A new pH-ISFET based dissolved oxygen sensor by employing electrolysis of oxygen

被引:43
|
作者
Sohn, BK [1 ]
Kim, CS [1 ]
机构
[1] KYUNGPOOK NATL UNIV,SENSOR TECHNOL RES CTR,PUK KU,TAEGU 702701,SOUTH KOREA
关键词
dissolved oxygen sensor; pH-ISFET; Clark type sensor; oxygen electrolysis; working electrode;
D O I
10.1016/S0925-4005(97)80017-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new dissolved oxygen sensor based on a pH-ISFET is discussed. A working electrode surrounding a pH-sensing gate of the pH-ISFET electrolytes dissolved oxygen, resulting in a corresponding pH change near the pH-sensing gate. The pH-ISFET is expected to determine dissolved oxygen concentration by detecting this pH change. The results suggest that the proposed sensor operated by a combined mechanism of an amperometric sensor (Clark type sensor) and a potentiometric sensor (pH-ISFET) is feasible to determine the content of dissolved oxygen. The dependence of sensor output on several factors, such as buffering capacity, bulk pH and stirring of the solution, is investigated and a comparison between the Clark type sensor and proposed sensor is also discussed.
引用
收藏
页码:435 / 440
页数:6
相关论文
共 50 条
  • [21] A very large integrated pH-ISFET sensor array chip compatible with standard CMOS processes
    Yeow, TCW
    Haskard, MR
    Mulcahy, DE
    Seo, HI
    Kwon, DH
    SENSORS AND ACTUATORS B-CHEMICAL, 1997, 44 (1-3) : 434 - 440
  • [22] Temperature influence on pH-ISFET sensor operating in weak and moderate inversion regime: Model and circuitry
    Naimi, S. E.
    Hajji, B.
    Humenyuk, I.
    Launay, J.
    Temple-Boyer, P.
    SENSORS AND ACTUATORS B-CHEMICAL, 2014, 202 : 1019 - 1027
  • [23] ISFET glucose sensor based on a new principle using the electrolysis of hydrogen peroxide
    Seo, HI
    Kim, CS
    Sohn, BK
    Yeow, T
    Son, MT
    Haskard, M
    SENSORS AND ACTUATORS B-CHEMICAL, 1997, 40 (01) : 1 - 5
  • [24] Development of a light-based dissolved oxygen sensor
    Kalantari, Armin
    Raval, Dhvanit
    Sheaff, Robert J.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2018, 255
  • [25] Fabrication and characterization of micro dissolved oxygen sensor-activated on demand using electrolysis
    Lee, Hyun-Jin
    Kim, Hyoung-Min
    Park, Jae-Hyoung
    Lee, Seung-Ki
    SENSORS AND ACTUATORS B-CHEMICAL, 2017, 241 : 923 - 930
  • [26] Continuous accurate pH measurements of human GI tract using a digital pH-ISFET sensor inside a wireless capsule
    Xu, Fei
    Yan, Guozheng
    Wang, Zhiwu
    Jiang, Pingping
    MEASUREMENT, 2015, 64 : 49 - 56
  • [27] Analysis and Solution on the Correlation between Dissolved Oxygen Sensor and PH Sensor for Water Quality Monitoring
    Liu, Liya
    Zhao, Dean
    Qin, Yun
    INFORMATION TECHNOLOGY APPLICATIONS IN INDUSTRY II, PTS 1-4, 2013, 411-414 : 1445 - +
  • [28] Use of an optical oxygen sensor to measure dissolved oxygen in seawater
    Hiroshi Hasumoto
    Tsuneo Imazu
    Toshiaki Miura
    Kazuhiro Kogure
    Journal of Oceanography, 2006, 62 : 99 - 103
  • [29] Use of an optical oxygen sensor to measure dissolved oxygen in seawater
    Hasumoto, H
    Imazu, T
    Miura, T
    Kogure, K
    JOURNAL OF OCEANOGRAPHY, 2006, 62 (01) : 99 - 103
  • [30] A Dissolved Oxygen Analyzer with an Optical Sensor
    N. K. Zaitsev
    V. I. Dvorkin
    P. V. Melnikov
    A. E. Kozhukhova
    Journal of Analytical Chemistry, 2018, 73 : 102 - 108