Performance of different abutment/implant joints as a result of a sealing agent

被引:4
作者
Seloto, Camila Berbel [1 ]
Strazzi-Sahyon, Henrico Badaoui [1 ]
dos Santos, Paulo Henrique [1 ]
Assuncao, Wirley Goncalves [1 ]
机构
[1] Sao Paulo State Univ, Aracatuba Sch Dent, Dept Dent Mat & Prosthodont, Sao Paulo, Brazil
基金
巴西圣保罗研究基金会;
关键词
Abutment; Implant-supported prosthesis; Preload; Screw; IN-VITRO EVALUATION; BACTERIAL LEAKAGE; MICROBIAL LEAKAGE; PRELOAD MAINTENANCE; RETENTION SCREWS; MARGINAL FIT; INTERFACE; IMPLANTS; TORQUE; MISFIT;
D O I
10.2186/jpr.JPR_D_20_00025
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: This in vitro study aimed to evaluate the effectiveness of a sealing agent in sealing the abutment/implant interface and the preload maintenance of retaining screws after mechanical cycling. Methods: Six groups (n = 12) were evaluated according to the abutment/implant system (external-hexagon implant and UCLA abutments, EHU; Morse taper implant and UCLA abutments, MTU; and Morse taper implant and flexcone abutments, MTF) and the presence of an anaerobic gel sealing agent (control group, no sealing agent; experimental group, sealing agent). Toluidine blue (0.7 mu L) was inserted into each implant and the abutments were attached to the implants using a digital torque wrench to evaluate the sealing of the abutment/implant interface. The specimens were tested through mechanical cycling (1 x 10(6) cycles, 2 Hz, and 130 N). Dye release from the abutment/implant interface was analyzed using a spectrophotometer, and the reverse torque values were obtained using a digital wrench. Reverse torque and dye release data were measured after mechanical cycling and analyzed using ANOVA and Tukey's test (alpha =.05). Results: All experimental groups showed higher reverse torque values than the control groups (P <.05). In general, the MTU and MTF experimental groups, as well as the MTF control group, showed no significant dye release at different periods (P >.05). Conclusions: The use of a sealing agent improved the preload maintenance of screw-retained implant-supported prostheses. The sealing agent was effective in sealing the Morse taper connection.
引用
收藏
页码:489 / 494
页数:6
相关论文
共 46 条
  • [1] Microbial leakage through the implant-abutment interface of morse taper implants in vitro
    Aloise, Joao Pedro
    Curcio, Ricardo
    Laporta, Marcia Zorello
    Rossi, Liliane
    Alvares da Silva, Adriana Madeira
    Rapoport, Abrao
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (03) : 328 - 335
  • [2] Preload Evaluation of Different Screws in External Hexagon Joint
    Assuncao, Wirley Goncalves
    Delben, Juliana Aparecida
    Tabata, Lucas Fernando
    Barao, Valentim Adelino Ricardo
    Gomes, Erica Alves
    Garcia, Idelmo Rangel, Jr.
    [J]. IMPLANT DENTISTRY, 2012, 21 (01) : 46 - 50
  • [3] Effect of unilateral misfit on preload of retention screws of implant-supported prostheses submitted to mechanical cycling
    Assuncao, Wirley Goncalves
    Ricardo Barao, Valentim Adelino
    Delben, Juliana Aparecida
    Gomes, Erica Alves
    Garcia, Idelmo Rangel, Jr.
    [J]. JOURNAL OF PROSTHODONTIC RESEARCH, 2011, 55 (01) : 12 - 18
  • [4] Fracture resistance of single-tooth implant-supported all-ceramic restorations: An in vitro study
    Att, W
    Kurun, S
    Gerds, T
    Strub, JR
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2006, 95 (02) : 111 - 116
  • [5] Besimo CE, 1999, INT J ORAL MAX IMPL, V14, P654
  • [6] Binon PP, 1996, INT J PROSTHODONT, V9, P149
  • [7] Dynamic nature of abutment screw retightening: Finite element study of the effect of retightening on the settling effect
    Bulaqi, Haddad Arabi
    Mashhadi, Mahmoud Mousavi
    Safari, Hamed
    Samandari, Mohammad Mandi
    Geramipanah, Farideh
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2015, 113 (05) : 412 - 419
  • [8] TIGHTENING CHARACTERISTICS FOR SCREWED JOINTS IN OSSEOINTEGRATED DENTAL IMPLANTS
    BURGUETE, RL
    JOHNS, RB
    KING, T
    PATTERSON, EA
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1994, 71 (06) : 592 - 599
  • [9] Survival rate, fracture strength and failure mode of ceramic implant abutments after chewing simulation
    Butz, F
    Heydecke, G
    Okutan, M
    Strub, JR
    [J]. JOURNAL OF ORAL REHABILITATION, 2005, 32 (11) : 838 - 843
  • [10] Byrne Declan, 2006, J Prosthodont, V15, P164, DOI 10.1111/j.1532-849X.2006.00096.x