Evaluation of two pediatric polymethyl pentene membrane oxygenators with pulsatile and non-pulsatile perfusion

被引:14
作者
Qiu, Feng [1 ]
Khan, Sameer [1 ]
Talor, Jonathan [1 ]
Kunselman, Allan
Uendar, Akif [1 ,2 ,3 ]
机构
[1] Penn State Hershey Childrens Hosp, Penn State Hershey Coll Med, Penn State Milton S Hershey Med Ctr, Penn State Hershey Pediat Cardiovasc Res Ctr, Hershey, PA USA
[2] Penn State Hershey Childrens Hosp, Penn State Hershey Coll Med, Penn State Milton S Hershey Med Ctr, Dept Surg, Hershey, PA USA
[3] Penn State Hershey Childrens Hosp, Penn State Hershey Coll Med, Penn State Milton S Hershey Med Ctr, Dept Bioengn, Hershey, PA USA
来源
PERFUSION-UK | 2011年 / 26卷 / 03期
关键词
polymethyl pentene membrane oxygenator; extracorporeal life support; pulsatile perfusion; pediatric; EXTRACORPOREAL LIFE-SUPPORT; NONPULSATILE FLOW; IMPACT; QUALITY; PUMP;
D O I
10.1177/0267659110396116
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: This experiment sought to compare two polymethyl pentene (PMP) hollow-fiber membrane oxygenators: the Medos HILITE 2400 LT and the Maquet Quadrox-iD Pediatric in terms of transmembrane pressure gradients and hemodynamic energy preservation under both pulsatile and non-pulsatile conditions. Methods: A simulated pediatric extracorporeal life support (ECLS) circuit was used to test these two oxygenators. The circuit consisted of a roller pump, 1/4 inch tubing for both arterial and venous lines, an oxygenator, and a venous reservoir served as a pseudo-patient. Three pressure transducers were placed upstream and downstream of the oxygenator and the distal arterial line. The experimental system was primed with lactated Ringer's solution and packed human red blood cells to maintain a hematocrit of 40%. The total volume was 600 ml, including the 350 ml volume of the pseudo-patient. The tests were performed at six flow rates (250, 500, 750, 1000, 1250, 1500 ml/min) and three distal arterial line pressures (MAP) (60, 80, 100 mmHg), under both pulsatile and non-pulsatile perfusion modes. The temperature was kept constant at 37 degrees C for all tests. Results: Both oxygenators had adequate performances in pressure drop and hemodynamic energy preservation. There were no significant differences between pre- and post-oxygenators for mean pressure (MP), energy equivalent pressure (EEP) and total hemodynamic energy (THE). During the pulsatile perfusion mode, the HILITE 2400 LT retained a greater percentage of surplus hemodynamic energy (SHE) across the oxygenator. Conclusions: Both the Quadrox-iD Pediatric and HILITE 2400LT PMP membrane oxygenators are suitable for pediatric ECLS therapy under both non-pulsatile and pulsatile perfusion. An optimized combination of flow rate and MAP should be achieved in order to deliver the maximal pulsatile energy in the extracorporeal circuit.
引用
收藏
页码:229 / 237
页数:9
相关论文
共 14 条
[1]  
[Anonymous], 2004, Applied Longitudinal Analysis
[2]   Comparison of Two Types of Neonatal Extracorporeal Life Support Systems With Pulsatile and Nonpulsatile Flow [J].
Haines, Nikkole ;
Wang, Shigang ;
Myers, John L. ;
Undar, Akif .
ARTIFICIAL ORGANS, 2009, 33 (11) :958-966
[3]   Extracorporeal Life Support Registry Report 2008: Neonatal and Pediatric Cardiac Cases [J].
Haines, Nikkole M. ;
Rycus, Peter T. ;
Zwischenberger, Joseph B. ;
Bartlett, Robert H. ;
Uendar, Akif .
ASAIO JOURNAL, 2009, 55 (01) :111-116
[4]   Performance of polymethyl pentene oxygenators for neonatal extracorporeal membrane oxygenation: a comparison with silicone membrane oxygenators [J].
Khoshbin, E ;
Westrope, C ;
Pooboni, S ;
Machin, D ;
Killer, H ;
Peek, GI ;
Sosnowski, AW ;
Firmin, RK .
PERFUSION-UK, 2005, 20 (03) :129-134
[5]   A comparison of radiographic signs of pulmonary inflammation during ECMO between silicon and poly-methyl pentene oxygenators [J].
Khoshbin, Espeed ;
Dux, Anthony E. W. ;
Killer, Hilliary ;
Sosnowski, Andrzej W. ;
Firmin, Richard K. ;
Peek, Giles J. .
PERFUSION-UK, 2007, 22 (01) :15-21
[6]   The effects of pulsatile flow upon renal tissue perfusion during cardiopulmonary bypass: A comparative study of pulsatile and nonpulsatile flow [J].
Kim, HK ;
Son, HS ;
Fang, YH ;
Park, SY ;
Hwang, CM ;
Sun, K .
ASAIO JOURNAL, 2005, 51 (01) :30-36
[7]   Evolution of the Extracorporeal Life Support Circuitry [J].
Palanzo, David ;
Qiu, Feng ;
Baer, Larry ;
Clark, J. Brian ;
Myers, John L. ;
Uendar, Akif .
ARTIFICIAL ORGANS, 2010, 34 (11) :869-873
[8]   Impact of Tubing Length on Hemodynamics in a Simulated Neonatal Extracorporeal Life Support Circuit [J].
Qiu, Feng ;
Uluer, Mehmet C. ;
Kunselman, Allen ;
Clark, J. Brian ;
Myers, John L. ;
Uendar, Akif .
ARTIFICIAL ORGANS, 2010, 34 (11) :1003-1009
[9]   The Impact of Pump Settings on the Quality of Pulsatility [J].
Rider, Alan R. ;
Ressler, Noel M. ;
Karkhanis, Tushar R. ;
Kunselman, Allen R. ;
Wang, Shigang ;
Uendar, Akif .
ASAIO JOURNAL, 2009, 55 (01) :100-105
[10]   Comparison of Perfusion Quality in Hollow-Fiber Membrane Oxygenators for Neonatal Extracorporeal Life Support [J].
Talor, Jonathan ;
Yee, Stella ;
Rider, Alan ;
Kunselman, Allen R. ;
Guan, Yulong ;
Undar, Akif .
ARTIFICIAL ORGANS, 2010, 34 (04) :E110-E116