Fabrication of Non-enzymatic Ni-Au Alloy Nanowire Glucose Sensor

被引:6
作者
Hsieh, Yi-Shu [1 ]
Wang, Po-Wen [1 ]
Li, Chien Yu [1 ]
Hsieh, Shang-Ju [1 ]
Wang, Ching-Yu [1 ]
Chou, Dei-Wei [2 ]
Wang, Na-Fu [3 ]
Houng, Mau-Phon [1 ]
机构
[1] Natl Cheng Kung Univ, Inst Microelect, Dept Elect Engn, 1 Univ Rd, Tainan 701, Taiwan
[2] Air Force Inst Technol, Dept Mil Meteorol, Kaohsiung 820, Taiwan
[3] Cheng Shiu Univ, Super Micro Mass Res & Technol Ctr, Ctr Environm Toxin & Emerging Contaminant Res, Dept Elect Engn, 840 Chengcing Rd, Kaohsiung 83347, Taiwan
关键词
non-enzymatic; glucose sensor; Ni-Au alloy nanowire; anodic aluminum oxide (AAO); ARRAYS;
D O I
10.18494/SAM.2020.2479
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We fabricated a non-enzymatic Ni-Au alloy nanowire electrochemical glucose sensor on a p-silicon-based anodic aluminum oxide (AAO) template. The advantages of the Ni-Au glucose sensor arc its high sensitivity and stability, and fast reaction. In contrast to a traditional enzymatic glucose sensor, the non-enzymatic glucose sensor is reusable, enabling its long-term use by patients with diabetes. Although the reduction voltages of Ni and Au are considerably different, the Ni-Au alloy nanowires used in the sensor were fabricated successfully via the adjustment of the electrodeposition parameters. The Ni-Au alloy nanowires exhibited a high crystallinity and a uniform arrangement with an average height of 750 nm. The Ni-Au alloy glucose sensor exhibited a linear range of 0-3 mM, a sensitivity of 1893 mu A/mMcm(2), and a detection limit of 1 mu M. Superior selectivity and stability over at least 30 days were also observed. The characteristics show that Ni-Au alloy nanowires have excellent glucose-sensing performance compared with single noble metal nanowires. The silicon-based AAO template also serves as a strong mechanical support of the Ni-Au nanowires used for sensing. To our knowledge, our fabricated sensor is the first non-enzymatic glucose sensor to use Ni-Au alloy nanowires.
引用
收藏
页码:1843 / 1850
页数:8
相关论文
共 20 条
[11]   A novel non-enzymatic glucose sensor based on Cu nanoparticle modified graphene sheets electrode [J].
Luo, Jing ;
Jiang, Sisi ;
Zhang, Hongyan ;
Jiang, Jinqiang ;
Liu, Xiaoya .
ANALYTICA CHIMICA ACTA, 2012, 709 :47-53
[12]   A MINIATURE OPTICAL GLUCOSE SENSOR BASED ON AFFINITY BINDING [J].
MANSOURI, S ;
SCHULTZ, JS .
BIO-TECHNOLOGY, 1984, 2 (10) :885-890
[13]   Electrochemical non-enzymatic glucose sensors [J].
Park, S ;
Boo, H ;
Chung, TD .
ANALYTICA CHIMICA ACTA, 2006, 556 (01) :46-57
[14]   Synthesis of Ni/Au multilayer nanowire arrays for ultrasensitive non-enzymatic sensing of glucose [J].
Qin, Lirong ;
He, Lizhong ;
Zhao, Jianwei ;
Zhao, Binglin ;
Yin, Yingying ;
Yang, Yu .
SENSORS AND ACTUATORS B-CHEMICAL, 2017, 240 :779-784
[15]  
Qu S., 2014, SENSOR ACTUAT B-CHEM, V203, P588
[16]  
Slate J. B., 1997, U. S. Patent, Patent No. 5605152
[17]  
Weiwei L., 2016, ELECTROCHIM ACTA, V188, P197
[18]   GLUCOSE-OXIDASE - AN IDEAL ENZYME [J].
WILSON, R ;
TURNER, APF .
BIOSENSORS & BIOELECTRONICS, 1992, 7 (03) :165-185
[19]  
Zhi-Da G., 2013, SCI REP, V3, P3323
[20]   CNT-network modified Ni nanostructured arrays for high performance non-enzymatic glucose sensors [J].
Zhu, Jianhui ;
Jiang, Jian ;
Liu, Jinping ;
Ding, Ruimin ;
Li, Yuanyuan ;
Ding, Hao ;
Feng, Yamin ;
Wei, Guangming ;
Huang, Xintang .
RSC ADVANCES, 2011, 1 (06) :1020-1025