共 50 条
Aptamer Biosensors for Label-Free Colorimetric Detection of Human IgE Based on Polydiacetylene (PDA) Supramolecules
被引:10
|作者:
Kim, Jun Pyo
[2
]
Park, Cheol Hee
[2
]
Sim, Sang Jun
[1
]
机构:
[1] Korea Univ, Dept Chem & Biol Engn, Seoul 136713, South Korea
[2] Sungkyunkwan Univ, Dept Chem Engn, Suwon 440746, South Korea
关键词:
Polydiacetylene;
Biosensor;
Aptamer;
IgE;
IN-VITRO SELECTION;
VESICLES;
CHROMISM;
COLOR;
BIND;
D O I:
10.1166/jnn.2011.3664
中图分类号:
O6 [化学];
学科分类号:
0703 ;
摘要:
In this study, we demonstrate an aptamer-based biosensor (apta-biosensor) using PDA liposomes for label-free detection of allergy diagnosis by hIgE detection. In order to detect the target hIgE, the surface of PDA liposome were functionalized with hIgE antibody and anti-hIgE aptamer as a receptor, and the target hIgE onto the receptors was detected by the change of fluorescence signal. The hIgE antibody-modified PDA liposome biosensor had a serious problem that the immune reaction between receptor and target could not powerfully affect the change of florescence signal on PDA liposorne. In order to solve this problem, the anti-hIgE aptamer which was far smaller than whole antibody was introduced as the receptor for the PDA liposome system. An aptamer-based PDA liposome biosensor was able to measure a quantity of target protein with various concentrations and at this time the detection limit was 141 ng/mL of the hIgE concentration. These results enabled diagnosis of allergy disease by an aptamer-based PDA liposome biosensor because real allergic patients showed high concentration of hIgE in serum (greater than 290 ng/mL). Therefore, we suggest that aptamer-modified PDA supramolecules as promising candidates for development of label-free colorimetric biosensors.
引用
收藏
页码:4269 / 4274
页数:6
相关论文