Use of temperature programming to improve resolution of inorganic anions, haloacetic acids and oxyhalides in drinking water by suppressed ion chromatography

被引:33
作者
Barron, L
Nesterenko, PN
Paull, B [1 ]
机构
[1] Dublin City Univ, Sch Chem Sci, Natl Ctr Sensor Res, Dublin 9, Ireland
[2] Moscow MV Lomonosov State Univ, Dept Analyt Chem, Moscow 119899, Russia
关键词
suppressed gradient ion chromatography; haloacetic acids; oxylialides; drinking water; temperature programming;
D O I
10.1016/j.chroma.2005.03.028
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Temperature programming was used to improve selectivity in the suppressed ion chromatographic (IQ separation of inorganic anions, haloacetic acids and oxyhalides in drinking water samples when using NaOH gradient elution. The programme exploited varying responses of these anions to changes in temperature. Heats of adsorption (Delta H, kJ/mol) for 17 anionic species were calculated from van't Hoff plots. For haloacetic acids, both the degree of substitution and log P (log of n-octanol-water partition coefficient) values correlated well with the magnitude of the temperature effect, with monochloro- and monobromoacetic acids showing the largest effect (Delta H = -10.4 to -10.7 kJ/mol), dichloro-and dibromoacetic acids showing a reduced effect (Delta H= -6.8 to -8.4 kJ/mol) and trichloro-, bromodichloro- and chlorodibromoacetic acids showing the least effect (Delta H = -4.7 to -2.4 kJ/mol). The effect of temperature on oxyhalides ranged from Delta H = 8.4 kJ/mol for perchlorate to Delta H = -9.1 kJ/mol for iodate. The effectiveness of two commercial column ovens was investigated for the application of temperature gradients during chromatographic runs, with the best system applied to improve the resolution of closely retained species at the start, middle and end of the, separation obtained using a previously optimised hydroxide gradient, in a real drinking water sample matrix. Retention time reproducibility of the final method ranged from 0.62 to 3.18 % RSD (n = 30) showing temperature programming is indeed a practically important parameter to manipulate resolution. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:207 / 215
页数:9
相关论文
共 27 条
[1]   EFFECT OF COLUMN TEMPERATURE ON HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC BEHAVIOR OF INORGANIC POLYPHOSPHATES .1. ISOCRATIC ION-EXCHANGE CHROMATOGRAPHY [J].
BABA, Y ;
YOZA, N ;
OHASHI, S .
JOURNAL OF CHROMATOGRAPHY, 1985, 348 (01) :27-37
[2]   Determination of haloacetic acids in drinking water using suppressed micro-bore ion chromatography with solid phase extraction [J].
Barron, L ;
Paull, B .
ANALYTICA CHIMICA ACTA, 2004, 522 (02) :153-161
[3]   Retention of alkali, alkaline earth and transition metals on an itaconic acid cation-exchange column - Eluent pH, ionic strength and temperature effects upon selectivity [J].
Bashir, W ;
Tyrrell, E ;
Feeney, O ;
Paull, B .
JOURNAL OF CHROMATOGRAPHY A, 2002, 964 (1-2) :113-122
[4]  
BOYD GE, 1970, ION EXCHANGE PROCESS, P261
[5]   High-speed ion chromatographic separation of cations at elevated temperature [J].
Chong, J ;
Hatsis, P ;
Lucy, CA .
JOURNAL OF CHROMATOGRAPHY A, 2003, 997 (1-2) :161-169
[6]   Observations on the effect of temperature on performance and stability of anion exchange columns in ion chromatography [J].
Dybczynski, R ;
Kulisa, K .
CHROMATOGRAPHIA, 2003, 57 (7-8) :475-484
[7]   Ion-exchange properties of glutamic acid-bonded silica [J].
Elefterov, AI ;
Kolpachnikova, MG ;
Nesterenko, PN ;
Shpigun, OA .
JOURNAL OF CHROMATOGRAPHY A, 1997, 769 (02) :179-188
[8]   THE EFFECT OF TEMPERATURE ON SINGLE-COLUMN ION-CHROMATOGRAPHY OF METAL-IONS [J].
FORTIER, NE ;
FRITZ, JS .
TALANTA, 1987, 34 (04) :415-418
[9]   Ion chromatography on chelating stationary phases: separation of alkali metals [J].
Haidekker, A ;
Huber, CG .
JOURNAL OF CHROMATOGRAPHY A, 2001, 921 (02) :217-226
[10]   Effect of temperature on retention and selectivity in ion chromatography of anions [J].
Hatsis, P ;
Lucy, CA .
JOURNAL OF CHROMATOGRAPHY A, 2001, 920 (1-2) :3-11