Design optimization of cementless metal-backed cup prostheses using the concept of functionally graded material

被引:6
|
作者
Hedia, H. S. [1 ]
El-Midany, T. T. [1 ]
Shabara, M. A. N. [1 ]
Fouda, N. [1 ]
机构
[1] Mansoura Univ, Fac Engn, Prod Engn & MC Design Dept, Mansoura, Egypt
关键词
D O I
10.1088/1748-6041/1/3/006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Metal backing has been widely used in acetabular cup design. A stiff backing for a polyethylene liner was initially believed to be mechanically favourable. Yet, recent studies of the load transfer around acetabular cups have shown that a stiff backing causes two problems. It generates higher stress peaks around the acetabular rim than those caused by full polyethylene cups and reduces the stresses transferred to the dome of the acetabulum causing stress shielding. The aim of this study is to overcome these two problems by improving the design of cementless metal-backed acetabular cups using the two-dimensional functionally graded material (FGM) concept through finite-element analysis and optimization techniques. It is found that the optimal 2D FGM model must have three bioactive materials of hydroxyapatite, Bioglass and collagen. This optimal material reduces the stress shielding at the dome of the acetabulum by 40% and 37% compared with stainless steel and titanium metal backing shells, respectively. In addition, using the 2D FGM model reduces the maximum interface shear stress in the bone by 31% compared to the titanium metal backing shell.
引用
收藏
页码:127 / 133
页数:7
相关论文
共 50 条
  • [1] Development of cementless metal-backed acetabular cup prosthesis using functionally graded material
    Hedia H.S.
    El-Midany T.T.
    Shabara M.A.N.
    Fouda N.
    International Journal of Mechanics and Materials in Design, 2005, 2 (3-4) : 259 - 267
  • [2] Outcome of metal-backed cementless patellar components - The effect of implant design
    Kraay, MJ
    Darr, OJ
    Salata, MJ
    Goldberg, VM
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2001, (392) : 239 - 244
  • [3] Design optimization of cementless hip prosthesis coating through functionally graded material
    Hedia, H. S.
    Fouda, N.
    COMPUTATIONAL MATERIALS SCIENCE, 2014, 87 : 83 - 87
  • [4] A Method of Material Optimization of Cementless Stem Through Functionally Graded Material
    H.S. Hedia
    M.A.N. Shabara
    T.T. El-Midany
    N. Fouda
    International Journal of Mechanics and Materials in Design, 2004, 1 (4) : 329 - 346
  • [5] Survivorship, Clinical and Radiographic Outcomes of a Novel Cementless Metal-Backed Patella Design
    Grau, Luis C.
    Ong, Alvin C.
    Restrepo, Santiago
    Griffiths, Sean Z.
    Hozack, William J.
    Smith, Eric B.
    JOURNAL OF ARTHROPLASTY, 2021, 36 (07): : S221 - S226
  • [6] Design and analysis of cementless hip-joint system using functionally graded material
    Asiri, Saeed
    ARCHIVE OF MECHANICAL ENGINEERING, 2022, 69 (01) : 129 - 146
  • [7] Improved Stress Shielding on a Cementless Tibia Tray using Functionally Graded Material
    Hedia, Hassan S.
    Fouda, Noha
    MATERIALS TESTING, 2013, 55 (11-12) : 845 - 851
  • [8] Material optimization of a cemented tibia tray using functionally graded material
    Hedia, Hassan S.
    Aldousari, Saad M.
    Fouda, Noha
    MATERIALS TESTING, 2016, 58 (03) : 260 - 268
  • [9] Design optimization of a functionally graded overlay using FEA
    Fouquet, V.
    Tapie, L.
    Attal, J. P.
    Benoit, A.
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2020, 23 : S110 - S112
  • [10] Design of functionally graded structures using topology optimization
    Paulino, GH
    Silva, ECN
    FUNCTIONALLY GRADED MATERIALS VIII, 2005, 492-493 : 435 - 440