Development and characterization of a dynamic lesion phantom for the quantitative evaluation of dynamic contrast-enhanced MRI

被引:11
|
作者
Freed, Melanie [1 ,2 ]
de Zwart, Jacco A. [3 ]
Hariharan, Prasanna [4 ]
Myers, Matthew R. [4 ]
Badano, Aldo [1 ]
机构
[1] US FDA, Div Imaging & Appl Math, Off Sci & Engn Labs, Ctr Devices & Radiol Hlth, Silver Spring, MD 20993 USA
[2] Univ Maryland, Dept Bioengn, College Pk, MD 20742 USA
[3] Natl Inst Neurol Disorders & Stroke, Lab Funct & Mol Imaging, Adv MRI Sect, NIH, Bethesda, MD 20892 USA
[4] US FDA, Div Solid & Fluid Mech, Off Sci & Engn Labs, Ctr Devices & Radiol Hlth, Silver Spring, MD 20993 USA
关键词
dynamic phantom; DCE-MRI; breast imaging; SPOILED GRADIENT-ECHO; MYOCARDIAL-PERFUSION; DIAGNOSTIC-ACCURACY; BREAST MRI; PARAMETERS; SEQUENCES; RESONANCE; DTPA; STANDARDIZATION; QUANTIFICATION;
D O I
10.1118/1.3633911
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To develop a dynamic lesion phantom that is capable of producing physiological kinetic curves representative of those seen in human dynamic contrast-enhanced MRI (DCE-MRI) data. The objective of this phantom is to provide a platform for the quantitative comparison of DCE-MRI protocols to aid in the standardization and optimization of breast DCE-MRI. Methods: The dynamic lesion consists of a hollow, plastic mold with inlet and outlet tubes to allow flow of a contrast agent solution through the lesion over time. Border shape of the lesion can be controlled using the lesion mold production method. The configuration of the inlet and outlet tubes was determined using fluid transfer simulations. The total fluid flow rate was determined using x-ray images of the lesion for four different flow rates (0.25, 0.5, 1.0, and 1.5 ml/s) to evaluate the resultant kinetic curve shape and homogeneity of the contrast agent distribution in the dynamic lesion. High spatial and temporal resolution x-ray measurements were used to estimate the true kinetic curve behavior in the dynamic lesion for benign and malignant example curves. DCE-MRI example data were acquired of the dynamic phantom using a clinical protocol. Results: The optimal inlet and outlet tube configuration for the lesion molds was two inlet molds separated by 30 degrees and a single outlet tube directly between the two inlet tubes. X-ray measurements indicated that 1.0 ml/s was an appropriate total fluid flow rate and provided truth for comparison with MRI data of kinetic curves representative of benign and malignant lesions. DCE-MRI data demonstrated the ability of the phantom to produce realistic kinetic curves. Conclusions: The authors have constructed a dynamic lesion phantom, demonstrated its ability to produce physiological kinetic curves, and provided estimations of its true kinetic curve behavior. This lesion phantom provides a tool for the quantitative evaluation of DCE-MRI protocols, which may lead to improved discrimination of breast cancer lesions. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3633911]
引用
收藏
页码:5601 / 5611
页数:11
相关论文
共 50 条
  • [1] Feasibility testing of a motion phantom for quantitative Dynamic contrast-enhanced MRI
    Lye, Jessica
    Cahoon, Glenn
    Shelton, Nikki
    Alinaghi-Zadeh, Reza
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S3895 - S3898
  • [2] Quantitative evaluation of contrast agent uptake in standard fat-suppressed dynamic contrast-enhanced MRI examinations of the breast
    Kousi, Evanthia
    Smith, Joely
    Ledger, Araminta E.
    Scurr, Erica
    Allen, Steven
    Wilson, Robin M.
    O'Flynn, Elizabeth
    Pope, Romney J. E.
    Leach, Martin O.
    Schmidt, Maria A.
    MEDICAL PHYSICS, 2018, 45 (01) : 287 - 296
  • [3] Comparison of Dynamic Contrast-Enhanced MRI and Dynamic Contrast-Enhanced CT Biomarkers in Bladder Cancer
    Naish, J. H.
    McGrath, D. M.
    Bains, L. J.
    Passera, K.
    Roberts, C.
    Watson, Y.
    Cheung, S.
    Taylor, M. B.
    Logue, J. P.
    Buckley, D. L.
    Tessier, J.
    Young, H.
    Waterton, J. C.
    Parker, G. J. M.
    MAGNETIC RESONANCE IN MEDICINE, 2011, 66 (01) : 219 - 226
  • [4] Dynamic contrast-enhanced MRI for oncology drug development
    Sung, Yu Sub
    Park, Bumwoo
    Choi, Yoonseok
    Lim, Hyeong-Seok
    Woo, Dong-Cheol
    Kim, Kyung Won
    Kim, Jeong Kon
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2016, 44 (02) : 251 - 264
  • [5] Uncertainty estimation in dynamic contrast-enhanced MRI
    Garpebring, Anders
    Brynolfsson, Patrik
    Yu, Jun
    Wirestam, Ronnie
    Johansson, Adam
    Asklund, Thomas
    Karlsson, Mikael
    MAGNETIC RESONANCE IN MEDICINE, 2013, 69 (04) : 992 - 1002
  • [6] Comparison of dynamic contrast-enhanced MRI and quantitative SPECT in a rat glioma model
    Skinner, Jack T.
    Yankeelov, Thomas E.
    Peterson, Todd E.
    Does, Mark D.
    CONTRAST MEDIA & MOLECULAR IMAGING, 2012, 7 (06) : 494 - 500
  • [7] Dynamic Contrast-Enhanced (DCE) MRI
    Li, Xin
    Huang, Wei
    Holmes, James H.
    MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2024, 32 (01) : 47 - 61
  • [8] Dynamic contrast-enhanced MRI in oncology: how we do it
    Petralia, Giuseppe
    Summers, Paul E.
    Agostini, Andrea
    Ambrosini, Roberta
    Cianci, Roberta
    Cristel, Giulia
    Calistri, Linda
    Colagrande, Stefano
    RADIOLOGIA MEDICA, 2020, 125 (12): : 1288 - 1300
  • [9] Quantitative evaluation of Crohn's disease using dynamic contrast-enhanced MRI in children and young adults
    Lee, Seunghyun
    Choi, Young Hun
    Cho, Yeon Jin
    Cheon, Jung-Eun
    Moon, Jin Soo
    Kang, Gyeong Hoon
    Kim, Woo Sun
    EUROPEAN RADIOLOGY, 2020, 30 (06) : 3168 - 3177
  • [10] Quantitative Diffusion-Weighted Imaging and Dynamic Contrast-Enhanced Characterization of the Index Lesion With Multiparametric MRI in Prostate Cancer Patients
    Yuan, Qing
    Costa, Daniel N.
    Senegas, Julien
    Xi, Yin
    Wiethoff, Andrea J.
    Rofsky, Neil M.
    Roehrborn, Claus
    Lenkinski, Robert E.
    Pedrosa, Ivan