A new pH-ISFET based dissolved oxygen sensor by employing electrolysis of oxygen

被引:43
|
作者
Sohn, BK [1 ]
Kim, CS [1 ]
机构
[1] KYUNGPOOK NATL UNIV,SENSOR TECHNOL RES CTR,PUK KU,TAEGU 702701,SOUTH KOREA
关键词
dissolved oxygen sensor; pH-ISFET; Clark type sensor; oxygen electrolysis; working electrode;
D O I
10.1016/S0925-4005(97)80017-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new dissolved oxygen sensor based on a pH-ISFET is discussed. A working electrode surrounding a pH-sensing gate of the pH-ISFET electrolytes dissolved oxygen, resulting in a corresponding pH change near the pH-sensing gate. The pH-ISFET is expected to determine dissolved oxygen concentration by detecting this pH change. The results suggest that the proposed sensor operated by a combined mechanism of an amperometric sensor (Clark type sensor) and a potentiometric sensor (pH-ISFET) is feasible to determine the content of dissolved oxygen. The dependence of sensor output on several factors, such as buffering capacity, bulk pH and stirring of the solution, is investigated and a comparison between the Clark type sensor and proposed sensor is also discussed.
引用
收藏
页码:435 / 440
页数:6
相关论文
共 50 条
  • [1] pH-ISFET based urea biosensor
    Pijanowska, DG
    Torbicz, W
    SENSORS AND ACTUATORS B-CHEMICAL, 1997, 44 (1-3) : 370 - 376
  • [2] pH-ISFET based urea biosensor
    Polish Acad of Sciences, Warsaw, Poland
    Sens Actuators, B Chem, 1 -3 pt 5 (370-376):
  • [4] Study on the hysteresis effect of pH-ISFET based on Beckman Φ™ 110 (Si3N4 gate pH-ISFET)
    Chou, JC
    Tseng, YN
    OPTOELECTRONIC MATERIALS AND DEVICES II, 2000, 4078 : 793 - 800
  • [5] A pH-ISFET based micro sensor system on chip using standard CMOS technology
    Yang, HG
    Sun, HG
    Han, JH
    Wei, JB
    Lin, ZJ
    Xia, SH
    Zhong, H
    Fifth International Workshop on System-on-Chip for Real-Time Applications, Proceedings, 2005, : 180 - 183
  • [6] 64 pH-ISFET averaging array employing global negative current feedback
    Chan, W. P.
    Premanode, B.
    Toumazou, C.
    ELECTRONICS LETTERS, 2009, 45 (11) : 536 - U16
  • [7] A PH-ISFET AND AN INTEGRATED PH-PRESSURE SENSOR WITH BACK-SIDE CONTACTS
    VANDENVLEKKERT, HH
    KLOECK, B
    PRONGUE, D
    BERTHOUD, J
    HU, B
    DEROOIJ, NF
    GILLI, E
    DECROUSAZ, P
    SENSORS AND ACTUATORS, 1988, 14 (02): : 165 - 176
  • [8] Robust Dual Optical Sensor for pH and Dissolved Oxygen
    Wang, Lu
    Jensen, Kim
    Hatzakis, Nikos
    Zhang, Min
    Sorensen, Thomas Just
    ACS SENSORS, 2022, 7 (05) : 1506 - 1513
  • [9] Labview Based Dissolved Oxygen Sensor
    Chu, Fenghong
    Yang, Junjie
    2009 INTERNATIONAL CONFERENCE ON ARTIFICIAL INTELLIGENCE AND COMPUTATIONAL INTELLIGENCE, VOL II, PROCEEDINGS, 2009, : 101 - 103
  • [10] The influence of isothermal annealing on tin oxide thin film for pH-ISFET sensor
    Liao, HK
    Chou, JC
    Chung, WY
    Sun, TP
    Hsiung, SK
    TECHNICAL DIGEST OF THE SEVENTH INTERNATIONAL MEETING ON CHEMICAL SENSORS, 1998, : 497 - 499