea, Grad Sch, Dept Prevent Med, Seoul, South Korea
关键词
REPEATABILITY; AGREEMENT;
D O I
10.1371/journal.pone.0223114
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose To compare biometry and prediction of postoperative refractive outcomes obtained by two swept-source optical coherence tomography (SS-OCT) biometers (IOLMaster 700 and Argos), and a partial coherence interferometry (IOLMaster ver 5.4) Methods Biometric values were measured using two SS-OCT and PCI device and evaluated against one another. Predictive errors were compared at one month after cataract surgery. Results One hundred forty six eyes were considered. Axial length (AXL) measurements were not successful in 3 eyes measured by IOLMaster 700 and Argos devices, and in 17 eyes measured by IOLMaster ver. 5.4 devices. AXL as measured by Argos showed a tendency to be shorter in long eyes with AXL more than 26.0 mm (p<.001) and to be longer in short eyes with AXL less than 22.5 mm (p=.005). Anterior chamber depth as measured by IOLMaster ver. 5.4 was longer than that measured by the other two SS-OCT devices (vs. IOLMaster 700: p=.003; vs. Argos: p=.006). White-to-white diameter measured using Argos was significantly different measurements obtained using two IOLMaster (p<.001, respectively). The mean absolute postoperative prediction errors were 0.41 +/- 0.31 diopters (D), 0.42 +/- 0.32 D, and 0.35 +/- 0.30 D for IOLMaster ver. 5.4, IOLMaster 700, and Argos, respectively. Conclusion The ocular biometric measurements using three devices showed high agreement. AXL measured by Argos showed a significant difference compared with the measurements from two IOLMaster. ACD was highly correlated between two SS-OCT devices except IOLMaster ver 5.4. LT and CCT values between IOLMaster 700 and Argos were different significantly. SS-OCT devices demonstrated a superior ability to successfully perform measurements compared with PCI device.
引用
收藏
页数:13
相关论文
共 15 条
[1]   Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500 [J].
Akman, Ahmet ;
Asena, Leyla ;
Gungor, Sirel Gur .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2016, 100 (09) :1201-1205
[2]   Agreement and clinical comparison between a new swept-source optical coherence tomography-based optical biometer and an optical low-coherence reflectometry biometer [J].
Arriola-Villalobos, P. ;
Almendral-Gomez, J. ;
Garzon, N. ;
Ruiz-Medrano, J. ;
Fernandez-Perez, C. ;
Martinez-de-la-Casa, J. M. ;
Diaz-Valle, D. .
EYE, 2017, 31 (03) :437-442
[3]   Partial coherence interferometry: A novel approach to biometry in cataract surgery [J].
Drexler, W ;
Findl, O ;
Menapace, R ;
Rainer, G ;
Vass, C ;
Hitzenberger, CK ;
Fercher, AF .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1998, 126 (04) :524-534
[4]   The impact of cataract severity on measurement acquisition with the IOLMaster [J].
Freeman, G ;
Pesudovs, K .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2005, 83 (04) :439-442
[5]   Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis [J].
Haigis, W ;
Lege, B ;
Miller, N ;
Schneider, B .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2000, 238 (09) :765-773
[6]   Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry [J].
Hoffer, Kenneth J. ;
Hoffmann, Peter C. ;
Savini, Giacomo .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (08) :1165-1172
[7]   Protocols for Studies of Intraocular Lens Formula Accuracy [J].
Hoffer, Kenneth J. ;
Aramberri, Jaime ;
Haigis, Wolfgang ;
Olsen, Thomas ;
Savini, Giacomo ;
Shammas, H. John ;
Bentow, Stanley .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2015, 160 (03) :403-405
[8]   Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry [J].
Kunert, Kathleen S. ;
Peter, Monika ;
Blum, Marcus ;
Haigis, Wolfgang ;
Sekundo, Walter ;
Schuetze, Juliane ;
Bueehren, Tobias .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (01) :76-83
[9]   Biometry with a new swept-source optical coherence tomography biometer: Repeatability and agreement with an optical low-coherence reflectometry device [J].
Kurian, Mathew ;
Negalur, Nikhil ;
Das, Sudeep ;
Puttaiah, Narendra K. ;
Haria, Dhawal ;
Tejal, S. J. ;
Thakkar, Maithil M. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (04) :577-581
[10]   Axial Length Measurement Failure Rates with the IOLMaster and Lenstar LS 900 in Eyes with Cataract [J].
McAlinden, Colm ;
Wang, Qinmei ;
Pesudovs, Konrad ;
Yang, Xin ;
Bao, Fangjun ;
Yu, Ayong ;
Lin, Shishi ;
Feng, Yifan ;
Huang, Jinhai .
PLOS ONE, 2015, 10 (06)