Short-term biological variation and diurnal rhythm of cardiac troponin I (Access hs-TnI) in healthy subjects

被引:33
作者
Zaninotto, Martina [1 ]
Padoan, Andrea [1 ,2 ]
Mion, Monica Maria [1 ]
Marinova, Mariela [1 ]
Plebani, Mario [1 ,2 ]
机构
[1] Univ Hosp, Dept Lab Med, Via Giustiniani 2, I-35128 Padua, Italy
[2] Univ Padua, Dept Med, Padua, Italy
关键词
High-sensitivity troponin I; Intra-individual variation; Inter-individual variation; Index of individuality; Circadian variability; Diurnal rhythm; CARDIOVASCULAR EVENTS; MYOCARDIAL-INFARCTION; DIAGNOSIS; VALUES; ASSAY; CTNI; RISK;
D O I
10.1016/j.cca.2020.02.004
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The availability of high-sensitivity cTnI (hs-cTnI) assays has improved the accuracy of cTn measurements at concentrations around and below the 99th percentile, allowing the evaluation of biological variation. Methods: cTnI concentrations have been measured in blood samples of 35 reference subjects collected at time 0 (between 8 and 9 AM) and after 1,2,3 and 7 h using a high-sensitive assay (Access hs-TnI). Repeated measure ANOVA and lognormal transformation followed by Nested ANOVA were used to assess differences in cTnI concentration and to estimate biological variation components, respectively. Circadian variability was modelled by sine-wave functions fitting. Results: At time 0, cTnI concentrations were significantly higher than those measured at other times in overall population, as well as in subjects subdivided by biological sex. The concentrations exhibit a strong circadian variability in males and females, with a predicted interval of around 5.4 h (R-2 0.949 and 0.999 for males and females, respectively). Conclusions: Troponin I demonstrates a diurnal rhythm with decreasing values throughout daytime and the peak concentrations in the morning. The circadian variability is statistically significant, but not relevant from a clinical viewpoint. The intra-individual variation (CVI) is lower than that reported in the literature and the index of individuality lower than 0.6 suggests a scarce value of reference interval.
引用
收藏
页码:163 / 167
页数:5
相关论文
共 32 条
[1]   Harmonization initiatives in the generation, reporting and application of biological variation data [J].
Aarsand, Aasne K. ;
Roraas, Thomas ;
Bartlett, William A. ;
Coskun, Abdurrahman ;
Carobene, Anna ;
Fernandez-Calle, Pilar ;
Jonker, Niels ;
Diaz-Garzon, Jorge ;
Braga, Federica ;
Sandberg, Sverre .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (10) :1629-1636
[2]   Cardiac Troponin Assays: Guide to Understanding Analytical Characteristics and Their Impact on Clinical Care [J].
Apple, Fred S. ;
Sandoval, Yader ;
Jaffe, Allan S. ;
Ordonez-Llanos, Jordi .
CLINICAL CHEMISTRY, 2017, 63 (01) :73-81
[3]   High-Sensitivity Cardiac Troponin I Assay for Early Diagnosis of Acute Myocardial Infarction [J].
Boeddinghaus, Jasper ;
Nestelberger, Thomas ;
Twerenbold, Raphael ;
Koechlin, Luca ;
Meier, Mario ;
Troester, Valentina ;
Wussler, Desiree ;
Badertscher, Patrick ;
Wildi, Karin ;
Puelacher, Christian ;
de Lavallaz, Jeanne du Fay ;
Gimenez, Maria Rubini ;
Zimmermann, Tobias ;
Hafner, Benjamin ;
Potlukova, Eliska ;
Miro, Oscar ;
Javier Martin-Sanchez, F. ;
Keller, Dagmar, I ;
Reichlin, Tobias ;
Mueller, Christian ;
Walter, Joan Elias ;
Strebe, No ;
Kozhuharov, Nikola ;
Freese, Michael ;
Fuenzalida, Carolina ;
Stelzig, Claudia ;
Gualandro, Danielle M. ;
Michou, Eleni ;
Meissner, Kathrin ;
Kulangara, Caroline ;
Shrestha, Samyut ;
Fahrni, Regor ;
Osswald, Stefan ;
Lopez, Beatriz ;
Rodriguez Adrada, Esther ;
Ganovska, Eva ;
Lohrmann, Jens ;
Kloos, Wanda ;
Steude, Jana ;
User, Andreas B. ;
von Eckardstein, Arnold ;
Morawiec, Beata ;
Kawecki, Damian ;
Nowalany-Kozielska, Ewa ;
Muzyk, Piotr ;
Geigy, Nicolas ;
Rentsch, Katharina .
CLINICAL CHEMISTRY, 2019, 65 (07) :893-904
[4]   Distribution of plasma cardiac troponin I values in healthy subjects: pathophysiological considerations [J].
Clerico, Aldo ;
Fortunato, Antonio ;
Ripoli, Andrea ;
Prontera, Concetta ;
Zucchelli, Gian Carlo ;
Emdin, Michele .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2008, 46 (06) :804-808
[5]   Evaluation of analytical performance of immunoassay methods for cTnI and cTnT: From theory to practice [J].
Clerico, Aldo ;
Zaninotto, Martina ;
Padoan, Andrea ;
Masotti, Silvia ;
Musetti, Veronica ;
Prontera, Concetta ;
Ndreu, Rudina ;
Zucchelli, Giancarlo ;
Passino, Claudio ;
Migliardi, Marco ;
Plebani, Mario .
ADVANCES IN CLINICAL CHEMISTRY, VOL 93, 2019, 93 :239-262
[6]   Evaluation of 99th percentile and reference change values of a high-sensitivity cTnI method: A multicenter study [J].
Clerico, Aldo ;
Ripoli, Andrea ;
Masotti, Silvia ;
Musetti, Veronica ;
Aloe, Rosalia ;
Dipalo, Mariella ;
Rizzardi, Sara ;
Dittadi, Ruggero ;
Carrozza, Cinzia ;
Storti, Simona ;
Belloni, Lucia ;
Perrone, Marco ;
Fasano, Tommaso ;
Canovi, Simone ;
Correale, Mario ;
Prontera, Concetta ;
Guiotto, Cristina ;
Cosseddu, Domenico ;
Migliardi, Marco ;
Bernardini, Sergio .
CLINICA CHIMICA ACTA, 2019, 493 :156-161
[7]   Assessment of Access hsTnI 99th percentiles upper reference limits following IFCC recommendations [J].
Di Pietro, Martina ;
Dipalo, Mariella ;
Rocchi, Marco Bruno Luigi ;
Mus, Roberta ;
Avanzini, Paola ;
Gnocchi, Cecilia ;
Anelli, Maria Chiara ;
Aloe, Rosalia .
CLINICA CHIMICA ACTA, 2019, 492 :26-28
[8]   Circadian rhythm of blood cardiac troponin T concentration [J].
Fournier, Stephane ;
Iten, Lea ;
Marques-Vidal, Pedro ;
Boulat, Olivier ;
Bardy, Daniel ;
Beggah, Ahmed ;
Calderara, Rachel ;
Morawiec, Beata ;
Lauriers, Nathalie ;
Monney, Pierre ;
Iglesias, Juan F. ;
Pascale, Patrizio ;
Harbaoui, Brahim ;
Eeckhout, Eric ;
Muller, Olivier .
CLINICAL RESEARCH IN CARDIOLOGY, 2017, 106 (12) :1026-1032
[9]   Reference change values [J].
Fraser, Callum G. .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2012, 50 (05) :807-812
[10]  
FRASER CG, 1989, CRIT REV CL LAB SCI, V27, P409