Patient doses from barium meal and barium enema examinations and potential for reduction through proper set-up of equipment

被引:11
作者
Yakoumakis, E
Tsalafoutas, IA
Sandilos, P
Koulentianos, H
Kasfiki, A
Vlahos, L
Proukakis, C
机构
[1] Univ Athens, Sch Med, Dept Med Phys, Athens 11527, Greece
[2] Konstantopoulio Agia Olga Hosp, Dept Radiol, Athens 14233, Greece
[3] Areteion Hosp, Dept Radiol, Athens 11528, Greece
[4] Konstantopoulio Agia Olga Hosp, Med Phys Unit, Athens 14233, Greece
关键词
D O I
10.1259/bjr.72.854.10365069
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Patient doses for barium meal and barium enema examinations, performed at two Greek hospitals, were measured using a dose-area product meter. The results were analysed to obtain the contributions of fluoroscopy and radiography to the dose as well as a number of other dose related parameters for each examination. The doses observed are within the range of values reported by other authors and comply with the dose reference levels (DRLs), proposed from relevant surveys in the UK and The Netherlands. However, comparison between the two hospitals revealed significant differences in the contributions to dose from the various parts of the examinations. To determine the reasons for these differences, measurements of dose related parameters were made using a Plexiglas phantom and standard clinical X-ray machine settings. Factors contributing to increased dose delivery were determined and recommendations have been made concerning ways in which doses might be reduced in each hospital, without degradation of the diagnostic quality of these examinations.
引用
收藏
页码:173 / 178
页数:6
相关论文
共 14 条
  • [1] ESTIMATION OF DOSES TO PATIENTS FROM COMPLEX CONVENTIONAL X-RAY EXAMINATIONS
    CALZADO, A
    VANO, E
    MORAN, P
    CASTELLOTE, C
    RUIZ, S
    GONZALEZ, L
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (762) : 539 - 546
  • [2] Dosimetry Working Party of the Institute of physical sciences in medicine, 1992, NAT PROT PAT DOS MEA
  • [3] RADIATION-DOSE RECEIVED BY PATIENTS DURING BARIUM MEAL INVESTIGATIONS UNDER AUTOMATIC BRIGHTNESS CONTROL
    FAULKNER, K
    BRAMALL, G
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1985, 58 (685) : 31 - 34
  • [4] Patient dose due to colon examination: Dose assessment and results from a survey in the Netherlands
    Geleijns, J
    Broerse, JJ
    Shaw, MPC
    Schultz, FW
    Teeuwisse, W
    vanUnnik, JG
    Zoetelief, J
    [J]. RADIOLOGY, 1997, 204 (02) : 553 - 559
  • [5] ESTIMATION OF EFFECTIVE DOSE FROM DOSE-AREA PRODUCT MEASUREMENTS FOR BARIUM MEALS AND BARIUM ENEMAS
    HART, D
    WALL, BF
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (797) : 485 - 489
  • [6] Hart D, 1993, NRPBR262
  • [7] HILES PA, 1996, 32 IPEMB, P1
  • [8] ICRP, 1991, ANN ICRP, V21
  • [9] Jones D G., 1985, NRPBR186
  • [10] LEHERON JC, 1992, PHYS MED BIOL, V37, P2117, DOI 10.1088/0031-9155/37/11/008