Postmortem In Situ Micro-CT Evaluation of Coronary Stent Fracture

被引:5
作者
Foerst, Jason [1 ]
Ball, Timothy [1 ]
Kaplan, Aaron V. [1 ]
机构
[1] Dartmouth Hitchcock Med Ctr, Cardiol Sect, Dartmouth Med Sch, Lebanon, NH 03756 USA
关键词
angiography coronary (ANGO); complications adult cath/intervention (COMP); imaging (IMAG); (CT/MR); SIROLIMUS-ELUTING STENT; CLINICAL IMPACT; IMPLANTATION;
D O I
10.1002/ccd.22498
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Coronary stent fracture has been postulated as an important mechanism for in-stent restenosis and late-stent thrombosis. We have developed a nondestructive microcomputed tomography (micro-CT) technique to image en bloc stented arterial segments. Case Report: A 43-year-old man died of sepsis 5 months following placement of a paclitaxel-eluting stent (PES) in the left circumflex coronary artery to treat in-stent restenosis of a sirolimus-eluting stent (SES). Micro-CT three-dimensional images of the overlapping stents were notable for numerous connecting element fractures along the length of the SES. No fractures were observed in the PES. The greatest area of luminal narrowing is related to the region of most extensive stent fracture. Micro-CT of en bloc specimens provides high resolution (16 pm), three-dimensional images allowing for visualization of the stented arterial segment in a nondestructive manner. (c) 2010 Wiley-Liss, Inc.
引用
收藏
页码:527 / 531
页数:5
相关论文
共 8 条
  • [1] Allie DE., 2004, Endovasc. Today, V7, P22
  • [2] Incidence and clinical impact of coronary stent fracture after sirolimus-eluting stent implantation
    Aoki, Jiro
    Nakazawa, Gaku
    Tanabe, Kengo
    Hoye, Angela
    Yamamoto, Hirosada
    Nakayama, Tomohiro
    Onuma, Yoshinobu
    Higashikuni, Yasutomi
    Otsuki, Syuji
    Yagishita, Atsuhiko
    Yachi, Sen
    Nakajima, Hiroyoshi
    Hara, Kazuhiro
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2007, 69 (03) : 380 - 386
  • [3] Stent fracture associated with drug-eluting stents: Clinical characteristics and implications
    Lee, Michael S.
    Jurewitz, Daniel
    Aragon, Joseph
    Forrester, James
    Makkar, Raj R.
    Kar, Saibal
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2007, 69 (03) : 387 - 394
  • [4] Very late stent fracture associated with a sirolimus-eluting stent
    Leong, D. P.
    Dundon, B. K.
    Puri, R.
    Yeend, R. A. S.
    [J]. HEART LUNG AND CIRCULATION, 2008, 17 (05) : 426 - 428
  • [5] Incidence and Predictors of Drug-Eluting Stent Fracture in Human Coronary Artery A Pathologic Analysis
    Nakazawa, Gaku
    Finn, Aloke V.
    Vorpahl, Marc
    Ladich, Elena
    Kutys, Robert
    Balazs, Isidora
    Kolodgie, Frank D.
    Virmani, Renu
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 54 (21) : 1924 - 1931
  • [6] A Qualitative and Quantitative Angiographic Analysis of Stent Fracture Late Following Sirolimus-Eluting Stent Implantation
    Popma, Jeffrey J.
    Tiroch, Klaus
    Almonacid, Alexandra
    Cohen, Sidney
    Kandzari, David E.
    Leon, Martin B.
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2009, 103 (07) : 923 - 929
  • [7] Prevalence and clinical impact of stent fractures after femoropopliteal stenting
    Scheinert, D
    Scheinert, S
    Sax, J
    Piorkowski, C
    Bräunlich, S
    Ulrich, M
    Biamino, G
    Schmidt, A
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 45 (02) : 312 - 315
  • [8] Stent fracture, an incidental finding or a significant marker of clinical in-stent restenosis?
    Shaikh, Fareed
    Maddikunta, Rajesh
    Djelmami-Hani, Mohamed
    Solis, Joaquin
    Allaqaband, Suhail
    Bajwa, Tanvir
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2008, 71 (05) : 614 - 618