共 6 条
High reliability linear drive device for artificial hearts
被引:15
作者:
Ji, Jinghua
[1
]
Zhao, Wenxiang
[1
]
Liu, Guohai
[1
]
Shen, Yue
[1
]
Wang, Fangqun
[1
]
机构:
[1] Jiangsu Univ, Sch Elect & Informat Engn, Zhenjiang 212013, Peoples R China
基金:
中国国家自然科学基金;
关键词:
Inductance - Flux linkage - Magnetic actuators - Reliability - Magnetic circuits - Permanent magnets - Stators - Artificial heart;
D O I:
10.1063/1.3678302
中图分类号:
O59 [应用物理学];
学科分类号:
摘要:
In this paper, a new high reliability linear drive device, termed as stator-permanent-magnet tubular oscillating actuator (SPM-TOA), is proposed for artificial hearts (AHs). The key is to incorporate the concept of two independent phases into this linear AH device, hence achieving high reliability operation. The fault-tolerant teeth are employed to provide the desired decoupling phases in magnetic circuit. Also, as the magnets and the coils are located in the stator, the proposed SPM-TOA takes the definite advantages of robust mover and direct-drive capability. By using the time-stepping finite element method, the electromagnetic characteristics of the proposed SPM-TOA are analyzed, including magnetic field distributions, flux linkages, back-electromotive forces (back-EMFs) self- and mutual inductances, as well as cogging and thrust forces. The results confirm that the proposed SPM-TOA meets the dimension, weight, and force requirements of the AH drive device. (C) 2012 American Institute of Physics. [doi:10.1063/1.3678302]
引用
收藏
页数:3
相关论文