First report of a late type III endoleak from fabric tears of a Zenith stent graft

被引:39
作者
Wanhainen, Anders [1 ]
Nyman, Rickard [2 ]
Eriksson, Mats-Ola [2 ]
Bjorck, Martin [1 ]
机构
[1] Univ Uppsala Hosp, Dept Surg, Vasc Surg Sect, SE-37185 Uppsala, Sweden
[2] Univ Uppsala Hosp, Dept Radiol, Sect Intervent Radiol, SE-37185 Uppsala, Sweden
关键词
D O I
10.1016/j.jvs.2008.03.047
中图分类号
R61 [外科手术学];
学科分类号
摘要
We report a case of a late type III endolcak front a hole in the fabric of the main body of a Zenith bifurcated endograft 7 years after implantation. Abdominal pain and a rapidly expanding aneurysm were eventually followed by rupture. The defect was detected at open surgery, whereas no evidence of endoleak was found at preoperative computed tomography (CT) or angiogram. The defect was repaired by a relining procedure with all Excluder stent graft. The patient, however, died 3 weeks after admission.
引用
收藏
页码:723 / 726
页数:4
相关论文
共 13 条
[1]  
Biebl Matthias, 2005, Vasc Endovascular Surg, V39, P267, DOI 10.1177/153857440503900308
[2]   Endoleaks after endovascular aneurysm repair lead to nonuniform intra-aneurysm sac pressure [J].
Dias, Nuno V. ;
Ivancev, Krassi ;
Resch, Timothy A. ;
Malina, Martin ;
Sonesson, Bjorn .
JOURNAL OF VASCULAR SURGERY, 2007, 46 (02) :197-203
[3]   Comparison of endovascular aneurysm repair with open repair in patients with abdominal aortic aneurysm (EVAR trial 1), 30-day operative mortality results: randomised controlled trial [J].
Greenhalgh, RM ;
Brown, LC ;
Kwong, GPS ;
Powell, JT ;
Thompson, SG .
LANCET, 2004, 364 (9437) :843-848
[4]   Incidence and risk factors of late rupture, conversion, and death after endovascular repair of infrarenal aortic aneurysms: The EUROSTAR experience [J].
Harris, PL ;
Vallabhaneni, SR ;
Desgranges, P ;
Bacquemin, JP ;
van Marrewijk, C ;
Laheij, RJF .
JOURNAL OF VASCULAR SURGERY, 2000, 32 (04) :739-749
[5]   Long-term outcome and reintervention after endovascular abdominal aortic aneurysm repair using the Zenith stent graft [J].
Hiramoto, Jade S. ;
Reilly, Linda M. ;
Schneider, Darren B. ;
Sivamurthy, Nayan ;
Rapp, Joseph H. ;
Chuter, Timothy A. M. .
JOURNAL OF VASCULAR SURGERY, 2007, 45 (03) :461-465
[6]   Late type III endoleak from graft erosion of an Excluder stent graft: A case report [J].
Lee, W. Anthony ;
Huber, Thomas S. ;
Seeger, James M. .
JOURNAL OF VASCULAR SURGERY, 2006, 44 (01) :183-185
[7]  
Parodi J C, 1991, Ann Vasc Surg, V5, P491, DOI 10.1007/BF02015271
[8]   Fabric tears as a new cause of type III endoleak with Ancure endograft [J].
Teutelink, A ;
van der Laan, MJ ;
Milner, R ;
Blankensteijn, JD .
JOURNAL OF VASCULAR SURGERY, 2003, 38 (04) :843-846
[9]   Type III endoleak caused by fabric tear of a zenith endograft after low-pressure balloon modeling [J].
van der Vliet, JA .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2005, 16 (07) :1042-1044
[10]  
Volodos N L, 1991, Vasa Suppl, V33, P93