Novel hexacyanoferrate (III)-modified carbon electrodes: Application in miniaturized biosensors with potential for in vivo glucose sensing

被引:44
作者
Jaffari, SA [1 ]
Pickup, JC [1 ]
机构
[1] GUYS HOSP, UMDS, DIV CHEM PATHOL, LONDON SE1 9RT, ENGLAND
关键词
miniature glucose biosensor; glucose oxidase; 4-acetamidophenol; acetaminophen; paracetamol; ascorbic acid; modified carbon; potassium hexacyanoferrate (III);
D O I
10.1016/0956-5663(96)82339-6
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
We adapted a new technology for the modification of carbon base electrodes for use as probe-type, potentially-implantable glucose sensors. Carbon rods (diameter 0.3 mm) were modified by repeated potential cycling in 0.1 M potassium hexacyanoferrate (III). The modified-carbon electrodes were sealed in plastic pipette tips with an exposed reaction area where glucose oxidase was immobilized using glutaraldehyde. An outer membrane of Nafion, followed by 15% (w/v) polyurethane, was applied over the enzyme layer. The miniature modified-carbon glucose sensors displayed a sensitivity to glucose in phosphate-buffered saline of 91.4+/-19 nA/mM (mean+/-SEM) and a linear range up to 5.3+/-1 mM glucose when operated at 750 mV versus a silver/silver chloride reference. Corresponding, unmodified-carbon based glucose sensors displayed a lower sensitivity of 20.7+/-3 nA/mM with a linear range up to 3.8+/-0.5 mM. The modified-carbon glucose sensors responded to glucose when operated in plasma but with a reduced sensitivity compared with that in buffered saline. Glucose sensors displayed good stability for up to 6.5 days during continuous operation in 5 mM buffered glucose solution. Interference from ascorbate and 4-acetamidophenol at both physiological and pharmacological ranges was significantly lower at the modified-carbon base electrodes than that at the unmodified-carbon base electrodes. Also, the relatively large effect of ascorbate and 4-acetamidophenol at the unmodified-carbon base electrode was reduced considerably when the base electrode was coated with glucose oxidase, Nafion and polyurethane membranes. (C) 1996 Elsevier Science Limited
引用
收藏
页码:1167 / 1175
页数:9
相关论文
共 23 条
[1]   DESIGN AND INVITRO STUDIES OF A NEEDLE-TYPE GLUCOSE SENSOR FOR SUBCUTANEOUS MONITORING [J].
BINDRA, DS ;
ZHANG, YN ;
WILSON, GS ;
STERNBERG, R ;
THEVENOT, DR ;
MOATTI, D ;
REACH, G .
ANALYTICAL CHEMISTRY, 1991, 63 (17) :1692-1696
[2]   FERROCENE-MEDIATED ENZYME ELECTRODE FOR AMPEROMETRIC DETERMINATION OF GLUCOSE [J].
CASS, AEG ;
DAVIS, G ;
FRANCIS, GD ;
HILL, HAO ;
ASTON, WJ ;
HIGGINS, IJ ;
PLOTKIN, EV ;
SCOTT, LDL ;
TURNER, APF .
ANALYTICAL CHEMISTRY, 1984, 56 (04) :667-671
[3]  
CSOREGI E, 1994, ANAL CHEM, V66, P3131
[4]   EVALUATION OF NAFION AS MEDIA FOR GLUCOSE-OXIDASE IMMOBILIZATION FOR THE DEVELOPMENT OF AN AMPEROMETRIC GLUCOSE BIOSENSOR [J].
FORTIER, G ;
VAILLANCOURT, M ;
BELANGER, D .
ELECTROANALYSIS, 1992, 4 (03) :275-283
[5]   IMMOBILIZATION OF GLUCOSE-OXIDASE IN FERROCENE-MODIFIED PYRROLE POLYMERS [J].
FOULDS, NC ;
LOWE, CR .
ANALYTICAL CHEMISTRY, 1988, 60 (22) :2473-2478
[6]   TWO-DIMENSIONAL ENZYME ELECTRODE SENSOR FOR GLUCOSE [J].
GOUGH, DA ;
LUCISANO, JY ;
TSE, PHS .
ANALYTICAL CHEMISTRY, 1985, 57 (12) :2351-2357
[7]   CROSS-LINKED REDOX GELS CONTAINING GLUCOSE-OXIDASE FOR AMPEROMETRIC BIOSENSOR APPLICATIONS [J].
GREGG, BA ;
HELLER, A .
ANALYTICAL CHEMISTRY, 1990, 62 (03) :258-263
[8]   ROLE OF CARBONYL SITES IN THE ELECTROCHEMICAL RESPONSE OF 2,6-DIAMINOPURIN-8-OL [J].
BODALBHAI, L ;
BRAJTERTOTH, A .
ANALYTICA CHIMICA ACTA, 1991, 243 (01) :33-42
[9]   POLYFERROCENES AS MEDIATORS IN AMPEROMETRIC BIOSENSORS FOR GLUCOSE [J].
HENDRY, SP ;
CARDOSI, MF ;
TURNER, APF ;
NEUSE, EW .
ANALYTICA CHIMICA ACTA, 1993, 281 (03) :453-459
[10]   IMMOBILIZED ENZYME CHEMICALLY MODIFIED ELECTRODE AS AN AMPEROMETRIC SENSOR [J].
IANNIELLO, RM ;
YACYNYCH, AM .
ANALYTICAL CHEMISTRY, 1981, 53 (13) :2090-2095