Histomorphometric assessments of peri-implant bone around Ti-Nb-Sn alloy implants with low Young's modulus

被引:7
|
作者
Shiraishi, Naru [1 ,2 ]
Masumoto, Hiroshi [3 ]
Takahashi, Kenta [1 ]
Tenkumo, Taichi [4 ]
Anada, Takahisa [5 ]
Suzuki, Osamu [6 ]
Ogawa, Toru [1 ]
Sasaki, Keiichi [1 ]
机构
[1] Tohoku Univ, Grad Sch Dent, Div Adv Prosthet Dent, Aoba Ku, 4-1 Seiryo Machi, Sendai, Miyagi 9808575, Japan
[2] Niigata Univ, Div Dysphagia Rehabil, Grad Sch Med & Dent Sci, Chuo Ku, 2-5274 Gakkocho Dori, Niigata 9518514, Japan
[3] Tohoku Univ, Frontier Res Inst Interdisciplinary Sci, Aoba Ku, Aramaki Aza Aoba 6-3, Sendai, Miyagi 9808578, Japan
[4] Tohoku Univ, Lab Redox Regulat, Grad Sch Dent, Aoba Ku, 4-1 Seiryo Machi, Sendai, Miyagi 9808575, Japan
[5] Kyushu Univ, Inst Mat Chem & Engn, Nishi Ku, CE41 744 Motooka, Fukuoka 8190395, Japan
[6] Tohoku Univ, Grad Sch Dent, Div Craniofacial Funct Engn, Aoba Ku, 4-1 Seiryo Machi, Sendai, Miyagi 9808575, Japan
关键词
Ti-Nb-Sn; Implant; Low Young's modulus; FINITE-ELEMENT-ANALYSIS; FIXED PARTIAL DENTURE; BIOMECHANICAL EVALUATION; MANDIBULAR BONE; DENTAL IMPLANTS; CONFIGURATIONS; INTEGRATION; INTERFACE; TITANIUM; BEHAVIOR;
D O I
10.4012/dmj.2018-376
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Many beta-Ti alloys have been developed for, and used in, medical devices because of the corrosion resistance, biocompatibility, and exceptionally low Young's modulus. The aim of the present study was to investigate the histomorphometric aspects of peri-implant bone around Ti-Nb-Sn alloy implants and compare them with those in the case of commercially pure Ti (Ti). Fluorescent morphological observations of ST-2 cells on the substrate were performed and bone morphogenesis around implants in rat femur was evaluated. There was no difference between the cell morphology on Ti and those on the Ti-Nb-Sn alloy during observation for 24 h. A comparison of the Ti-Nb-Sn alloy implant and the Ti implant showed no significant differences between the bone-to-implant contact ratios or the bone fractions. These results suggest that the biological adaptations with Ti-Nb-Sn implants during a healing period are similar to those with Ti. Ti-Nb-Sn is therefore suitable for use in dental implants.
引用
收藏
页码:148 / 153
页数:6
相关论文
共 50 条
  • [41] Peri-implant bone density around implants of different lengths: A 3-year follow-up of a randomized clinical trial
    Sahrmann, Philipp
    Schoen, Patrizia
    Naenni, Nadja
    Jung, Ronald
    Attin, Thomas
    Schmidlin, Patrick R.
    JOURNAL OF CLINICAL PERIODONTOLOGY, 2017, 44 (07) : 762 - 768
  • [42] Biological properties of a novel β-Ti alloy with a low young's modulus subjected to cold rolling
    Kuczynska-Zemla, D.
    Kijenska-Gawronska, E.
    Chlanda, A.
    Sotniczuk, A.
    Pisarek, M.
    Topolski, K.
    Swieszkowski, W.
    Garbacz, H.
    APPLIED SURFACE SCIENCE, 2020, 511
  • [43] Histological evaluation of the peri-implant bone around plasma-sprayed non-submerged titanium implants retrieved from man: A report of two cases
    Piattelli, A
    Scarano, A
    Vaia, E
    Matarasso, S
    BIOMATERIALS, 1996, 17 (23) : 2219 - 2224
  • [44] Metastable β-type Ti-30Nb-1Mo-4Sn Alloy with Ultralow Young's Modulus and High Strength
    Hu, Liang
    Guo, Shun
    Meng, Qingkun
    Zhao, Xinqing
    METALLURGICAL AND MATERIALS TRANSACTIONS A-PHYSICAL METALLURGY AND MATERIALS SCIENCE, 2014, 45A (02): : 547 - 550
  • [45] Evaluation of peri-implant marginal tissues around tissue-level and bone-level implants in patients with a history of chronic periodontitis
    Vianna, Tiago T.
    Taiete, Tiago
    Casarin, Renato C. V.
    Giorgi, Maria C. C.
    Aguiar, Flavio Henrique B.
    Silverio, Karina G.
    Nociti Junior, Francisco H.
    Sallum, Enilson A.
    Casati, Marcio Z.
    JOURNAL OF CLINICAL PERIODONTOLOGY, 2018, 45 (10) : 1255 - 1265
  • [46] Crystal orientation dependence of Young's modulus in Ti-Nb-based β-titanium alloy
    Cui WenFang
    Guo AiHong
    Zhou Lian
    Liu ChunMing
    SCIENCE CHINA-TECHNOLOGICAL SCIENCES, 2010, 53 (06) : 1513 - 1519
  • [47] Peri-implant bone response to retrieved human zirconia oral implants after a 4-year loading period: A histologic and histomorphometric evaluation of 22 cases
    Kohal, Ralf-Joachim
    Schwindling, Franz Sebastian
    Baechle, Maria
    Spies, Benedikt Christopher
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2016, 104 (08) : 1622 - 1631
  • [48] Development of new β/α"-Ti-Nb-Zr biocompatible coating with low Young's modulus and high toughness for medical applications
    Frutos, E.
    Karlik, M.
    Jimenez, J. A.
    Langhansova, H.
    Lieskovska, J.
    Polcar, T.
    MATERIALS & DESIGN, 2018, 142 : 44 - 55
  • [49] Diagnostic accuracy of imaging examinations for peri-implant bone defects around titanium and zirconium dioxide implants: A systematic review and meta-analysis
    Chagas, Mariana Murai
    Kobayashi-Velasco, Solange
    Gimenez, Thais
    Paraiso Cavalcanti, Marcelo Gusmao
    IMAGING SCIENCE IN DENTISTRY, 2021, 51 (04) : 363 - 372
  • [50] Texture Development in Cold Deformed and Recrystallized Ti-30Nb-4Sn Alloy and Its Effects on Hardness and Young's Modulus
    Fanton, Leonardo
    de Lima, Nelson Batista
    Franca Hayama, Alexandra de Oliveira
    Caram, Rubens
    Fogagnolo, Joao Batista
    ADVANCED ENGINEERING MATERIALS, 2017, 19 (06)