Investigation of the reliability of semi-automatic annotation by the Geri time-lapse system

被引:0
作者
Vandame, Jessica [1 ]
Fossard, Camille [1 ]
Filali, Meryem [1 ]
Benammar, Achraf [1 ]
Ranga, Stephanie [1 ]
Pirtea, Paul [1 ]
Racowsky, Catherine [1 ]
Ayoubi, Jean-Marc [1 ,2 ]
Poulain, Marine [1 ,2 ]
机构
[1] Hosp Foch, Dept Obstet Gynecol & Reprod Med, Suresnes, France
[2] Univ Versailles, Paris Saclay Univ, UMR BREED 1198, INRAE,UVSQ ENVA, Jouy En Josas, France
关键词
Algorithm; Automatic annotation; Geri (R); IVF; Kinetic events; Time-lapse; DAY; 3; MORPHOLOGY; EMBRYO; IMPLANTATION; MORPHOKINETICS; MICROSCOPY; PREDICTOR; SELECTION;
D O I
10.1016/j.rbmo.2022.02.012
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Research question: What is the reliability of Geri (R) Assess 2.0 software time-lapse technology for annotating kinetic events and identifying abnormal phenotypes in preimplantation human embryos? Design: Embryos were annotated using Assess 2.0 for the appearance and fading of pronuclei, and for progression to the 2-, 3-, 4-, 5- and 6-cell stages and to three blastocyst stages. Identification of reverse cleavage and direct cleavage phenotypes was also recorded. Manual annotation was undertaken after these events in a blinded fashion. Embryo scores were compared between Assess 2.0 and manual annotation. Results: A total of 513 oocytes from 34 women were included. Detection rates for Assess 2.0 versus manual annotation among the 10 kinetic events and including direct cleavage and reverse cleavage ranged between 0% and 94.4%. The percentage of discordant pairs was significantly different for all 12 events analysed (P-value range 0.036 to <0.0001). The sensitivity of Assess 2.0 ranged from 68.2% to 94.4% and specificity ranged from 63.8% to 97.3%. Assess 2.0 called for verification by the embryologist for at least one event in 55.2% of oocytes assessed. Of the 297 embryos scored by manual annotation, Assess 2.0 assigned the same score for only 125 (42.1%), although after manual corrections, concordance with manual annotation scores was raised to 66.0%. Conclusions: The results reveal striking differences between Assess 2.0 and manual annotation for kinetic annotations. Failure of Assess 2.0 to detect direct cleavage events and the low detection rate of reverse cleavage are further limitations. These collective findings highlight the importance of validating time-lapse annotation software before clinical implementation. Manual verification of Assess 2.0 outputs remains essential for accurate data interpretation.
引用
收藏
页码:35 / 45
页数:11
相关论文
共 35 条
[1]   Improved implantation rates of day 3 embryo transfers with the use of an automated time-lapse-enabled test to aid in embryo selection [J].
Adamson, G. David ;
Abusief, Mary E. ;
Palao, Lourella ;
Witmer, Jennifer ;
Palao, Lonyl M. ;
Gvakharia, Marina .
FERTILITY AND STERILITY, 2016, 105 (02) :369-+
[2]   External validation of a time-lapse model; a retrospective study comparing embryo evaluation using a morphokinetic model to standard morphology with live birth as endpoint [J].
Adolfsson, Emma ;
Porath, Sandra ;
Andershed, Anna Nowosad .
JORNAL BRASILEIRO DE REPRODUCAO ASSISTIDA, 2018, 22 (03) :205-214
[3]   Interpretable, not black-box, artificial intelligence should be used for embryo selection [J].
Afnan, Michael Anis Mihdi ;
Liu, Yanhe ;
Conitzer, Vincent ;
Rudin, Cynthia ;
Mishra, Abhishek ;
Savulescu, Julian ;
Afnan, Masoud .
HUMAN REPRODUCTION OPEN, 2021, 2021 (04)
[4]   Novel noninvasive embryo selection algorithm combining time-lapse morphokinetics and oxidative status of the spent embryo culture medium [J].
Alegre, Lucia ;
Del Gallego, Raquel ;
Arrones, Sarai ;
Hernandez, Purificacion ;
Munoz, Manuel ;
Meseguer, Marcos .
FERTILITY AND STERILITY, 2019, 111 (05) :918-+
[5]  
Alpha Scientists in Reproductive Medicine and ESHRE Special Interest Group of Embryology, 2011, Hum Reprod, V26, P1270, DOI 10.1093/humrep/der037
[6]  
[Anonymous], TECHNOTES
[7]   Selection of preimplantation embryos using time-lapse microscopy in in vitro fertilization: State of the technology and future directions [J].
Aparicio-Ruiz, Belen ;
Romany, Laura ;
Meseguer, Marcos .
BIRTH DEFECTS RESEARCH, 2018, 110 (08) :648-653
[8]   Good practice recommendations for the use of time-lapse technology [J].
Apter, Susanna ;
Ebner, Thomas ;
Freour, Thomas ;
Guns, Yves ;
Kovacic, Borut ;
Le Clef, Nathalie ;
Marques, Monica ;
Meseguer, Marcos ;
Montjean, Debbie ;
Sfontouris, Ioannis ;
Sturmey, Roger ;
Coticchio, Giovanni .
HUMAN REPRODUCTION OPEN, 2020, 2020 (02)
[9]   The use of morphokinetics as a predictor of implantation: a multicentric study to define and validate an algorithm for embryo selection [J].
Basile, N. ;
Vime, P. ;
Florensa, M. ;
Aparicio Ruiz, B. ;
Garcia Velasco, J. A. ;
Remohi, J. ;
Meseguer, M. .
HUMAN REPRODUCTION, 2015, 30 (02) :276-283
[10]  
Bori L., FERTIL STERIL, V116, P1