Determination of Nd3+ Ions in Solution Samples by a Coated Wire Ion-Selective Sensor

被引:27
|
作者
Zamani, Hassan Ali [1 ]
机构
[1] Islamic Azad Univ, Quchan Branch, Dept Appl Chem, Quchan, Iran
关键词
Coated wire electrode; Ion-selective electrode; PVC membrane; Sensor; Potentiometry; PVC-MEMBRANE-ELECTRODE; SEDIMENT SAMPLES; WASTE-WATER; ERBIUM-ION; CONSTRUCTION; SOIL; FABRICATION; BASE; CARRIER;
D O I
10.1155/2012/360769
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A new coated wire electrode (CWE) using 5-(methylsulfanyl)-3-phenyl-1H-1,2,4-triazole (MPT) as an ionophore has been developed as a neodymium ion-selective sensor. The sensor exhibits Nernstian response for the Nd3+ ions in the concentration range of 1.0x10(-6)-1.0x10(-2) M with detection limit of 3.7x10(-7) M. It displays a Nernstian slope of 20.2 +/- 0.2 mV/decade in the pH range of 2.7-8.1. The proposed sensor also exhibits a fast response time of similar to 5 s. The sensor revealed high selectivity with respect to all common alkali, alkaline earth, transition and heavy metal ions, including members of the lanthanide family other than Nd3+. The electrode was used as an indicator electrode in the potentiometric titration of Nd(III) ions with EDTA. The electrode was also employed for the determination of the Nd3+ ions concentration in water solution samples.
引用
收藏
页码:83 / 88
页数:6
相关论文